video corpo

501 Series

501 Series

501 Series

Product catalog summary
Applications
The 501 Series Continuous Multi-Gas Ambient Air Monitor is designed for the simultaneous analysis of up to four gases in ambient air, including O2, CO2, NO, NO2, CO, temperature, and % RH. Single gas versions are available for H2S, SO2, HCL, Cl2, and NH3.
Features
  • All gas detectors are housed in a NEMA 12 or optional NEMA 4 wall-mounted enclosure.
  • Each gas has a separate sensor and a bright digital readout.
  • Includes a built-in sampling pump and a modular layout for easy maintenance.
Options
  • Recorder output options include 4-20mA or voltage for all gases.
  • Isolated analog 4-20mA output is available.
  • Alarm lights and relay contacts for each gas, with a common 'Sonalert' audible alarm and reset button.
  • Red strobe light mounted on top of the analyzer.
  • Low sample flow alarm and multiple alarm points for measured gases.
  • Multipoint sampling using Nova 402AS AutoSequencer.
Calibration
  • Zero calibration on clean air or nitrogen.
  • Span calibration on analyzed calibration gas.
Description
The 501 Series offers four gas versions with options for O2, CO2, CO, NO, and NO2. Single gas versions can include an O2 sensor for toxic gas plus oxygen deficiency monitoring. A built-in pump draws samples for analysis, and a ¼" plastic tube can be attached for sampling different areas.
Specifications
  • Detection Method: Electrochemical sensor for most gases, dielectric ceramic sensor for % RH, infrared sensor for CO2.
  • Ranges Available: Various ranges for different gases, e.g., 0-25.0% O2, 0-5000 PPM CO2.
  • Resolution: 0.1% on O2; 1 PPM on PPM gases; 1% RH; 0.1°C Temp.
  • Accuracy and Repeatability: ± 2% of full scale.
  • Drift: Less than 2% per month.
  • Response Time (T-90): Generally under 2 minutes.
  • Ambient Temperature Range: 40° to 105°F (4° to 40°C).
  • Linearity: Better than 2% of full scale.
  • Power: 115VAC 60 Hz (220VAC 50 Hz available).
  • Output Options: 4-20mA into 500 ohms non-isolated standard, optional isolated 4-20mA.
  • Alarms: Optional alarms with separate SPDT relay contacts and indicator lights, common 'Sonalert' audible alarm with reset button.
Unique Applications
Warm-up time is approximately four minutes after initial power-up. The operating ambient humidity range is 0-90% RH non-condensing.
See more

Catalog excerpts

501 Series-1

501 SERIES CONTINUOUS MULTI-GAS AMBIENT AIR MONITOR APPLICATIONS For simultaneous analysis of up to four gases in ambient air including O2, CO2, NO, NO2, CO, temperature, and % RH. Single gas versions are also available for H2S, SO2, HCL, Cl2, & NH3. FEATURES • All gas detectors contained in NEMA 12 or optional NEMA 4 wall-mounted enclosure • Separate sensor for each gas • Bright digital readout for each gas • Built-in sampling pump • Easy to maintain modular layout OPTIONS • Recorder output of 4-20mA or voltage for all gases • Isolated analog 4-20mA output available • Alarm lights and relay contacts for each gas with common 'Sonalert' audible alarm and reset button • Red strobe light mounted on top of analyzer • Low sample flow alarm • Multiple alarm points for measured gases • Multipoint sampling using Nova 402AS AutoSequencer CALIBRATION • On clean air or nitrogen for Zero • On analyzed calibration gas for Span NOVA ANALYTICAL SYSTEMS www.nova-gas.com NEMA 12 and NEMA 4 enclosures shown with optional red strobe light

 Open the catalog to page 1
501 Series-2

DESCRIPTION The 501 Series four gas versions offer a choice of O2, CO2, CO, NO, & NO2. Single gas verions are also available with H2S, SO2, Cl2, HCL, & NH3. An O2 sensor may be added to the single gas versions to produce a toxic gas plus oxygen deficiency monitor. A built-in pump draws in the sample for analysis. A ¼" plastic tube may be attached to the analyzer to sample a different area, such as a room or confined space. Standard system comes with NEMA 12 enclosure, built-in sampling pump, flow meter, pre-filter, and liquid blocker. Alarm option - alarms may be provided for all gases. An optional...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.