video corpo

Atomic Absorption

Atomic Absorption

Atomic Absorption

Product catalog summary
Overview
The AA500 Atomic Absorption Spectrometer is a high-performance automated instrument designed for modern laboratories. It is versatile and suitable for applications in agriculture, clinical, environmental, food, metal, mining, and petrochemical sectors.

Configurations
  • AA500F: Equipped with a Flame atomiser only. Offers three flame configurations: air/acetylene (standard), N2O/acetylene, and air/LPG (optional), all with coded burners for safety.
  • AA500G: Features a Graphite Furnace Atomiser only, with a fixed graphite head in the optical path to enhance performance and reduce drift.
  • AA500FG: Combines both Flame and Graphite Furnace Atomisers, allowing easy switching via the AA-Win software.

Specifications
  • Wavelength Range: 190nm – 900nm
  • Monochromator: Czerny-Turner Configuration
  • Spectral Bandwidth: 0.1nm to 2.0nm (software selectable)
  • Wavelength Accuracy: ± 0.15nm
  • Resolution: 0.2nm ± 0.02nm
  • Sensitivity (Cu): 2 µg/ml Absorption > 0.28A (flame)
  • Detection Limit: Cu < 0.004 µg/ml (flame), Cd < 0.4 x 10-12g (graphite furnace)
  • Repeatability: Cu < 0.7% (Air/Acetylene flame), Ba < 1.0% (N2O/Acetylene flame)
  • Background Correction: Deuterium Arc, Self reversal

Flame Atomiser Features
  • Three flame systems: Air/acetylene, N2O/acetylene, Air/Propane (LPG)
  • Safety features include gas pressure monitoring, burner identification, flame sensor, and gas leak detector.

Graphite Furnace Atomiser Features
  • Available in two formats: fixed in light path (AA500G) or motorised (AA500FG)
  • Efficient heating and cooling with feedback system
  • Safety features include Argon Gas Pressure Sensor and Over Temperature Sensor

Standard Features
  • Embedded PC system with Windows 7 Professional
  • Full software control with pre-installed AA-Win software
  • Automatic 8 lamp turret and universal autosampler
See more

Catalog excerpts

Atomic Absorption-1

AA500 DATA SHEET – 011 27/2/12 14:33 Page 1 instruments instruments Analytical Instruments for Science Analytical Instruments for Science Specifications Wavelength Range 190nm – 900nm Monochromator Czerny-Turner Configuration Spectral Bandwidth 0.1nm, 0.2nm, 0.4nm, 1.0nm, 2.0nm, (software selectable). Wavelength Accuracy ± 0.15nm Wavelength Reproducibility < 0.05nm Resolution 0.2nm ± 0.02nm Baseline Stability 0.005A/30min Sensitivity (Cu) 2 µg/ml Absorption > 0.28A (flame) Detection Limit Cu < 0.004 µg/ml (flame) Cd < 0.4 x 10-12g (graphite furnace) Repeatability Cu < 0.7% (Air/Acetylene flame) Ba < 1.0% ( Nitrous oxide/Acetylene flame) Cu < 2.0% Cd < 2.0% (Graphite Furnace) Background Correction Deuterium Arc, Self reversal Characteristic Concentration Cu < 0.02 µg/ml, Ba < 0.15 mg/ml (N2O/Acetylene) Burner Heads Titanium Alloy Nebuliser High-efficiency glass, Acid proof available as an option Atomization Chamber Corrosion-resistant material Position Adjustment Automatic changeover (AA500GF) Manual (AA500F) Automatic Setting of Optimum Height for Flame Burner. Safety Functions Burner Identification, Flame Sensor, Gas leak Sensor, Low Gas Pressure Sensor, Drain Trap Sensor, Power Loss Protection, Circulation Water (graphite), Over Temperature Sensor (graphite) AA500 AT O M I C A B S O R P T I O N S P E C T R O M E T E R The AA500 Atomic Absorption Spectrometer is a high performance automated instrument designed to meet the requirements of the modern laboratory. Due to its versatility and performance it can be used for a wide range of applications including: Agriculture, Clinical, Environmental, Food, Metal, Mining and Petrochemical. The Instrument is available in three configurations: ● AA500F – The Instrument is equipped with a Flame atomiser only. 3 flames are available with the air/acetylene as standard configuration and N2O/acetylene or air/LPG(natural gas) available as an option. All 3 flame configurations offer coded burner for full safety protection. ● AA500G – The Instrument is equipped with a Graphite Furnace Atomiser only. The Graphite head is fixed into the optical path to maximise performance and eliminate drift. The transversally heated graphite tube is efficiently heated and cooled due to the precision feedback system. ● PG Instruments Ltd, Alma Park, Wibtoft, Leicestershire, LE17 5BH, England Tel: 0044 (0) 1455 220131 Fax: 0044 (0) 1455 220025 www.pginstruments.com Email: [email protected] AA500FG – The Instrument is equipped with both Flame and Graphite Furnace Atomiser’s as detailed in the previous description. Both configurations are installed into the instrument and can be changed over by a simple selection in the versatile AA-Win software.

