Overview: The Nova Series Digital Laser Doppler Vibrometer is a high-performance device utilizing an invisible SWIR laser at 1550 nm, offering ten times the laser power of a HeNe system while maintaining eye safety. It is designed for maximum dynamic range and is ideal for surfaces that poorly reflect visible light, such as dark, rough surfaces, biological tissue, and for long-distance measurements.
Key Features:
  • Operates with a SWIR laser for high signal quality on challenging surfaces.
  • Suitable for high-frequency vibrations, biological surfaces, and long-distance measurements.
  • Maximum velocity measurement up to 25 m/s and frequency bandwidth up to 25 MHz.
Technical Specifications:
  • Measured quantities include velocity, displacement, and acceleration.
  • Digital signal processing with FPGA-based technology.
  • Variable working distances from 6 mm to over 300 m.
  • Laser safety class 1 for measurement laser and class 2 for target laser.
Model Options:
  • Various models available, each with specific capabilities such as high speed, high frequency, and all-rounder features.
  • Optional decoders for displacement and acceleration measurements.
Accessories and Software:
  • Includes transport case, optional telescope, IR-camera, and mobile battery.
  • OptoGUI software for data acquisition, analysis, and remote control.
Maintenance and Warranty:
  • Standard 12-month warranty with optional extensions.
  • Free software updates within the warranty period.
  • Optional recalibration and cleaning services.
Contact Information: Optomet GmbH, Pfungstaedter Strasse 92, 64297 Darmstadt, Germany. Phone: +49(0)6151-38432-0, Email: [email protected], Website: www.optomet.com
See more

Catalog excerpts

Digital Laser Doppler Vibrometer Nova-Series-1

Digital Laser Doppler Vibrometer Nova-Series SWIR Single-Point Laser Vibrometer Vibrometry in the short-wave infrared range for maximum signal-to-noise ratio. Delivers best signal quality for dark or rough surfaces, biological tissue, as well as long working distances. IDEAL FOR • Dark / rough surfaces • Very high-frequency vibrations • Biological surfaces • Moving or rotating objects • High speed vibrations up to 25 m/s • Long distance measurements HIGH-END AND BEST-PERFORMANCE The Nova Series Laser Vibrometer operates with an invisible SWIR laser (1550 nm). At a wavelength of 1,550 nm, ten times the laser power of a HeNe system is available for measurements without compromising the eye safety. The Nova Series comes in when maximum dynamic range is required or every decibel of signal level counts. Typical use cases are surfaces that poorly reflect visible light, for example dark and rough surfaces or biological tissue. A high signal strength is also of vital importance for measurements over long distances, e.g. in civil engineering or

 Open the catalog to page 1
Digital Laser Doppler Vibrometer Nova-Series-2

Technical Data Measured Quantities – Performance Parameters Max.2 Frequency The Resolution is defined as the signal amplitude (rms) corresponding to a signal-to-noise ratio (SNR) of 0dB with 1 Hz spectral resolution at 50 % fmax of smallest measurement range. 1a The Resolution is defined as the signal amplitude (rms) corresponding to a signal-to-noise ratio (SNR) of 0dB with 1 Hz spectral resolution. 2 Actual specifications depend on the configured decoder. 3 Optional available Measurement Specifications Measured quantities Frequency Max. Velocity, displacement, acceleration 0 Hz - 25 MHz 25 m/s Signal...

 Open the catalog to page 2
Digital Laser Doppler Vibrometer Nova-Series-3

General Device Specifications User interface output Color screen 3.5" + 20 segment LED bargraph User interface input Touch screen, knobs with push-button, key switch (power) Operating temperature Storage temperature Relative Humidity Length × width × height (excluding handle and lens): 380 × 180 × 148 mm Power supply Portable Operation Portable power supply - Produce various preset functions (sin, chirp, gaussian, ...) - Import of arbitrary functions and audio wave-files Every 12 month (recommended) Optional available Model Options - Decoders 5, 6 5For details see decoder data-sheets. 6 Variations...

 Open the catalog to page 3
Digital Laser Doppler Vibrometer Nova-Series-4

Characteristics of the optional Displacement Decoder7 Decoder Required Vel.Decoder Largest Range Frequency Bandwidth Best displacement resolution: 0.05 pm. Characteristics of the optional Acceleration Decoder7 Decoder Required Vel.Decoder Largest Range Frequency Bandwidth Best acceleration resolution: 1.8µ g /

 Open the catalog to page 4
Digital Laser Doppler Vibrometer Nova-Series-5

Set-up O-AF-XLR-N Ethernet Digital Output Objective Lenses Vibrometer Standard Objective lens Specification Short-Range Autofocus O-AF-SR-N Mid-Range Autofocus O-AF-MR-N Long-Range Autofocus O-AF-LR-N Super-Long Range Autofocus O-AF-SLR-N Objective lens for Extra-Long distance (O-AF-XLR-N) Stand-off distance 14 (m) 2.49 3.79 6.48 9.64 10.2 20.5 48.5 With each additional meter plus [µm] Fiber Objective lens Specification M-F-OBJ Micro Fiber Objective Min. stand-off distance (mm) Spot size in µm at: 15 mm 30 mm 35 mm 50 mm 90 mm 200 mm 860 mm F-OBJ Fiber Objective Measured from the front of the objective...

 Open the catalog to page 5
Digital Laser Doppler Vibrometer Nova-Series-6

Standard included, O Optional available Transport Case for Single Point Vibrometer Stable and waterproof Peli case for safe keeping and transport of vibrometer. External dimension (L x B x H): 61.9 x 49.2 x 22.3 cm Compact und light transport bag for outdoor measurements or transport as carry-on baggage in an plane. For measurements from a greater distance, a telescope makes it easier to find the measuring point. The telescope can be easily mounted on the gauge rail at the top of the vibrometer. For digital visualizing of the laser point. The camera can be easily mounted on the gauge rail at the...

 Open the catalog to page 6
Digital Laser Doppler Vibrometer Nova-Series-7

Software OptoGUI Analysis Software Software for data acquisition, analysis and remote control. Live animation of measured time and frequency data. OptoGUI software includes Remote control Remotely control all vibrometer settings via ethernet. Read out data via ethernet with up to 80 MS/s Time data Live animation of measured vel. /disp. /accel. data Export data Export time data as .csv, .h5, .wav or .mat files -Real-Time Fast Fourier Transformation -up to 536 Mio. FFT lines Peak identification Automatically identify signal peaks in the frequency spectrum Show a live Spectrogram of the FFTs of the...

 Open the catalog to page 7
Digital Laser Doppler Vibrometer Nova-Series-8

DIGITAL LASER DOPPLER VIBROMETER Optomet GmbH | Pfungstaedter Strasse 92 64297 Darmstadt | Germany Phone +49(0)6151-38432-0 | [email protected] www.optomet.com

 Open the catalog to page 8
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.