PAL Systems Specifications

PAL Systems Specifications

PAL Systems Specifications

Product catalog summary
Overview: This document outlines the specifications and test procedures for various PAL Systems used in gas and liquid chromatography. It includes detailed descriptions of test conditions, equipment specifications, and performance criteria.
1. Gravimetric Test:
  • Focuses on repeatability with a relative standard deviation (s-rel) of 0.30%.
  • Carry-over is not applicable.
  • Test conditions involve iso-Octane in 2 mL vials, using a Hamilton syringe.
2. Gas Chromatographic Tests:
  • High Split Ratio: Uses iso-Octane with a retention gap. Repeatability s-rel is ≤ 0.80%.
  • Split Mode: Utilizes an alkene test mix with a repeatability s-rel of ≤ 1.00%.
  • Headspace Test: Involves iso-Octane with a high split ratio, achieving a repeatability s-rel of < 1.00%.
3. HPLC Chromatographic Test:
  • Tests include loop overfill and partial loop filling with a repeatability s-rel of 0.10% and 0.20% respectively.
  • Linearity is tested across 5 levels with a correlation coefficient R > 0.9999.
4. General Remarks:
  • Repeatability tests involve multiple specimens, with specific conditions for each test.
  • Temperature and barometric pressure are not considered in the tests.
  • Data acquisition is performed using Agilent ChemStation Software.
5. Equipment and Parameters:
  • GC and HPLC systems from Agilent are used, with specific settings for injectors, detectors, and columns.
  • Carrier gases and solvents are specified for each test type.
6. Formulas:
  • Includes formulas for calculating mean values, standard deviation, and relative standard deviation.
See more

Catalog excerpts

PAL Systems Specifications-1

SYSTEM Prep and Load Platform Specifications for PAL Systems COMBI | GC | LC HTC | HTS | LHX | HTX Not Valid for PAL-xt Systems PAL is a registered trademark of CTC Analytics AG | Switzerland

 Open the catalog to page 1
PAL Systems Specifications-2

Gas Chromatographic Test: High Split Ratio | Retention Gap | Solvent iso-Octane 4 Gas Chromatographic Test: Split Mode | Analytical Column | Alkene Test Mix 5 Gas Chromatographic Headspace Test: High Split Ratio | Retention Gap | Solvent iso-Octane 6 HPLC Chromatographic Test: Loop Overfill and Partial Loop Filling 7 Formulas used to calculate the Specification Values 8

 Open the catalog to page 2
PAL Systems Specifications-3

Syringe Volume Specification Test Conditions 10 pi Repeatability 15 times 8 pi Water transferred into Relative Standard Deviation target vial. s-rel :_0.20 % 100 pL Repeatability 100 pL Water transferred in a target vial. Relative Standard Deviation s-rel :_0.15 % Linearity 6 levels and 3 replicates of each level. R: >0.9999 1000 pL Repeatability 500 pL Water transferred in a target vial. Relative Standard Deviation s-rel : 0.10 % Linearity 6 levels and 3 replicates of each level. R: > 0.9999 2500 pL Repeatability 1000 pL Water transferred in a target Relative...

 Open the catalog to page 3
PAL Systems Specifications-4

Specification: Repeatability: Relative Standard Deviation s-rel: 0.30 % Carry-over: not applicable General Remarks: Number of specimens for Repeatability Tests: n = 20 Carry-over: Test: Remaining solvent in syringe can not be eliminated from injection to injection. Test Conditions: Test Sample: iso-Octane puriss., 2 mL vials filled to 50 % of nominal content Syringe: Hamilton Art. No.: 203205 | CTC Art. No.: SyrC L10-26S-AS PAL Parameter Settings: CYCLE GC-Inj Fill Strokes 5 SYRINGE10 |jlPullup Del5 s Sample Volume 1.0 p.L Inject to GC Inj1 Fill Speed 2 |jl/s Eject...

 Open the catalog to page 4
PAL Systems Specifications-5

Specification: Repeatability: Relative Standard Deviation s-rel: 0.80 % General Remarks: Number of specimens for Repeatability Tests: n = 8 Repeatability determination based on C14 Peak Carry-over Test: Peak < 1 area count, not detectable. Test Conditions: Test Sample: Alkane Test Mix C14,G5,G6 | 0.3 p.g dissolved in iso-Octane. 2 mL vials filled to 50 % of nominal content Syringe: Hamilton PNo.: 203205 | CTC Analytics PNo.: SyrC L10-26S-AS PAL Parameter Settings: CYCLE GC-Inj Fill Strokes 5 Sample Volume 1.0 |jl Inject to GC Inj1 Air Volume 0.5 |jl Inject Speed 50...

 Open the catalog to page 5
PAL Systems Specifications-6

Specification: Repeatability: Relative Standard Deviation s-rel: < 1.00 % General Remarks: Number of specimens for Repeatability Tests: n = 7 10 |jl are measured by using a glass capillary, Drummond Microcaps (Supplier Supelco, PNo. P-1924) Test Conditions: Test Sample: Iso-Octane, puriss., Microcaps Glass Capillary hold with Tweezers, filled and put into 20 mL Headspace vial, sealed immediately. Syringe: Hamilton PNo.: 203084 | CTC Analytics PNo.: SyrC HS 2.5 -23-5 GC Agilent 6890: Injector: split/splitless, Liner: Single Taper 4 mm ID, Needle Penetration 40 mm Silanized glass wool packing...

 Open the catalog to page 6
PAL Systems Specifications-7

Specification: Syringe VolumeSpecificationTest Conditions 100 pL Repeatability Loop Overfill Loop size 20 pi Relative Standard Deviation Loop fill volume: 80 pL 100 pL Repeatability Partial Loop Filling Loop size 20 pL Relative Standard Deviation Loop fill volume: 10 pL s-rel :_0.20 % 100 pL Linearity 5 levels and 3 replicates of each level. General Remarks: Number of specimens for Repeatability Tests: n = 9 Carry-over: Test: Peak < 1 area count, not detectable. Test Conditions: Test Sample: Benzophenone, Supplier Fluka PNo. 12750, 50 pg/mL dissolved in...

 Open the catalog to page 7
PAL Systems Specifications-8

Mean Value (average). Average of measured area counts. x = Mean Value of area counts. Xt = Peak area counts, i th observation of n n = Number of injections, calibration points. Standard Deviation O n - 1 = Standard Deviation Relative Standard Deviation x RSD = Relative Standard Deviation

 Open the catalog to page 8