Product catalog summary
Introduction
The document discusses the DLW (Dynamic Load and Wash) Option for the PAL system, designed to enhance LC/MS injection processes by minimizing carryover and improving reproducibility. This is particularly important given the sensitivity of modern mass spectrometers.

Specifications
The DLW Option features stainless steel and PEEK wetted parts, a 100µl syringe size, and offers reproducibility better than 1% RSD under specific conditions. Typical carryover is less than 0.003% (30ppm), with a clean cycle time of less than one minute.

Procedures
The DLW system operates by aspirating the sample into a holding loop, with the syringe acting solely as an aspirator and dispenser. After injection, the sample path is washed with up to two solvents using active micro pumps and a solenoid valve to ensure thorough cleaning and near-zero carryover.

Performance Tests
Carryover tests using UV and MS detection methods demonstrate the DLW's effectiveness, with no detectable carryover in MS tests and less than 30ppm in UV tests. The tests involved specific conditions such as flow rates, injection volumes, and solvent compositions.

Ordering Details
The document outlines various DLW upgrade kits, including the DLW-2 Option, HTC-xt, and HTS-xt kits, each containing necessary components like the injection adapter, wash station, and solvent pumps.

General Specifications
The PAL system is an XYZ robot with an injection unit, featuring a local user interface and remote control capabilities. It supports various sample capacities and instrument options, and adheres to electrical safety standards.

Conclusion
The DLW Option provides a clean and efficient solution for LC/MS sample handling, with integrated pumps for active wash solvent delivery and the ability to upgrade existing PAL systems.
See more

Catalog excerpts

PAL DLW-1

DLW Option Prep and Load Platform DLW – Fast and Clean LC/MS Loading High Throughput Screening Environmental, Food Safety, Forensics Preclinical Research, Metabolomics Drug Metabolism, Pharmacokinetics Protein Biomarker Discovery

 Open the catalog to page 1
PAL DLW-2

DLW Upgrade kit including Syringe set, Pump and Wash Station Near Zero Carryover – Fast Cycles and exceptional reproducibility Wetted parts material: Stainless steel, PEEK Syringe Size: Today’s Mass Spectrometers are so sensitive that carryover becomes an issue again. To meet today’s requirements for fast, clean and reproducible LC/MS injection; CTC developed the DLW Option for the PAL system (Dynamic Load and Wash). Reproducibility (partial loop): better than 1 % RSD (under specific conditions) Carry over: Typically less than 0.003% (30ppm) Typical Clean Cycle Time: less than 1 Minute (Fast...

 Open the catalog to page 2
PAL DLW-3

These tests show the excellent performance of the DLW in both UV and MS mode. Carryover Test UV: Less than 30ppm carryover (Chlorhexidine) Chlorhexidin O.emp'ml Conditions for UV carryover test Injection: Full Loop 2^1 (PEEK) Injection of 1200 ng Chlorhexidine on column; followed by 2 blank injections using the new DLW option. Less than 0.003% (30ppm) of carryover could be detected. Carryover Test MS: No Carryover detected (Dextromethorphan) Carryover: Undetectable Blank after sample injection Conditions for MS carryover test Injection: Full Loop 2\i\ Eluent Isocratic: Injection of 272 .2 pg...

 Open the catalog to page 3
PAL DLW-4

Wash Station Syringe Solenoid / Actuator Valve Pump & Solvent 2 Holding Loop Spring loaded needle guide Holding Loop needle

 Open the catalog to page 4
PAL DLW-5

Dynamic load and Wash Cycle principle minutes Old Syringe DLW Standard Fast DLW Download pdf Standard Cycle Fast Cycle Stator Wash Cycle www.palsystem.com 4 minutes 67 seconds 51 seconds Pump 2 Solenoid / Actuator Valve Holding Loop Rear air gap Wash Station Sample Front volume Sample Loop Column Waste DLW Standard Cycle cycle start DLW Fast Cycle cycle start 1. aspirate rear air segment 1. aspirate rear air segment 2. get sample aspirate rear, inject and front volume. 2. get sample aspirate rear, inject and front volume. 3. aspirate front air segment 3. aspirate front air segment 4. passive...

 Open the catalog to page 5
PAL DLW-6

SYSTEM LC/LC-MS General Specifications Ordering details System Type XYZ robot with injection unit PAL DLW-2 Option Dynamic Load an Wash Option for fast and clean LC Cycles 1pc DLW-2 injection adapter with sample holding loop and syringe 100µl 1pc wash station 1pc wash solvent pump station with 2 active micro pumps 2 pc 1 liter wash solvent bottles 2 pc solvent bottle transfer line including solvent filter 1 pc system CD ROM DLW for xt-Systems 1 pc manual DLW Option requires but does not include xt-Upgrade for PAL and Firmware 4.1.X or higher Choose this kit if your PAL is already equipped with...

 Open the catalog to page 6
PAL DLW-7

Clean and efficient removal of carryover in the entire flow path Holding loop for effective rinsing of the complete sample path Integrated pumps for active wash solvent delivery selectable wash time for organic and aqueous solvents Spring loaded syringe needle positioning in needle guide – no dead volume Existing PAL Systems can be upgraded X-Type Syringe based SYSTEM www.palsystem.com