1. Catalogs
  2. Particle Measuring Systems
  3. HPD III High Pressure Diffuser

HPD III High Pressure Diffuser

HPD III High Pressure Diffuser

HPD III High Pressure Diffuser

Product catalog summary
Benefits
  • Measures gases from 5 to 100 psi, depending on the model.
  • No need for external flow meters.
  • Low contamination, maintaining less than 5% of ISO Class 5 cleanliness.
  • Cost-effective monitoring of pressurized gases using atmospheric pressure particle counters.
  • Compatible with Clean Dry Air (CDA), argon, nitrogen, carbon dioxide, and other non-toxic inert gases.
  • Available in seven versions for different flow rates: 0.1 CFM, 1.0 CFM, 50 LPM, and 100 LPM.
Features
  • No moving parts, reducing wear and tear.
  • Connects via a male 4-VCR fitting.
  • Includes an exhaust filter to prevent room contamination.
  • Supplied with mounting clamps.
  • Constructed from stainless steel for durability.
Applications
  • Compatible with Nano-ID® NPC10 nanoparticle counter.
  • Works with Lasair® III particle counter models 310B, 310C, 350C, and 5100.
  • Suitable for Airnet® II particle sensor models 301, 501, 510, and 510XR.
  • Can be used with IsoAir® 310P particle counter.
Specifications
  • Pressure: High pressure (20 – 100 psi for 0.1 CFM, 1.0 CFM, 50 LPM; 60 – 100 psi for 100 LPM) and low pressure (5 – 25 psi for 1.0 CFM and 50 LPM; 20 – 60 psi for 100 LPM).
  • Flow rate options: 0.1 CFM, 1.0 CFM, 50 LPM, or 100 LPM.
  • Particle size range: 0.3 – 25 µm.
  • Temperature range: 39 – 86 ºF (4 – 30 ºC).
  • Humidity: 0 – 85% RH non-condensing.
  • Material: 316L stainless steel body and exhaust filter; Buna-N O-rings.
  • Sample gas: Dry, inert, non-toxic, non-flammable gases.
  • Inlet fitting: Male 4-VCR fitting, class 316L stainless steel with Ruby orifice.
  • Exhaust fitting: Barb fitting for various ID sizes.
  • ISO class: Less than 5% of ISO Class 5.
  • Tubing length: Maximum 1 m (39.4 in).
  • Dimensions: 15 x 2.75 x 5.5 in (38 x 7 x 14 cm).
  • Weight: 2.7 lb (1.25 kg).
Additional Information
  • System components include mounting brackets and exhaust tubing to particle counter.
  • Not included: particle counter, ISPs sample tubing, communications cables, PC, FacilityNet or Pharmaceutical Net software.
Contact Information

Headquarters: Boulder, Colorado, USA. Global offices in Austria, Benelux, Brazil, China, France, Germany, Italy, Japan, Korea, Mexico, Nordic, Puerto Rico, Singapore, Switzerland, Taiwan.

See more

Catalog excerpts

HPD III High Pressure Diffuser-1

HPD III High Pressure Diffuser Aerosol Monitoring Accessory BENEFITS • Measures gases from 5 to 100 psi (range dependent on model) • No external flow meters needed • Low contamination, < 5% of ISO Class 5 cleanliness • Economical monitoring of pressurized gases with atmospheric pressure particle counters • Can be used with Clean Dry Air (CDA), argon, nitrogen, carbon dioxide or other non-toxic inert gases • Seven different versions for 0.1 CFM, 1.0 CFM, 50 LPM, and 100 LPM flow rates F E AT U R E S High Pressure Diffusers (HPDs) have been developed to address the challenge of sampling pressures of 5 – 100 psi, and the possible damage that can result from connecting pressurized air or gas to a particle counter. HPDs reduce the pressure of the gas to near atmospheric pressure by allowing some of the gas to diffuse out of the system, allowing the particle counter’s flow-metering system to function as designed. The HPD III Aerosol Monitoring Accessory is compatible with nontoxic, non-flammable gases such as Clean Dry Air (CDA), argon, nitrogen and carbon dioxide. The higher pressure gases are reduced to a level acceptable to particle counters designed to operate at atmospheric pressure. • No moving parts to wear • Connects via male 4-VCR fitting • Exhaust filter prevents possible room contamination • Mounting clamps supplied • Stainless steel construction A P P L I C AT I O N S • Nano-ID® NPC10 nanoparticle counter • Lasair® III particle counter models 310B, 310C, 350C, and 5100 • Airnet® II particle sensor models 301, 501, 510, and 510XR • IsoAir® 310P particle counter

 Open the catalog to page 1
HPD III High Pressure Diffuser-2

HPD III High Pressure Diffuser Aerosol Monitoring Accessory High Pressure 20 – 100 psi for 0.1 CFM, 1.0 CFM, 50 LPM flow rates 60 – 100 psi for 100 LPM flow rate Low Pressure 5 – 25 psi for 1.0 CFM and 50 LPM flow rates 20 – 60 psi for 100 LPM flow rate Flow rate Particle size range Temperature range Enclosure: 316L stainless steel body, exhaust filter Gaskets: 316 stainless steel, Buna-N O-rings Dry, inert, non-toxic, non-flammable gases (CDA, nitrogen, argon, carbon dioxide) NOTE : For carbon dioxide sampling, contact Particle Measuring Systems. Inlet fitting Male 4-VCR fitting, #4 size, class...

 Open the catalog to page 2

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.