video corpo

UV IQ/OQ

UV IQ/OQ

UV IQ/OQ

Product catalog summary
Introduction
The document outlines the purpose and components of the Qualification Kit designed for PG Instruments spectrophotometers. It is intended to provide flexibility for laboratory analysts and distributors to perform full installation and operational qualifications.
Specifications and Compliance
The Qualification Kit is developed in accordance with the European Pharmacopoeia requirements to ensure compliance with good laboratory practices. It includes detailed instructions for conducting tests, documenting results, and performing corrective actions if necessary.
Contents of the Qualification Kit
  • Holmium Oxide Solution: Used for measuring wavelength accuracy (NIST srm 2034).
  • Potassium Dichromate Solution: Used for measuring photometric accuracy (NIST srm 935a).
  • Potassium Chloride: Used for measuring stray light at 200nm.
  • Blank Solution: Provided for baseline measurements.
  • Two 10mm Path Length Quartz Cuvettes: Required for validation processes.
  • Certificate for Traceability: Ensures the traceability of measurements.
  • Flash Drive: Contains a printable Qualification Workbook and Worksheets.
Software Requirement
UVWin software is mandatory for performing instrument qualification, ensuring that all procedures are correctly followed and documented.
Manufacturer Information
PG Instruments Ltd, located in Leicestershire, England, reserves the right to modify or discontinue any product without notice.
See more

Catalog excerpts

UV IQ/OQ-1

Analytical Instruments for Science The purpose of the Qualification Kit is to offer both laboratory analysts and regional distributors the flexibility to perform a full installation and operational qualification on the range of PG Instruments spectrophotometers.

 Open the catalog to page 1
UV IQ/OQ-2

instruments Analytical Instruments for Science The work instructions within the qualification The contents of the package are as follows: workbook have been written with considerations Pharmacopoeia to ensure compliance with good Holmium Oxide Solution – For measuring wavelength accuracy (NIST srm 2034). Potassium Dichromate Solution – laboratory practice. For measuring photometric accuracy The documentation supplied with the kit guides the user through the qualification as it offers detailed instructions on how to carry out the Potassium Chloride – For measuring stray light @ 200nm. tests, document...

 Open the catalog to page 2

All PG INSTRUMENTS LIMITED catalogs and technical brochures

  1. UV-Win6

    102  Pages

  2. GC650

    12  Pages

  3. FP902

    4  Pages

  4. UV-WinGLP

    4  Pages

  5. T110+/T112+

    2  Pages

  6. T92+

    4  Pages

  7. T85

    4  Pages

  8. T80

    4  Pages

  9. T70

    4  Pages

  10. T75

    4  Pages

  11. T60

    4  Pages

  12. C30/C30M

    4  Pages

  13. Micron-Mist

    2  Pages

  14. AA990F

    2  Pages

  15. FP910

    4  Pages

  16. AF420

    4  Pages

  17. PG990

    2  Pages

  18. AA500

    4  Pages

  19. ICP 5000

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.