MaiYa MST-4 NH3 sensor

MaiYa MST-4 NH3 sensor

MaiYa MST-4 NH3 sensor

Product catalog summary
Sensor Specification
  • Product Model: 4NH3-100
  • Normal Range: 0-100 ppm
  • Max Range: 200 ppm
  • Sensitivity: 0.10±0.03 µA/ppm
  • Resolution: 0.5 ppm
  • Response Time (T90): <60 seconds
  • Max Zero Drift: 1 ppm (from +20℃ to +40℃)
  • Repeatability: 1% output signal
  • Linearity: Regression coefficient R2=0.999
  • Operating Temperature: -20℃ to +50℃
  • Humidity: 15%~90% RH, non-condensing
  • Pressure: Atmospheric pressure ±10%
  • Pressure Factor: < 0.02% signal /MBar
  • Baseline Extent: <±0.4µA
  • Long Term Drift: < 5% signal loss per year
  • Bias Voltage: No bias voltage
  • Load Resistance: 10Ω
  • Built-In Filters: No
  • Lifetime: 2 years (in air)
  • Warranty: 1 year from the date of shipment
  • Output Current Polarity: Positive current
Cross Interference

Cross-interference data is for reference only and not guaranteed for accuracy. The sensor response to various gases is as follows:

  • CO: 100 ppm results in 60 ppm sensor response
  • H2: 200 ppm results in 2 ppm sensor response
  • CO2, NO2, CH4, C2H4: No response
Contact Information

Ningxia MaiYa Sensor Technology Development Co., LTD
Address: Tower C, Yinchuan Zhongguancun Innovation Center, Helanshan Road, Xingzhou Street, Yinchuan, Ningxia, China 750000
Email: [email protected] / [email protected]
Phone: +86 189 9527 8133 / +86 182 0276 0413
Whatsapp: +86 18202760413
Website: maiyachuangan.com, maiya-sensor.com

Declaration
  • 4-Series sensors are for gas leak monitoring and alarms, not for long-term exposure or continuous detection.
  • Sensors must be connected via a PCB socket; direct welding will damage them.
  • Avoid contact with organic solvents, alcohol, paint, oil, and high concentrations of gas.
  • Sensors with positive output current require oxygen for reaction and should be calibrated with air as the background gas.
  • Do not remove or insert the sensor when powered on to avoid damage.
See more

Catalog excerpts

MaiYa MST-4 NH3 sensor-1

Electrochemical sensor series Sensor specification Product Model Max zero drift (+20℃ to 1ppm +40℃) Repeatability Linearity, regression coefficient R2=0.999 15%~90%RH (relative humidity), non-condensing Pressure factor Baseline extent Long term drift < 5% signal loss / year Bias voltage Load resistance Built-In Filters 1 year (from the date of shipment) Output current polarity Positive current (define the output current of carbon monoxide and hydrogen sulfide sensor as positive) Ningxia MaiYa Sensor Technology Developme

 Open the catalog to page 1
MaiYa MST-4 NH3 sensor-2

Electrochemical sensor series Product size (unit: mm) R : reference electrode W: working/sensing electrode C : counter electrode Cross interference: Due to the difference of standard gas, cross-interference data is for reference only. The measurement of intersecting gases is not linear. There is no guarantee for accuracy of the measurement of the crossing gases. GAS Sensor response(ppm) Ningxia MaiYa Sensor Technology Development Co.,LTD

 Open the catalog to page 2
MaiYa MST-4 NH3 sensor-3

Electrochemical sensor series Declaration: 1. 4-Series electrochemical gas sensors are designed for gas leak monitoring and alarm. They are not suitable for long-term exposure to the gas to be measured or long-term on-line continuous detection and analysis instruments. 2. Sensor pins must be connected through a PCB socket, Welding the pin directly will damage the sensor. 3. Sensors should avoid contact with organic solvents, alcohol, paint, oil and high concentration of gas, including silica gel and other adhesives. 4. Electrochemical sensors with positive output current (such as CO, H2S, SO2,...

 Open the catalog to page 3

All Ningxia Maiya Sensor Technology catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.