Q-MACS MPO HERRIOT

Q-MACS MPO HERRIOT

Product catalog summary
Overview: The Q-MACS MPO Herriott cell is an advanced optical device designed to enhance the efficiency of gas volume reduction through increased optical path lengths. This is achieved by optimizing the mirror area usage compared to traditional Herriott cell designs. Its compact structure is ideal for applications requiring extended interaction paths between gaseous media and radiation sources, such as mid-infrared lasers, while maintaining minimal optical setup dimensions. The low gas volume facilitates rapid gas exchange rates with high sensitivity.
Applications: The Herriott cell is integral to various sensor systems and OEM modules, particularly in monitoring ambient trace gas concentrations, exemplified by the Q-MACS Trace compact series. Its robust and compact design makes it suitable for integration as an OEM module in industrial products.
Specifications:
  • Type: Compact closed absorption cell of Herriott type
  • Dimensions: 195 mm x 83 mm x 83 mm (L x H x D) [excluding connectors]
  • Weight: Less than 1700 g
  • Mechanical Interface:
    • Mirror Material: Gold-coated copper (alternative materials available upon request)
    • Window Material: Potassium bromide (KBr) (alternative materials available upon request)
    • Cell Body: Stainless steel, Pyrex tube, FKM o-rings
    • Volume: 0.35 liters (custom solutions available)
    • Gas Connectors: Festo 4 mm Push-in fittings
    • Pressure Range: 10 mbar to 2 bar
    • Gas Leak Rate: Less than 1 µbar*l/s
  • Optical Interface:
    • Input Hole Diameter: 6.7 mm
    • Mirror Diameter: 50 mm
    • Wavelength Range: 2 µm to 12 µm
    • Reflectivity: Greater than 98%
    • Number of Passes: 184
    • Path Length: 27.55 meters (custom solutions available)
    • Optical Axis Height: 63 mm
    • In-Out Half-Angle: 4 degrees
  • Operating and Storage Conditions:
    • Operating Temperature: -15°C to 150°C (extended ranges available upon request)
    • Operating Humidity: 15% to 75% relative humidity
    • Storage Temperature: -55°C to 70°C
    • Storage Humidity: 10% to 100% relative humidity
Contact Information: neoplas control GmbH, Walther Rathenau Straße 49a, 17489 Greifswald, Germany. Website: www.neoplas-control.de, Email: [email protected]
See more

Catalog excerpts

Q-MACS MPO HERRIOT-1

Q-MACS MPO HERRIOT Our latest Herriott cell design allows to significantly reducing the gas volume by increased optical path lengths. This performance is achieved by utilizing the mirror area more efficiently compared to common Herriott cell layouts. The compact construction makes it well suited to applications, which require a long interaction path between a gaseous medium and emission by a radiation source like mid infrared lasers, while keeping the dimensions of the optical setup as small as possible. Furthermore, the low gas volume ensures fastest gas exchange rates at very high sensitivity. The Q-MACS Herriott cell is a central part in various sensor systems and OEM modules for the monitoring of ambient trace gas concentrations like the Q-MACS Trace compact series. With its outstanding features the QMACS Herriott cell eases the design and implementation of laser-based systems in trace gas monitoring applications. Its robustness and compactness makes it the ideal choice for the integration as an OEM module into industrial products. general description dimensions weight optical interface compact closed absorption cell of type Herriott 195 mm x 83 mm x 83 mm (L x H x D) [without conectors] <1700 g input hole diameter mirror diameter wavelength range reflectivity number of passes path length gold coated copper (other material on request) potassium bromide (KBr) (other material on request) stainless steel, pyrex tube, FKM o-rings 0.35 l (customized solution on request) Festo 4 mm Push-in fittings 10 mbar to 2 bar <1 µbar*l/s optical axis height in-out half-angle mechanical interface mirror material window material cell body volume gas connectors pressure range gas leak rate 6,7 mm 50 mm 2 µm to 12 µm >98 % 184 27.55 m (customized solution on request) 63 mm 4 operating and storage conditions operating temperature operating humidity storage temperature storage humidity -15 C to 150 C (extended temperature ranges on request) 15 % to 75 % (rel.) -55 C to 70 C 10 % to 100 % (rel.) neoplas control GmbH - Walther Rathenau Straße 49a - 17489 - Greifswald - Germany - www.neoplas-control.de - [email protected]

 Open the catalog to page 1
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About DirectIndustry