1. Catalogs
  2. Paratronix Instruments Co., Ltd.
  3. MLT-V100 Non-destructive Micro-Leakage Tester

MLT-V100 Non-destructive Micro-Leakage Tester

MLT-V100 Non-destructive Micro-Leakage Tester

MLT-V100 Non-destructive Micro-Leakage Tester

Product catalog summary
Overview: The MLT-V100 Non-destructive Micro-Leakage Tester, also known as the Container Closure Integrity Tester (CCIT), is designed for professional testing of micro-leakage in various containers using vacuum decay and pressure decay methods. It is suitable for ampoules, vials, pre-filled syringes, bottles, non-porous pouches, and medical devices.
Product Features:
  • Equipped with a built-in 10-inch touch-screen computer, with an option for an external computer.
  • Completes single sample testing in 15 seconds.
  • Detects micro-leakage apertures of ≥1μm.
  • Separate arrangement of test chamber and main instrument, accommodating over 5 sample types.
  • Test chamber made of aluminum alloy or stainless steel with a pneumatic clamping system.
  • Includes a built-in flow meter for easy calibration.
  • Offers three calibration methods: zero point, leakage aperture, and flow.
  • Provides three judgment modes for test results: pressure decay, leakage aperture, and leakage flow.
  • Leakage flow error is ≤ 0.1ccm, with vacuum resolution ≤1pa/0.01mbar/0.0001psi.
  • Supports multiple units: kpa/mbar/pa/psi.
  • Suitable for both flexible and rigid samples, including vials, infusion bags, and electronic accessories.
Test Standards: Conforms to ASTM F2338, the standard test method for nondestructive detection of leaks in packages by vacuum decay method.
Technical Specifications:
  • Test Method: Vacuum Decay, with options for Pressure Decay.
  • Measurement & Control: Micro-computer control with dual sensors.
  • Vacuum System: Vacuum pump.
  • Air Source: External air source required for some models.
  • Units: kPa/mbar/psi/Pa.
  • Pressure Resolution: 1 Pa or 0.1 Pa.
  • Test Range: -99 to 0 kPa or -99 to 400 kPa.
  • Measuring Range: ≥0.01ccm (about 1μm).
  • Test Accuracy: ±0.2μm (≥3μm) and ±0.3μm (<3μm).
  • Main Control Chip: STM Microcomputer Chip.
  • Acquisition Frequency: 50 to 200 Hz.
  • Operation System: 7” or 10.1” LCD touch screen with computer software.
  • Data Storage: 600 groups or unlimited.
  • Data Communication: USB/RS232/485.
  • Body Material: Engineering plastics or stainless steel.
  • Instrument Dimensions: Varies slightly by model.
  • Instrument Weight: 20 kg.
See more

Catalog excerpts

MLT-V100 Non-destructive Micro-Leakage Tester-1

MLT-V100 Non-destructive Micro-Leakage Tester aka. Container Closure Integrity Tester (CCIT), employs non-destructive testing technology to meet the requirements of both vacuum decay and pressure decay testing methods. It is a highly versatile instrument, used for the professional testing of micro-leakage in a wide range of containers, including ampoules, vials, pre-filled syringes, bottles, non-porous pouches, blow fill seal strips medical devices. Product Features • Built -in 10-inch touch-screen computer (External computer is optional) • Single sample testing process can be completed within 15 seconds • Micro-Leakage aperture (> 1pm) can be detected • The test chamber and the main instrument are arranged separately, and the test chamber is suitable for more than 5 kinds of samples • The test chamber is made of aluminum alloy or stainless steel, with pneumatic clamping system • Built-in flow meter, easy to calibrate • Three calibration methods for zero point, leakage aperture and flow • Three judgment modes for test results: pressure decay, leakage aperture and leakage flow • Leakage flow error < 0.1ccm, Vacuum resolution < 1pa/0.01mbar/0.0001psi, Multiple units are available kpa/mbar/pa/psi • Suitable for various flexible and rigid samples including vials, infusion bags, packages for contact lenses, milk powder cans and electronic accessories, etc. with vacuum decay or pressure decay method Paratronix Instruments Co., Ltd. Website:www.paratronixen.com The manufacturer reserves the right to improve product performance and parameters

 Open the catalog to page 1
MLT-V100 Non-destructive Micro-Leakage Tester-2

This instruments conforms to multiple international standards ASTM F2338: Standard Test Method for Nondestructive Detection of Leaks in Packages by Vacuum Decay Method. Test Method Measurement & Control Vacuum System Air Source Unit Pressure Resolution Test Range Measuring Range Test Accuracy Test Accuracy Main Control Chip Acquisition Frequency Operation System Data Storage Data Communication Body Material Instrument Dimension Instrument Weight MLT-V100 MLT-V100 (T) MLT-V100 (P) Vacuum Decay Vacuum Decay/Pressure Decay Micro-computer Control/Dual Sensors Vacuum Pump 7”LCD touch screen 10.1 ”...

 Open the catalog to page 2

All Paratronix Instruments Co., Ltd. catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About DirectIndustry
DirectIndustry is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. DirectIndustry is an entity of VirtualExpo Group.