HOBO UX100-014M
8Pages

{{requestButtons}}

Catalog excerpts

HOBO UX100-014M - 1

onset HOBO Data Loggers HOBO® Single Channel Thermocouple Data Logger (UX100-014M) Manual HOBO Single Channel Thermocouple Data Logger UX100-014M Included Items: • Command™ strip • Double-sided tape • Hook & loop strap Required Items: • HOBOware 3.4 or later • USB cable (included with software) • Thermocouple sensor type J, K, T, E, R, S, B, or N The HOBO Single Channel Thermocouple data logger records temperature in indoor environments with J, K, T, E, R, S, B, or N type thermocouple sensors (sold separately). The logger also has an internal 10K thermistor for cold-junction compensation of the thermocouple output. The thermocouple channel has a 20-bit resolution while the internal 10K thermistor has 12-bit resolution. Using HOBOware®, you can easily configure the logger alarm to trip for specific high or low temperatures. Or, you can set up burst logging in which the logger records data at a different interval during certain conditions. The logger can also calculate minimum, maximum, average, and standard deviation statistics. This compact data logger has a built-in LCD screen to monitor temperature, logging status, battery use, and memory consumption in between readouts. Specifications Thermocouple Type Range Accuracy Resolution Internal 10K Thermistor (Temperature) Range -20° to 70°C (-4° to 158°F) Accuracy ±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F), see Plot A Logger Logger Operating Range Logging: -20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing); Launch/Readout: 0° to 50°C (32° to 122°F) per USB specification Logging Rate 1 second to 18 hours, 12 minutes, 15 seconds Logging Modes Fixed interval (normal), burst, or statistics Memory Modes Wrap when full or stop when full Start Modes Immediate, push button, date & time, or next interval Stop Modes When memory full, push button, or date & time Restart Mode Push button Time Accuracy ±1 minute per month at 25°C (77°F), see Plot B Power Source One 3V CR2032 lithium battery and USB cable Battery Life 1 year, typical with logging rate of 1 minute and sampling interval of 15 seconds or greater Memory 512 KB (208,076 measurements, maximum)

Open the catalog to page 1
HOBO UX100-014M - 2

HOBO Single Channel Thermocouple Data Logger (UX100-014M) Manual Specifications (continued) Logger (continued) Download Type Full Memory Download Time LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range Environmental Rating IP50 The CE Marking identifies this product as complying with all relevant directives in the European Union (EU). Plot A: Internal Temperature Accuracy & Resolution Logger Components and Operation Plot B: Time Accuracy Battery Tray Start/Stop Button Alarm/Stats Button Internal 10K Thermistor (Temperature)...

Open the catalog to page 2
HOBO UX100-014M - 3

HOBO Single Channel Thermocouple Data Logger (UX100-014M) Manual HOBO Single Channel Thermocouple Data Logger (UX100-014M) Manual MEMORY ■HID USB Port: Use this port (not visible in the diagram) to connect the logger to the computer or the HOBO U-Shuttle via USB cable (see Setting up the Logger and Reading Out the Logger). LCD Screen: This logger is equipped with an LCD screen that displays details about the current status. This example shows all symbols illuminated on the LCD screen followed by definitions of each symbol in the table below. Description The logger is waiting to be launched....

Open the catalog to page 3
HOBO UX100-014M - 4

HOBO Single Channel Thermocouple Data Logger (UX100-014M) Manual 2. Select the thermocouple type. Under Configure Sensors, select the correct type of thermocouple from the list as shown in this example. Type a label for the thermocouple if desired. 3. Select the internal 10k thermistor (temperature) channel to log (optional) and type a label if desired. This internal 10K thermistor measures ambient temperature and is used for cold-junction compensation of the thermocouple output. Important: Cold-junction compensation is always performed on the thermocouple measurements even if the internal...

Open the catalog to page 4
HOBO UX100-014M - 5

HOBO Single Channel Thermocouple Data Logger (UX100-014M) Manual Connecting the Thermocouple Sensor The logger can accept any standard thermocouple sensor of the appropriate type with standard thermocouple subminiature connectors. ANSI Standard Thermocouple Color Coding Type-J Iron-Constantan Black Type-K Chromel-Alumel Yellow Type-S Platinum-Rhodium Green Type-T Copper-Constantan Blue Type-E Chromel-Constantan Purple Type-R Platinum-Rhodium Green Type-B Platinum-Rhodium Gray Type-N Nicrosil-Nisil Orange Refer to http://www.omega.com/techref/colorcodes.html for additional details on color...

Open the catalog to page 5
HOBO UX100-014M - 6

HOBO Single Channel Thermocouple Data Logger (UX100-014M) Manual was furthest out of range for the sensor before the alarm cleared (see the Points table for the actual value). • Alarms can only be configured on the thermocouple channel; they are not available for the internal 10K thermistor (temperature) channel. Burst Logging Burst logging is a logging mode that allows you to set up more frequent logging when a specified condition is met. For example, let's say the logger is recording data at a 5-minute logging interval and burst logging is configured to log every 10 seconds when the...

Open the catalog to page 6
HOBO UX100-014M - 7

HOBO Single Channel Thermocouple Data Logger (UX100-014M) Manual logger will take a temperature sample every 30 seconds and temporarily store it in memory. The logger will then calculate the maximum, minimum, average, and standard deviation using the samples gathered over the previous 5-minute period and log the resulting value(s). When reading out the logger, this would result in 5 data series: one temperature series (with data logged every 5 minutes) plus four maximum, minimum, average, and standard deviation series (with values calculated and logged every 5 minutes based on the 30-second...

Open the catalog to page 7
HOBO UX100-014M - 8

HOBO Single Channel Thermocouple Data Logger (UX100-014M) Manual Battery Information The logger contains a 3V CR2032 battery (HRB-TEMP). Expected battery life varies based on the ambient temperature where the logger is deployed, the logging or sampling interval, frequency of offloading to the computer, number of channels that are active, if burst or statistics logging modes are active, and battery performance. A new battery typically lasts 1 year with logging intervals greater than 1 minute. Deployments in extremely cold or hot temperatures, a logging interval faster than 1 minute, or a...

Open the catalog to page 8