Product catalog summary
Overview: This document provides technical specifications for two high power thermal sensors, models L250W and L300W-LP1-50, designed for general purpose and high power density applications respectively.
Specifications:
  • Model L250W: Suitable for general purpose use with a broadband absorber type. It operates within a spectral range of 0.19 - 20 µm and has a power range of 1W - 250W. The maximum average power density is 10 kW/cm² at 250W.
  • Model L300W-LP1-50: Designed for high power densities and long pulses with an LP1 absorber type. It covers a spectral range of 0.35-2.2 µm and 10.6 µm, with a power range of 1W - 300W. The maximum average power density is 23 kW/cm² at 300W.
Energy Mode:
  • L250W: Energy range from 120mJ to 200J with energy scales of 200J / 30J / 3J.
  • L300W-LP1-50: Energy range from 200mJ to 300J with energy scales of 300J / 30J / 3J.
Cooling and Accessories:
  • Both models require water cooling with a minimum flow rate of 1 liter/min. The water temperature should be between 18-30°C with a rate of change less than 1°C/min.
  • Accessories for high power sensors are available on pages 59, 60, and 61.
Additional Notes:
  • LP1 sensors have a calibrated spectral curve for accuracy across their spectral range. When used with Nova, Orion, and LaserStar meters, accuracy is ±3% for 1.06µm and 10.6µm, and ±6% for other wavelengths in the 600-1100nm range.
See more

Catalog excerpts

High Power Thermal Sensors-1

1.1.2.7 High Power Thermal Sensors 1.1.2.7.2 High Power Water Cooled Thermal Sensors 1W to 300W Features ֺ ֺ High powers ֺ ֺ Water cooled ֺ ֺ Up to 300W ֺ ֺ Ø50mm aperture Use Absorber Type Spectral Range µm Aperture mm Power Mode Power Range Power Scales Power Noise Level Maximum Average Power Density kW/cm2 Response Time with Meter (0-95%) typ. s Power Accuracy +/-% Linearity with Power +/-% Energy Mode Energy Range Energy Scales Minimum Energy mJ Maximum Energy Density J/cm2 <100ns 1µs 0.5ms 2ms 10ms Cooling Minimum Water Flow Rate at Full Power Accessories for High Power Sensors Weight kg Version Part number General purpose Broadband 0.19 - 20 Ø50mm High power densities and long pulses LP1 0.35-2.2, 10.6 Ø50mm Calibrated for 1.064µm and 10.6µm. LP1 sensors have relatively large spectral variation in absorption and have a calibrated spectral curve at all wavelengths in their spectral range to the above specified accuracy. Nova, Orion and LaserStar meters do not support this feature and when used with those meters, accuracy will be ±3% for 1.06µm and 10.6µm, and ±6% for other wavelengths in the spectral range 600-1100nm. Water temperature range 18-30°C. Water temperature rate of change <1°C/min. Water Connector R1/4-20 BSW x6 deep Water Connector for 1/4" Plastic Hose For latest updates pleas

 Open the catalog to page 1

All Ophir Optronics catalogs and technical brochures

  1. SP932U

    1  Page

  2. ConoPoint-3

    2  Pages

  3. nova catalog

    2  Pages

  4. vega catalog

    2  Pages

  5. BeamWatchAM

    2  Pages

  6. StarBright

    2  Pages

  7. 680274

    1  Page

  8. StarLite

    2  Pages

  9. M200s

    3  Pages

  10. NanoScan 2s

    14  Pages

  11. BeamWatch®

    5  Pages

  12. ModeCheck ®

    3  Pages

  13. Nova

    2  Pages

  14. Power Pucks

    1  Page

  15. StarBrigh

    2  Pages

  16. Orion

    2  Pages

  17. Laserstar

    2  Pages

  18. RP Sensors

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.