 Open the catalog to page 1
Atomic Absorption-2

AA500 DATA SHEET – 011 27/2/12 14:33 Page 3 instruments Analytical Instruments for Science S TA N D A R D F E AT U R E S ● ● ● ● ● ● ● ● Embedded PC system built into the instrument as standard. Windows 7 professional operating system. Full software control of the instrument and autosampler. Pre-installed AA-Win software. Automatic 8 lamp turret controlled and optimized by the AA-Win software. D2 lamp and Self reversal background systems fitted as standard. High precision minimal optics ensures maximum light throughput to the computer controlled Czerny-Turner monochromator. A universal autosampler...

 Open the catalog to page 2
Atomic Absorption-3

AA500 DATA SHEET – 011 27/2/12 14:33 Page 3 instruments Analytical Instruments for Science S TA N D A R D F E AT U R E S ● ● ● ● ● ● ● ● Embedded PC system built into the instrument as standard. Windows 7 professional operating system. Full software control of the instrument and autosampler. Pre-installed AA-Win software. Automatic 8 lamp turret controlled and optimized by the AA-Win software. D2 lamp and Self reversal background systems fitted as standard. High precision minimal optics ensures maximum light throughput to the computer controlled Czerny-Turner monochromator. A universal autosampler...

 Open the catalog to page 3
Atomic Absorption-4

AA500 DATA SHEET – 011 27/2/12 14:33 Page 1 instruments instruments Analytical Instruments for Science Analytical Instruments for Science Specifications Wavelength Range 190nm – 900nm Monochromator Czerny-Turner Configuration Spectral Bandwidth 0.1nm, 0.2nm, 0.4nm, 1.0nm, 2.0nm, (software selectable). Wavelength Accuracy ± 0.15nm Wavelength Reproducibility < 0.05nm Resolution 0.2nm ± 0.02nm Baseline Stability 0.005A/30min Sensitivity (Cu) 2 µg/ml Absorption > 0.28A (flame) Detection Limit Cu < 0.004 µg/ml (flame) Cd < 0.4 x 10-12g (graphite furnace) Repeatability Cu < 0.7% (Air/Acetylene flame)...

 Open the catalog to page 4

All PG INSTRUMENTS LIMITED catalogs and technical brochures

  1. UV-Win6

    102  Pages

  2. GC650

    12  Pages

  3. FP902

    4  Pages

  4. UV-WinGLP

    4  Pages

  5. T110+/T112+

    2  Pages

  6. T92+

    4  Pages

  7. T85

    4  Pages

  8. T80

    4  Pages

  9. T70

    4  Pages

  10. T75

    4  Pages

  11. T60

    4  Pages

  12. C30/C30M

    4  Pages

  13. Micron-Mist

    2  Pages

  14. AA990F

    2  Pages

  15. FP910

    4  Pages

  16. AF420

    4  Pages

  17. PG990

    2  Pages

  18. AA500

    4  Pages

  19. ICP 5000

    4  Pages

  20. UV IQ/OQ

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.