Ultrabright 90kV Microfocus-1

Technical Data Sheet 90kv Microfocus X-ray Source The UltraBright Microfocus System 96000 Series is a 90kV, 80W X-ray source designed for applications where high power, high magnication and small spot size are important. Operated by an external high voltage Smart Controller capable of providing variable voltage and power control, the UltraBright Microfocus System delivers exceptional magnication and image quality with full control of “Brightness” Maximum ux output is maintained . through automatic matching of a given power setting to a corresponding optimal spot size. • Exceptional magnication and image quality • High power operation — ideal for high ux applications and experiments • Integrated package eliminates HV cable for improved reliability • Complete range of user control — ideal for research applications • Compact, lightweight design — ideal for portable applications Microtomography Microdiffraction Microuorescence CT imaging for life sciences and industrial inspection Product Ordering Table Target Material (Part#) Voltage Power Density Specifications Operating voltage range: Maximum power: Maximum beam current: Focal spot size: Focus to Object Distance (FOD): Cone of illumination: Window material and thickness: Window diameter (unobstructed): Window conguration: Target material: Ambient operating temperature: Maximum operating temperature (anode): Cooling method: Shielding: Dimensions: Weight: See product ordering table See product ordering table 2.0mA 13-20µm @ maximum voltage and minimum power 4mm 50° x 74° (nominal) See chart p. 2 Be, 245µm 9.5mm (0.37”) End window See product ordering table 10oC to 40oC 70°C Forced air (150 CFM @ 4” recommended for continuous operation) Not shielded 392.4mmL x Ø114.30mm (15.5”L x Ø4.5”) ≤ 4kg (8.81lbs.) Storage conditions: -10°C to 55°C Barometric Pressure: 50-106kPa; Humidity: 10-90% (no condensation) Condensation on Be window will cause window corrosion, vacuum loss and X-ray tube failure X-Ray Technology www.oxford-instruments.com/xt 360 El Pueblo Way Scotts Valley, CA 95066 Phone: (831) 439-9729 Email: [email protected]

 Open the catalog to page 1
Ultrabright 90kV Microfocus-2

UltraBright 96000 Series 90kv Microfocus X-ray Source REFERENCE DRAWING 8236 Dimensions: [mm],Inches [114.30] Ø4.50 Oxford Label Warning Label Clamp Here Cone Angle [4.92] .19 Target to Cu Flat DETAIL A SCALE 1:1 Radiation Pattern as Seen Normal to Window Line Tangent to Target Emitted Cone and Spot Position Dimension Location of radiation cone center Window to spot distance

 Open the catalog to page 2
Ultrabright 90kV Microfocus-3

Technical Data Sheet CONTROLLER UNIT SPECIFICATIONS Baud rate: Flow control: Data bits: Stop bits: Anode voltage READ BACK: X-ray ON/OFF Stand-by, warm-up, output, fault modes in ASCII format Display panel information except remote/local mode will be in ASCII Power on can be accomplished remotely by X-ray on/ off command. However, if cathode emitter is turned off, power is restored only through front panel on/off switch. Safety interlocks available on rear panel. Control Operation rear panel ation, front panel on/off switch must ation, front panel on/off switch must Front panel switch Front panel...

 Open the catalog to page 3
Ultrabright 90kV Microfocus-4

UltraBright 96000 Series 90kvMicrofocus x-ray source Software Control Option XTG Ultrabright Microfocus Source HighVoltage | Brightness Focus) Output Monitor Morel Instnjxients X-Ray Technology lm Runs under LabVIEW RT & MS-Windows Works in conjunction with RS232 control interface Complete control of voltage, power, and focus Dynamic status display Dynamic fault display Open software architecture allows for modification to user interface with available additional development software Software Control Manual Operation Switch is located on rear panel For software operation, front panel On/Off switch...

 Open the catalog to page 4