video corpo

T110+/T112+

T110+/T112+

T110+/T112+

Product catalog summary
Introduction
The T110+/T112+ Series Spectrophotometer is designed for advanced UV-Visible spectroscopy applications, featuring dual monochromator technology for precise measurements. It supports wireless control and data acquisition via Wi-Fi, enhancing laboratory mobility.

Specifications
  • Wavelength Range: 185–900 nm
  • Wavelength Accuracy: ±0.2 nm
  • Wavelength Reproducibility: ≤0.1 nm (D2 lamp)
  • Spectral Bandwidth: 0.1–50 nm, continually adjustable
  • Stray Light: ≤0.00004%T (NaI, 220 nm)
  • Photometric Range: -8.0 to 8.0 Abs
  • Photometric Accuracy: ±0.004A @2.0 A
  • Photometric Reproducibility: ≤0.002A @2.0 A
  • Baseline Flatness: ±0.0008 Abs
  • Noise: ≤0.01% at 0% T, ≤0.1% at 100% T
  • Communication Ports: RS232C, USB, Wi-Fi


Features
The spectrophotometer includes a range of accessories for specific sample requirements, such as integrating spheres for diffuse reflectance, polarizing optics, thermostatic cell holders, and automated cell changers. It also supports tablet dissolution for pharmaceutical quality control.

Optical System
The dual monochromator double beam system ensures low stray light and extensive photometric range, with capabilities for deep UV measurements using nitrogen-purged optics.

Contact Information
PG Instruments Ltd, Alma Park, Wibtoft, Leicestershire, LE17 5BH, England. Tel: 0044 (0) 1455 220131, Fax: 0044 (0) 1455 220025, Email: [email protected], Website: www.pginstruments.com
See more

Catalog excerpts

T110+/T112+-1

instruments Analytical Instruments for Science The T110+/T112+ Series Spectrophotometer incorporates dualmonochromator technology making it well suited to even themost demanding of applications in all areas of UV-VisibleSpectroscopy including: • Pharmaceutical • Material Science • Metrological Verification • Food Safety Wireless instrument control and data acquisition can be achieved from a Tablet PC using Wi-Fi technology allowing the analyst to move freely around the laboratory whilst also keeping track of sample me

 Open the catalog to page 1
T110+/T112+-2

The optical design of both the T110+ and T112+ offerextremely low stray light characteristics (<0.00004%TNaI,220 nm) which allows for an extensive photometricrange (-8.0 - 8.0Abs). Measurements at deep ultra-violetwavelengths can also be achieved with use of Nitrogenpurged optics. The instrument can be optically configured to suit the needs of the sample by means of a continually adjustable slit for precise control of spectral resolution and beam size adjustment by means of an attenuating wheel. Precise wavelength accuracy is ensured by the integrated Mercury Emissions Lamp used for automatic correction...

 Open the catalog to page 2

All PG INSTRUMENTS LIMITED catalogs and technical brochures

  1. UV-Win6

    102  Pages

  2. GC650

    12  Pages

  3. FP902

    4  Pages

  4. UV-WinGLP

    4  Pages

  5. T92+

    4  Pages

  6. T85

    4  Pages

  7. T80

    4  Pages

  8. T70

    4  Pages

  9. T75

    4  Pages

  10. T60

    4  Pages

  11. C30/C30M

    4  Pages

  12. Micron-Mist

    2  Pages

  13. AA990F

    2  Pages

  14. FP910

    4  Pages

  15. AF420

    4  Pages

  16. PG990

    2  Pages

  17. AA500

    4  Pages

  18. ICP 5000

    4  Pages

  19. UV IQ/OQ

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.