Catalog excerpts
Cooled frame transfer EMCCD camras PSL has supplied cooled frame transfer EMCCD cameras for the last 5 years to end users and OEMs. A selection of high responsitivity CCDs, combined with low noise high built EM gain characteristics, enables optimum photonic collection with best possible signal to noise ratio. Special read whilst expose mode allows 100% duty cycle and high sensitivity operation in low light level conditions. Applications: ' Fluorescence imaging ' Astronomy 1 TIRF microscopy ' Confocal microscopy / cell screening 1 Chemiluminescence ' Spectroscopy 1 Single molecule imaging 1 Cell motility / live cell recording ' Solar panel characterisation ' Electron microscopy ' Biochip reader ' Raman spectroscopy Photonic Science ,y\/\/v.-- Information / products and services Scientific detector systems
Open the catalog to page 1Cooled EMCCD camras Photonic Science Ltd selects premium grade EMCCD sensors : Small pixel size less < 10 microns pixel size 镕 Cooled CCD sensor with 55 degrees C delta T 17.5 MHz frequency Օ Sensor size from 6 and 11 mm diagonal Quartz, glass and or fibre optic input windows Very low readout noise < 25 electrons High EM gain : up to > 2000 enabling sub electron readout noise with near single photon counting detection capability EM gain calibration and gain compensation for quantitative measurements Very low dark current with less than 0.5 electrons per pixel per second (cooling options...