1. Catalogs
  2. Photo Research, Inc.
  3. PR-920 Digital Video Photometer

PR-920 Digital Video Photometer

PR-920 Digital Video Photometer

PR-920 Digital Video Photometer

Product catalog summary
Introduction
The PR-920 Digital Video Photometer (DVP) is the latest innovation from Photo Research, building on over 15 years of experience in video photometry. It is designed for a wide range of applications, including FAA testing and ISO certification.

Specifications
The PR-920 features a thermoelectrically cooled 1024 x 1024 element digital CCD camera, ensuring high accuracy and spatial resolution. It includes hand-matched filters to the CIE V(λ) response, with optional CIE Tristimulus filters available. The device is equipped with a focusable objective lens and five automated neutral density filters, offering a broad luminance range.

Standard Equipment
The system includes an optical head, PCI frame grabber, cables, and dual filter turrets. Calibration certification and an instruction manual are provided, while VideoWin™3 software is sold separately.

Features
The PR-920 offers luminance measurements, pseudo color luminance display, 3D luminance profile graph, and Active X control for integration with other applications. It also supports symbol luminance analysis.

Options
Optional features include CIE Tristimulus filters, a Pentium-based PC controller, high-resolution positioning stages, and absolute spatial calibration.

Measurement Capabilities
The device measures luminance, luminance uniformity, pseudo color representation, multiple area of regard analysis, line width, and CIE chromaticity.

Performance Specifications
The PR-920 offers 16-bit resolution with a pixel size of 0.012 mm. It supports a variable focusing range from 73 mm to infinity and a minimum measuring area of 0.06 mm. The luminance range is 0.03 to 85,000 cd/m², with a calibration uncertainty of ±5% absolute and ±2% relative to NIST standards.

Environmental and Power Requirements
The device operates at temperatures between 5° to 30°C and requires 90 to 240 VAC power.

Lens Capability
The PR-920 supports various magnifications and working distances, with detailed specifications provided for each lens type.
See more

Catalog excerpts

PR-920 Digital Video Photometer-1

PR-920 Digital Video Photometer Innovation - Photo Research developed the first commercially available video photometer over 15 years ago. We are proud to introduce our newest and most advanced video photometer - the PR-920 Digital Video Photometer (DVP). Our video photometers feature the patented Synthetic Aperture data analysis method and numerous other patents and are beng used world-wide for applications rangmg from FAA testing of radar screens to ISO certification of information dis* plays. Digital Prcision - The heart of the PR-920 is a thermo-electrically cooled. 1024 x 1024 響lment digital CCD camra which helps to make it the most accurate, sensitive and repeatable video photometer in rts class. Because the PR-920 detector consists of 1.048 million pixels, excellent spa驭tial resolution is guaranteed. Hand Matched Filters - In addition to the proprietary camra design, the detector response of each PR-920 is painstakingly matched to the CIE V(X) (photopic) response by our proprietary fitter matching techiques that have been devel- 6 32 m (16.1 cm» o p e d VideoWin 3* with 3D luminance graph over the last 50 years. Our filter matching exprience, combined with an individual calibration of each and every detector 驩lment means that the PR-920 delivers the most accurate measurement results of any commercially available video photometer In addition to the standard photopically matched filter, the PR-920 can be equipped with high* accuracy, hand-matched CIE Tristimulus filters (option CFS-302). Versatility - The PR-920 is the only commercially available video photometer that utilizes a focusable objective lens as standard equipment r 81 in (19 8 cm providing the capability of addressing virtually any application without changing lenses- The PR-920 is also the only video photometer that includes 5 intemally mounted, fijlly automated neu-tral density filters giving it the widest standard luminance range of any instrument in ifs class. Standard Equipment - System - Optical Head. PCI Frame Grabber and Cables; Optical Head - CCD Cam魩ra, Dual Filter Turrets (CIE Photopic and 5 Neutral Density filters); Calibration - Certification and Instruction Manual. VideoWin m3 software must be ordered separately. Features: * Luminance measurements ► Psuedo Color Luminance Display * 3D Luminance Profile graph ■ Active X Control from other applications (e.g. Microsoft Excel, Visual C++) * Symbol Luminance - Automatically select images for analysis based on luminance value Options: CIE Tristimulus (Xrt Xbf Z) Filters (P/N CFS-302) Controller; Pentium based PC running MS Windows XP High Resolution Positioning Stages Absolute spatial calibration (line width, profile capability) Measurement Capabilities Luminance Օ Multiple Area of Regard (AOR) Analysis Luminance Uniformity and Profile Օ Line Width, Profile (opt.) . Pseudo Color Luminance Reprsentation * CIE Chromaticity (opt)

 Open the catalog to page 1
PR-920 Digital Video Photometer-2

PR-920 Performance Spcifications Resolution: 16Bit CCD Pixels -1024 x 1024 CCD Pixel Size 0*012 mm x 0.012 mm -100% tll factor Standard Filters Photoplc(f1'*3%), ND-0.2, 0.5, 1.0,1.5, 2A OPEN Variable Focuslng Range: 73 mm to Inflnlty with standard MS-60 Lens Minimum Measurlng Area: a 0.06 mm with MS-60 uslng 5x5 pixels Field of View 鮣 12.2 mm with standard MS-60 Lens (see Lens chart) Measurement Time Range; 100 msec to 60 sec Calibration Uncertatnty: ± 5% (Absolute) 12% (Relative to NIST traceable 28S6 K standard) Luminance Prcision: 11.0% at 0.4 cd/m2 agalnst 2856K standard 110x10 pixel avg.)...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.