video corpo

NM18-45L
1 /12Pages

NM18-45L

NM18-45L
1 /12Pages

Catalog excerpts

NM18-45L-1

NM18-45L Site Preparation Guide

 Open the catalog to page 1
NM18-45L-2

Site Preparation GuideContents Customer Responsibilities 3 Safety Notice to Users 5

 Open the catalog to page 2
NM18-45L-3

NM18-45L Site Preparation Guide Customer Responsibilities To ensure a successful and timely installation of your Peak generator, please refer to this set of requirements. Correct site preparation is the first key step in ensuring that your generator and systems operate reliably over an extended lifetime. This document is an information guide and checklist that outlines the requirements for your site. It is essential your site meets the following specification prior to the installation date. For details, see specific sections within this document The necessary spatial requirements are met. The...

 Open the catalog to page 3
NM18-45L-4

Site Preparation GuideChange History Rev.

 Open the catalog to page 4
NM18-45L-5

Site Preparation Guide Safety Notices Symbols This manual uses the following symbols to highlight specific areas important to the safe and proper use of the Generator: This site preparation guide must be read thoroughly and understood before installation and operation of your Peak NM Generator. Incorrect preparation or use of the Generator in a manner not specified by Peak Scientific MAY impair the SAFETY provided by the equipment. When handling, operating or preparing for installation, personnel must employ safe practices and observe all relevant local health and safety requirements and regulations. The...

 Open the catalog to page 5
NM18-45L-6

Site Preparation GuideTechnical Specifications

 Open the catalog to page 6
NM18-45L-7

NM18-45L Site Preparation Guide Site preparation Environmental control The generator must remain on its feet to allow air intake from the bottom of the generator. If the generator is stored in an enclosed space the environment must be controlled via an air conditioner or extraction fan. Generator MUST NOT be stored or installed outside.

 Open the catalog to page 7
NM18-45L-8

Failing to provide adequate cooling space around the generator may cause damage to the membranes. This will reduce service life and invalidate warranty.

 Open the catalog to page 8
NM18-45L-9

NM18-45L Site Preparation Guide On-Site Transit When moving the generator in its shipping crate, doorways and other openings such as elevators must fit with the sizes in the figure below.

 Open the catalog to page 9
NM18-45L-10

Site Preparation Guide Unpacking Space Required The image below shows the minimum space required to unpack the generator from its shipping box. Unboxing The top of the shipping box should be carefully opened and the cardboard insert removed. The generator weighs 20 kg max and as such Peak recommends a two-man lift. Ensure safe lifting practices are adhered to. All packaging should be retained for future transit. Page 10 2014 © Peak Scientific - Rev. 0 - 15/04/14

 Open the catalog to page 10
NM18-45L-11

Site Preparation Guide Tubing lengths Tubing sizes should be chosen with accordance to the diagram below. The diameter of the tubing which will be connected to the gas outlet is important and is determined by the length of tubing required. Failure to follow these recommendations could lead to accelerated compressor wear. Copper Tubing

 Open the catalog to page 11
NM18-45L-12

Peak Scientific India 202, Amsri Shamira Old Lancer Line Opp. St. Mary's Degree Coll S.D. Road Secunderabad 500 025, India Tel: +91 40 2780 0663 Fax: +91 40 2780 0663 Peak Scientific Mexico Solon 352 Col. Los Morales Polanco 11530 Mexico, D.F. Mexico Tel: +1 866 647 1649 Peak Scientific Japan K.K. 2-7-56, 2F Fuji Building 28 Kita Aoyama, Minato-Ku Tokyo, Japan 107-0061 Tel: +81 3-6864-0468 Peak Scientific Singapore 3 Science Park Drive #03-14 The Franklin Science Park Drive 1 Singapore, 118223 Tel: +65-6777 1966 Fax: +65-6777 1978

 Open the catalog to page 12
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.