Chilled mirror hygrometer Chilled mirror hygrometers are used in standards and metrology labs as well as in industrial applications where precise and repeatable humidity measurement and control are required. The inherent accuracy and long term stability provide many advantages over other types of humidity measurement technologies.
Open the catalog to page 1OptiSonde™ Datasheet I PANAMETRICS The OptiSonde chilled mirror hygrometer is a National Institute of Standards and Technology (NIST) traceable humidity transfer standard at an exceptional value. The compact rugged analyzer is connected to either a single-or two-stage chilled mirror sensor and PRTD. Chilled mirrors are the humidity sensors of choice when laboratory and process measurements require high precision without long-term drift. An LCD indicates two simultaneous user-configured parameters. OptiSonde provides two user-configured analog (4 to 20 mA) outputs. The display and front panel...
Open the catalog to page 2OptiSonde™ Datasheet Chilled mirrors require a nominal flow rate across the mirror to achieve the optimal dew/frost formation and response time. The sensor should either be installed in a duct with moving air or be equipped with a sampling pump. The optimum flow rate is 0.5 to 2.5 SCFH (0.25 to 1.2 L/min). Panametrics provides sampling systems to condition the temperature, regulate the pressure and filter process air before reaching the chilled mirror. Our applications engineers would be pleased to discuss your application in detail and provide a recommendation for the system best suited to your...
Open the catalog to page 3OptiSonde™ Datasheet OptiSonde specifications Power • 100 - 240 VAC, 50 to 60 Hz, 60 watts • 18-32 VDC (bench top only) Enclosure material • Wall mount: Cast aluminum • Bench top: Aluminum body with plastic front bezel • Wall mount: 26.6 x 20.8 x 11.4 cm (h x w x d) • Bench top: 9.4 x 20.3 x 22.4 cm (h x w x d) • Wall mount: 2.4 kg • Bench mount: 1.4 kg Ingress • Wall mount: IP-65 protection rating • Bench top: IP-20 Front panel keys Rubber function keys—may be locked out via software Dependent on sensor used Temperature accuracy Other humidity parameters The range and accuracy of absolute humidity...
Open the catalog to page 4OptiSonde™ Datasheet 1111H single stage chilled mirror specifications Dew/frost point accuracy Typical dew point measurement range -15ºC to 25ºC Td at 25ºC, 1 atmosphere for standard air.(Equivalent to 6-100% RH) Sample flow rate Operating temperature Operating pressure Derived from analyzer Sensor body Epoxy-coated aluminum PTFE filter optional (standard on 1111H) Rhodium-plated copper (solid platinum optional) Vapor barrier Sensor wetted material Epoxy-coated aluminum, rhodium copper, mylar Electrical connector MS style multipin keyed connector • MB-11: Wall mounting bracket (1111H) • PTFE:...
Open the catalog to page 5OptiSonde™ Datasheet D2 two-stage chilled mirror specifications Sensing element Dew/frost point ± 0.2ºC accuracy 65ºC at 25ºC, 1 atmosphere for standard air Typical dew point measurement range -35ºC to 25ºC Td at 25ºC, 1 atmosphere for standard air.(Equivalent to 1 to 100 % RH) Sample flow rate Operating temperature Operating pressure Derived from analyzer View port Equipped with illuminated view port that enables observation of mirror Sensor body Cast aluminum with 314 SS flow cell Rhodium-plated copper (solid platinum optional) Vapor barrier Stainless steel Sensor wetted material 302, 314,...
Open the catalog to page 6OptiSonde™ Datasheet 1211H two stage chilled mirror specifications Sensing element Dew/frost point ± 0.2ºC accuracy -35ºC to 25ºC Td at 25ºC, 1 atmosphere for standard air. (Equivalent to 6 to 100% RH) Sample flow rate Operating temperature Operating pressure Derived from analyzer Sensor body Cast aluminum with 314 SS flow cell Rhodium-plated copper (solid platinum optional) Vapor barrier Sensor wetted material 302, 314, 316 stainless steel, silicone O-ring, rhodium plated copper or platinum mirror Electrical connector MS style multipin keyed connector Platinum mirror Stainless steel vapor barrier...
Open the catalog to page 7OptiSonde™ Datasheet I PANAMETRICS Sensing Four-wire 1/3 Class A DIN 43760 PRT, 100 W Dew/frost point Standard: ±0.2°C Hysteresis Negligible Cooling stages Two-stage TEC module Depression 85°C at 75°C body temperature and atmospheric pressure Electrical MS-style multipin connector connector IEC receptacle for AC power -10°C to 75°C dew/frost point (Td) in air @ 75°C body temperature in 25°C ambient temperature and atmospheric pressure. Equivalent to 0.7% to 100% RH. Other parameters based on calculations. Sample flow 0.5 to 5.0 SCFH (0.25 to 2.5 L/min) Heater control Thermostatically controlled....
Open the catalog to page 8OptiSonde™ Datasheet I PANAMETRICS Sensing Four-wire 1/3 Class A DIN 43760 PRT, 100 W Dew/frost point Standard: ±0.2°C Hysteresis Negligible Cooling stages Four-stage TEC module Auxiliary Liquid cooling jacket Depression, air 95°C at 25°C and atmospheric pressure Inlet/outlet 1/4 in (6 mm) OD tubing compression fittings Electrical MS-style multipin connector mates connector IEC receptacle for AC power S Stainless steel vapor barrier Depression, 105°C with 15°C cooling water liquid cooled Typical • -65°C to 25°C dew/frost point (Td) in air measurement @ 25°C and atmospheric pressure. range • -75°C...
Open the catalog to page 9OptiSonde™ Datasheet I PANAMETRICS Sensing Four-wire 1/3 Class A DIN 43760 PRT, Response time 7 seconds for 25°C to 70°C step change in fluid Sensor body Stainless steel Cable PTFE insulation rated to 150°C, 3 m (10 ft) length, spring strain relief Power Low voltage derived from analyzer PTFE: FILTER for 1111H Sensor • Hard shell carrying case for benchtop unit with accessories. • Extended warranty
Open the catalog to page 10Ordering information Analyzer 1 Benchtop 2 Wall mount Power 1 100-240 VAC, 50-60 Hz 2 18-32 VDC Chilled mirror sensor 0 None 1 1111H single stage sensor, 1” M-NPT fitting 2 1111H single stage sensor. Flange for duct mounting and PTFE filter 3 D2 two stage sensor 4 1211H two stage high temperature and pressure sensor 5 SIM-12H two stage heated sensor. 115 VAC 6 SIM-12H two stage heated sensor. 230 VAC 7 HSS-12H two stage heated sensor and sampling system. 115 VAC 8 HSS-12H two stage heated sensor and sampling system. 230 VAC 9 1311-DR four stage chilled mirror sensor. 100 VAC A 1311-DR four stage...