1. Catalogs
  2. Nira Srl
  3. Fast Residual solvent analyser Neptune 803

Fast Residual solvent analyser Neptune 803

Fast Residual solvent analyser Neptune 803

Fast Residual solvent analyser Neptune 803

Product catalog summary
Introduction
The Neptune 803 is a sophisticated instrument designed for quality control in production areas, particularly for rapid analysis by machine operators. It is equipped with a flame ionization detector and features an automatic hydrogen closing device for safety.

Specifications
The Neptune 803 includes an industrial PC with a Windows Embedded operating system and a touch screen interface. It supports Ethernet and USB connections, and features an integrated headspace without the need for an external sample transfer line. The system is designed to operate efficiently in solvent-polluted areas.

Features and Benefits
The instrument offers reduced maintenance due to its loop sampling method, which enhances analytical stability and reduces maintenance frequency. It provides real-time process corrections, minimizing waste and rework. The system includes a variety of components such as an Ultimetal capillary column, printer, sample receptacles, air purifier, and optional hydrogen generator.

New Features
Recent updates include a pulse washing system, electronic pressure control, an ultra-sensitive FID detector, and remote online service capabilities. The software interface has been improved for ease of use.

Accessories
Optional accessories include an air-conditioned pulpit assembly, zero air generator, circular sample cutter, and analytical micro syringe.

Compliance and Verification
The Neptune 803 complies with EN 13628 and ASTM 1884 guidelines and has been verified by third-party laboratories.

Technical Characteristics
Key technical specifications include:
  • Incubation time: 5, 10, 30 minutes
  • Operating temperature: 10 to 40°C (up to 50°C for process version)
  • Incubation temperature: 50°C to 150°C (± 0.1°C)
  • Measuring range for Ethyl Acetate: 0 - 135mg/m²
  • Minimum detectable level: 0.1 mg/m²
  • Accuracy: ±1%
  • Calibration frequency: 6 months
  • Response time: 7 minutes for single films, 12 minutes for laminated films
  • Hydrogen and air requirements: High purity and specific consumption rates
  • User interface: 10.5" color touch screen
  • Power supply: 230Vac 50/60 Hz
  • Dimensions and weight vary by version
See more

Catalog excerpts

Fast Residual solvent analyser Neptune 803-1

New Instruments and Research for Analysis NEPTUNE 803 is designed for the rapid execution of the analysis performed by the machine operators, directly next to production processes, even in solvent polluted areas. By setting off the waiting time of the laboratories (sampling performed directly by the operators) and compressing, thanks to innovative instrumental solutions, the technical analysis times, Neptune 803 is able to provide a residual solvent result in the time necessary to produce a single roll (12 minutes). The use of Neptune 803 allows process corrections in real time, greatly reducing waste and rework of products. Flame ionization detector with conveyed discharge. Automatic hydrogen closing device in case of anomalies. Industrial PC (OS Windows®) and touch screen interface. Integrated headspace without external sample transfer line. Ethernet and USB networking connection. Easy touch screen interface. The industrial PC housed in the analyzer, the Windows Embedded® operating system and the software have been developed to be managed by machine operators, they guide and support them in each required step. Any anomaly is recorded 24/7. It uses a known volume (Loop) sampling, with no mechanical moving part, that guarantee two exclusive advantages compared to traditional split injection systems: - Greater analytical stability over time - Less maintenance frequency

 Open the catalog to page 1
Fast Residual solvent analyser Neptune 803-2

NEW FEATURES - Exclusive headspace “pulse” washing system. - Full electronic pressures control (no more manual adjustments needed). - Easier new software interface and touch screen. - New ultra-sensitive FID detector, self-heated. - Remote online service: screen sharing with our technical staff. - Multiple incubation options. - Downloadable archives as .CSV and .pdf files. - Windows® operating system - Automatic self diagnostic NORMATIVES AND METHODS ACCESSORIES > Air conditioned pulpit assembly, to be used in production areas, IP55 protected. > Zero air generator to use factory compressed air...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.