Overview: The Lasair III particle counter is designed to set a new standard in aerosol particle counting, ensuring compliance with ISO 21501-4 through its on-board Pulse Height Analyzer (PHA). It features a user-friendly interface with a large IR touch screen and a chemically resistant, lightweight enclosure for portability and ease of cleaning.
Benefits:
  • ISO 21501 compliance with built-in PHA.
  • Available flow rates: 100 LPM, 50 LPM, or 1 CFM.
  • Large IR touch screen for ease of use.
  • Quiet pump for improved work environment.
  • Capability to generate reports for ISO 14644-1, EC GMP Annex #1, and FS209E.
  • Comprehensive validation manual for pharmaceutical regulatory compliance.
  • Operable in cleanrooms via web browser for setup, sampling, display, printing, and data download.
  • Supports nine languages for display and printout.
  • Removable batteries with optional external charger for continuous mobile use.
  • Stores sampling recipes to minimize operator errors.
  • Long-term data archiving with DataAnalyst software.
Applications:
  • Cleanroom monitoring.
  • Facility certification for ISO, EC GMP, or FS209E.
  • Trend analysis and statistical process control.
  • Troubleshooting particle excursions.
  • Compatible with manifolds for portable or dedicated use.
Specifications:
  • Channels: 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm.
  • Flow rate: 1.0 CFM (28.3 LPM) ± 5%, 50 LPM ± 5%, 100 LPM ± 5%.
  • Calibration meets ISO 21501-4 requirements.
  • Maximum concentration: > 1,200,000/ft³ for 310C, > 700,000/ft³ for 350L, > 575,000/ft³ for 5100.
  • Counting efficiency: 50% ± 20% for most sensitive threshold.
  • Zero count: 7.07 counts/m³ (1 CFM), 4.00 counts/m³ (50 LPM), 2.00 counts/m³ (100 LPM).
  • Data storage: 3000 complete data sets, 21 CFR Part 11 compliance.
  • Communication: Ethernet, RS-232, USB, optional wireless Ethernet.
  • Remote operation via web browser.
  • Environmental sensors: Four 4-20 mA inputs.
  • Display: 8.4” color VGA with IR touch screen.
  • Power: 100-240 V, 50-60 Hz, 150 W; Lithium battery operation for 3-6 hours.
  • Dimensions: 11.9 x 12.9 x 10.2 inches; Weight: 13.2 lbs without battery, 16.5 lbs with two batteries.
Contact Information: Particle Measuring Systems provides global support with contact details for various regions including Europe, Japan, Asia Pacific, Singapore, China, Mexico, and Puerto Rico.
See more

Catalog excerpts

Lasair III Particle Counter-1

ISO 21501 compliant with built-in Pulse Height Analyzer (PHA) > 100 LPM, 50 LPM or 1 CFM flow rates available і 100 LPM flow samples 1m > 3 in 10 minutes > Large IR touch screen for simplified use > Quiet pump improves work environment > Calculates and creates reports for ISO 14644-1, EC GMP Annex #1, and FS209E > Comprehensive validation manual makes it easy to meet pharmaceutical regulatory requirements > Operates in cleanrooms via a web browser: Setup, sample, display, print, download data > Choose from nine languages for display and printout > Removable batteries with optional external charger for continuous mobile use > Stores sampling recipes to reduce operator errors > Long term data archiving with DataAnalyst software The Lasair III particle counter sets the new standard for aerosol particle counters. The on-board Pulse Height Analyzer (PHA) ensures ISO 21501-4 compliance. The large IR touch screen controls and intuitive user interface make the system very easy to use. The particle counterђs rugged enclosure is chemically resistant and lightweight for portability. The enclosure also has a streamlined design that minimizes particle traps and makes the unit easy to clean. APPLICATIONS > Statistical process control > Troubleshooting particle excursions > Portable or dedicated use Without measurement, there is no control. >

 Open the catalog to page 1
Lasair III Particle Counter-2

310C 350L 5100 Channels 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 > ε m 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 > m 0.5, 1.0, 2.0, 5.0, 10.0, 25.0 > յ mFlow rate 1.0 CFM (28.3 LPM) 5% 50 LPM ѱ 5% 100 LPM 5% Calibration Meets ISO 21501-4 requirements Maximum concentration > a > 1,200,000/ft. > 3 > 700,000/ft. > 3 > 575,000/ft. > 3 Counting efficiency 50% ѱ 20% for most sensitive threshold. Meets ISO 21501-4 requirements. Zero count 7.07 counts/m > 3 (1 CFM); 4.00 counts/m > 3 (50 LPM); 2.00 counts/m > 3 (100 LPM)Data storage 3000 complete data sets. 21 CFR Part 11 compliance. Long term data storage/analysis with...