1. Catalogs
  2. neoplas control GmbH
  3. Q-MACS Process fibre

Q-MACS Process fibre

Q-MACS Process fibre

Product catalog summary
Overview
The Q-MACS Process fibre is a portable monitoring system designed for online measurements in industrial processes. It is based on the Q-MACS Basic and features a two-channel setup for enhanced long-term stability. The system is directly coupled to the measuring path via an infrared (IR) fibre and utilizes infrared absorption spectroscopy to measure absolute molecular concentrations.
General Specifications
  • Type: Two-path infrared spectrometer with IR-light source and IR-fibre coupling
  • Sensitivity: Down to parts per billion (ppb) range
  • Response Time: Down to milliseconds
  • Time Resolution: Down to milliseconds, with sub-microsecond resolution available upon request
  • Size: 554 mm x 962 mm x 600 mm
  • Weight: 125 kg
Components
  • Optical board
  • Light guide cable coupling
  • Industrial PC with acquisition hardware
  • Electronic supply system
  • Water cooling system
Power and Environmental Parameters
  • Power: 230 V, max. 2 A (switch-on current 6 A) or 115 V, max. 4 A (switch-on current 12 A)
  • Working Range: +5 °C to +40 °C
Quantum Cascade Laser (QCL) Specifications
  • Tuning Method: Inter pulse mode (laser sweep mode) and intra pulse mode (single pulse mode)
  • Pulse Width: 8 ns to 256 ns (longer pulses available on request)
  • Pulse Frequency: Up to 1 MHz
  • QCL Temperature Range: -35 °C to +40 °C
  • QCL Status: Tested and installed
Applications
The system is suitable for operations in gases and plasmas, with sensitivity dependent on species, temperature, and pressure.
See more

Catalog excerpts

Q-MACS Process fibre-1

solutions for your operations in gases and plasmas Q-MACS Process fibre technical specification The Q-MACS Process fibre is a portable monitoring system for online measurements of industrial processes. It is bases on the Q-MACS Basic and features a two channel setup to achieve long-term stability and can be directly coupled to the measuring path via an IR-fibre. The system uses infrared absorption spectroscopy to measure absolute molecular concentrations. general description two path infrared spectrometer with IRlight source, IR-fibre coupling response time time resolution down to milliseconds, sub-microsecond time resolution on request components parts ▪▪ optical board ▪▪ light guide cable coupling ▪▪ industrial PC with acquisition hardware ▪▪ electronic supply system ▪▪ water cooling system parameter power ▪▪ 230 V, max. 2 A (switch-on current 6 A) ▪▪ 115 V, max. 4 A (switch-on current 12 A) working range ▪▪ inter pulse mode (laser sweep mode) ▪▪ intra pulse mode (single pulse mode) pulse width 8 ns* … 256 ns** * depends on the QCL and QCL-voltage used ** longer pulses on request pulse frequency

 Open the catalog to page 1
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.