1. Catalogs
  2. Paratronix Instruments Co., Ltd.
  3. OTR-D3 Oxygen Transmission Rate Tester

OTR-D3 Oxygen Transmission Rate Tester

OTR-D3 Oxygen Transmission Rate Tester

OTR-D3 Oxygen Transmission Rate Tester

Product catalog summary
Introduction
The OTR-D3 Oxygen Transmission Rate Tester is designed to measure the oxygen permeability of medium and high barrier materials such as plastic films and packaging materials used across various industries including food, pharmaceuticals, and electronics. It utilizes the coulometric sensor method for precise testing.

Features
  • Equipped with a high precision infrared sensor for accurate results.
  • Automatic control of humidity and temperature ensures a stable testing environment.
  • Operated via computer software with a user-friendly interface.
  • Real-time display of test data and curves.
  • Three independent test chambers allow simultaneous testing of three samples.
  • One-button automatic test operation.
  • Patented pipeline flushing design for clean carrier gas and efficient testing.
  • Unique leak-proof seal technology enhances chamber performance.
  • Complies with USP 1207 standards for data traceability and rights management, suitable for pharmaceutical industry needs (optional).

Technical Specification
  • Model: OTR-D3
  • Measuring Range (Film): 0.01~6500(cc/㎡.24h)
  • Resolution: 0.001
  • Permeation Area: 50 cm2 (customizable)
  • Sample Thickness: <3 mm (accessories needed for thicker samples)
  • Sample Quantity: 3
  • Temperature Control Range: 15℃~55℃ (optional)
  • Temperature Control Accuracy: ±0.1℃
  • Carrier Gas: 99.999% high purity nitrogen (user-provided)
  • Carrier Gas Flow: 0~100 mL/min
  • Air Source Pressure: ≥0.2MPa
  • Port Size: 1/8 inch metal tube
  • Instrument Dimension: 740mm (L)×415 mm (W)×430mm (H)
  • Power Supply: AC 220V 50Hz
  • Net Weight: 50Kg

Testing Principle
The specimen is clamped in the test cell, with oxygen or air on one side and high-purity nitrogen on the other. Oxygen permeates through the specimen into the nitrogen side, where it is analyzed by the coulometric sensor to calculate the oxygen transmission rate.

Test Standards
Complies with ASTM D3985, ASTM F2622, ASTM F1307, ASTM F1927, ISO 15105-2, JIS K7126-B.

Configuration
  • Standard: Main instrument, sampler, nitrogen/oxygen cylinder pressure reducing valve, Dow Corning vacuum grease.
  • Optional: Package test accessories, package temperature control device, constant temperature control device.
  • Note: Air sources (nitrogen and oxygen) must be provided by the user.
See more

Catalog excerpts

OTR-D3 Oxygen Transmission Rate Tester-1

OTR-D3 Oxygen Transmission Rate Tester (OTR Tester) is manufactured based on the coulometric sensor method and is applicable to test the oxygen permeability of medium and high barrier materials such as plastic films, composite films, sheets, papers, and other packaging materials used in food, pharmaceutical, medical device, consumer products, and electronic industries, etc. • High precision infrared sensor is used to ensure the testing accuracy • Automatic humidity and temperature control to guarantee the stability of the testing environment • All test operations are performed on a computer with testing software. User-friendly interface for easy and simplified operation. • Test data and curves are displayed in the real time • 3 independent test chambers for 3 test samples • Automatic test operation by one-button clicking • Patented pipeline flushing design ensures cleanliness of carrier gas and high testing efficiency • Unique sample leak-proof seal installation technology enhances the seal performance of test chambers • Meeting the requirements of USP 1207 United States Pharmacopoeia Standard for data traceability, rights management and other functions, the computerized system fully meets the needs of the pharmaceutical industry (optional)

 Open the catalog to page 1
OTR-D3 Oxygen Transmission Rate Tester-2

Technical Specification The pre-conditioned specimen is clamped in the test cell, oxygen or air flows on one side of the specimen while a stream of high-purity nitrogen flows on the other side. Oxygen molecules permeate through the specimen into the nitrogen side and are carried to the coulometric sensor. The sensor analyzes the oxygen concentration and calculates the oxygen transmission rate and other factors.. ASTM D3985, ASTM F2622, ASTM F1307, ASTM F1927, ISO 15105-2, JIS K7126-B Standard Configuration: main instrument, sampler, nitrogen/oxygen cylinder pressure reducing valve, Dow Corning...

 Open the catalog to page 2

All Paratronix Instruments Co., Ltd. catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About DirectIndustry
DirectIndustry is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. DirectIndustry is an entity of VirtualExpo Group.