LCR-Active-Head-IAI2
1Pages

{{requestButtons}}

Catalog excerpts

LCR-Active-Head-IAI2 - 1

LCR Active Head and Impedance Analysis Interface The PSM1700 and PSM1735 can provide LCR measurements in either ‘passive mode’, where an external shunt is used for current detection or with optional impedance measurement accessories available from N4L. Passive mode: In this mode, an external shunt is placed in series with the component under test, then CH1 is used to measure the voltage across the component under test while CH2 is used to measure the voltage across the shunt. From these voltages and their relative phase angle, the PSM units compute impedance and all LCR functions. The LCR Active Head and IAI convert the PSM 1700/1735 units into high performance LCR meters with true 4 wire Kelvin connections that are taken directly to the component under test without the need for external shunts. Buffering, amplification and selectable shunts provide LCR measurements over a wide frequency and impedance range. Standard accessories and options LCR Active Head supplied with: BNC to Kelvin clip lead set (Power and comms via hard wired Auxiliary lead) Impedance Analysis Interface supplied with: BNC to Kelvin clip lead set, 3 x BNC link cables, 1 x comms link, 1 x IEC power cable IAI fixture option: IAI Kelvin fixture - to simplify connection and HF accuracy with axial / radial components Newtons4th Ltd 1 Bede Island Road Leicester LE2 7EA UK Tel: +44 (0)116 2301066

Open the catalog to page 1

All Newtons4th Ltd. catalogs and technical brochures

  1. PPA3500