GammaScreen-8™ 8-Channel Rapid Analysis Screening of Multiple Gamma Emitting Radionuclides in Clinical Samples ORTEC Gross Gamma Counter Sample Evaluation Counting System #1 Sample Date/Time: Friday, October 16 2009 11:13 Actual (CPS) Limit (CPS) Evaluation SampleID: 09-911-000065-1 GG Detector: 2 3.87 3.98 PASSED SampleID: 09-911-000066-1 GG Detector: 3 3.43 3.80 PASSED SampleID: 09-911-000067-1 GG Detector: 4 3.54 3.84 PASSED SampleID: 09-911-000068-1 GG Detector: 5 6.44 3.82 FAILED SampleID: 09-911-000069-1 GG Detector: 6 3.43 3.68 PASSED JL Easy to Use Rapid and Simple Gamma Radionuclide Screening • A very compact solution for rapid screening of hundreds of samples/day for possible radioactive contamination. • No sample processing: Pass/fail evaluations against analysis limits. • 700 samples may be counted in 20 hrs of operation. • MDA (LOD)1 = 52 Bq/L, 300 second count. • 8 samples (10 ml or 50 ml) counted simultaneously with automated data entry by bar code scanning (optional). • 8 completely independent NaI(Tl) well detectors and count chains. • Digitally stable electronics. • Full spectral data retained for detailed analysis. 'Minimum Detectable Activity or Limit of Detection.
Open the catalog to page 1The post 9-11 world differs drastically from previous nuclear age history. The possibility of a large accidental release of radioactivity affecting a civilian population is now added to by the possibility of an intentional exposure as a result of a terrorist attack involving a radioactive dispersal device or RDD. Following an exposure event or environmental release of radioactive material, large numbers of people may require or demand internal monitoring. Failure to respond quickly can increase the level of public anxiety. Even when the risk is slight, large numbers of subjects may be...
Open the catalog to page 2GammaScreen-8™ 8-Channel Rapid Analysis Screening of Multiple Gamma Emitting Radionuclides in Clinical Samples System Specifications Lead Shield, floor mounted Weight 1780 lbs (810 kg) Dimensions (HxWxL) 34.5 x 24 x 28 in (87.6 x 61.0 x 71.1 cm) Width with top open 38.54 in (97.9 cm) Minimum thickness between detectors 1 in (2.54 cm) NaI Detectors (8) Crystal size 3” diameter x 5” long Sample well 0.67” diameter x 3.9” deep Sample Tubes (not included but supplied at extra cost if required) Maximum dimensions 17 mm diameter x 120 mm length with nominal 15 ml volume. ORTEC digiBase...
Open the catalog to page 3Specifications subject to change 080317 www. ortec-online. com Tel. (865) 482-4411 • Fax (865) 483-0396 • ortec.info@ametek.com 801 South Illinois Ave., Oak Ridge, TN 37830 U.S.A. For International Office Locations, Visit Our Website
Open the catalog to page 4All ORTEC catalogs and technical brochures
-
RADEAGLET-R
12 Pages