Overview: The document describes a syringe sampling system equipped with the LiQuilaz® II Particle Counter, ideal for precise, small-volume sampling applications. It highlights the system's cost-effectiveness, ease of use, and broad application range.
Advantages:- Cost-effective due to minimal sample volume requirements, especially for expensive products.
- Operated via Windows-based software, reducing operator workload.
- Accommodates a wide range of measurement samples.
Ease of Use:- Allows recipe creation for each sample, ensuring consistent measurement conditions across multiple operators.
- User-level settings and security password options available.
- Supports up to 15 particle size channels.
Applications:- Chemical quality control.
- Component cleanliness testing.
- Ultrapure water and chemical testing in laboratories.
Specifications:- Product Models: SLS-2000, SLS-1200, SLS-1300, SLS-1500.
- Sensor Types: LiQuilazⅡE15P, LiQuilazⅡE20P, LiQuilazⅡS02, LiQuilazⅡS03, LiQuilazⅡS05.
- Particle Size Range: 0.2 – 125 µm depending on the model.
- Flow Rate: 20 ml/min with 100% measurement efficiency.
- Maximum Particle Concentration: 10,000 particles/ml.
- Sample Temperature Range: 10 – 50 ºC.
- Materials: Fused Silica, Black Quartz, Teflon, Kel-F®, Kalrez®.
- Calibration: Traceable to NIST and JIS standards.
- Software: SamplerSight.
- Syringe Sizes: 1, 5, 10 (standard), 25 ml.
- Sampling Mode: Suction.
- Minimum Sample Volume: 0.4 ml at 20 ml/min.
- Power Supply: 100 – 240 V, 50 – 60Hz, 1.6A.
- Dimensions: 38.1 x 22.0 x 43.4 cm; Weight: 13.6 kg.
- Operating Environment: Temperature 10 – 30 ºC, non-condensing humidity.
Additional Information: The document includes contact details for Spectris Co., Ltd., and notes on trademarks and rights reserved by Particle Measuring Systems, Inc.