ECS8022
11Pages

{{requestButtons}}

Catalog excerpts

ECS8022 - 2

ECS 8022 – NC Organic Elemental Analysis ECS 8022 is the specialized N and C Elemental Analyzer Model based on the Dumas combustion method. ECS 8022 is prepared and dedicated to the carbon and nitrogen determination in a wide range of samples. ECS 8022 is based on sample combustion and separation of gases with a chromatographic column. The combustion products, i.e. CO2 and N2 are separated and quantified by a high resolution TCD detector. ECS 8022 is fully automatic, from the available samplers to the oxygen dosage, from the monitoring of consumables status to the automatic leak test

Open the catalog to page 2
ECS8022 - 3

Dual furnace combustion system (for a better combustion and optimization of catalysts) Safety quick fit system for reactors connection (easy and safe way to set the instrument) Customized GC column Detection system Data acquisition and handling Several configurations may be set for the determination of the target element. Instrument is already prepared for the determination of carbon and nitrogen. ECS 8022 could be also equipped with other tools for the upgrade of the instrument to the CHNS-O analyzer. Consumables kit are prepared for obtaining the best analysis performances. Prepacked...

Open the catalog to page 3
ECS8022 - 4

Three different samplers are available: Pneumatic autosampler up to 147 positions Electronic autosampler with 32, 50, 100 positions Manual sampler Leak Test inning leak Test. Automatisms make it particularly user-friendly: Automatic oxygen dosage system (for a better consumption of oxygen and consumables) Automatic consumables status monitoring Automatic leak test Standby mode (gas, energy and time savings) Automatic Amount Automatic oxygen dosage: main highlights. Details of sample are registered: Position Type Weight ECS 8024 will dose the oxygen for the right combustion. No oxygen or...

Open the catalog to page 4
ECS8022 - 5

EAS Clarity Powerful software for powerful instrument Getting data, analyzing data, presenting data Integration There is extensive possibility to modify chromatograms. The chromatogram can be changed by entering global parameters or interactively, through direct graphic modification of the baseline. Overlay Simultaneously displays a virtually unlimited number of chromatograms and their mathematical modification. Calibration Internal and external standard calculation methods. Automated measuring support Sequence tables for any set of samples.

Open the catalog to page 5
ECS8022 - 6

ECS 80 SERIES What the user can do Summary result tables Displays and prints selected results from all simultaneously displayed chromatograms. User settings User selects parameters for peak display and the specification for axes, including color from an extensive array of color settings. Text labels and lines, either as part of the area or anchored to a chromatogram, may also be inserted. Export Optional exportation of all results, with or without the chromatogram, in various formats, into a file or clipboard. Import Imports chromatograms or mathematical curves, which have been saved in...

Open the catalog to page 6
ECS8022 - 7

ECS 8022 key points  Fully automated analysis system  Consumables status monitoring for an optimization of catalysts usage  High sensitivity, accuracy and precision  Flexibility applications  Detector does not require reference gas  Three types of samplers available: electronic / pneumatic / manual.  Easy connection to Mass Spectrometers and other detectors for stable isotopes analysis  Powerful software for viewing results from a computer  Low operation and management costs  Touch-screen display settings management  Standby mode for gas, energy and time savings

Open the catalog to page 7
ECS8022 - 8

ECS 8022 application fields  Environmental analysis  Organic chemistry and pharmaceutical  Petrol chemistry and energy  Soil science and geology- marine science (distinction between organic and inorganic carbon is available through the previous acidification of the sample)  Materials characterization  Food (Special configuration for big size sample is easily a

Open the catalog to page 8
ECS8022 - 9

Analytical and Technical Features Accuracy* <0,2% (certified standard; purity >99.9%) Precision* <0,1% (certified standard; purity >99.9%) Pneumatic autosampler Electronic autosampler Manual sampler Dual furnace system Safety quick fit system Touch screen display Standby mode * Accuracy and precision are related to samples nature and homogeneity .

Open the catalog to page 9
ECS8022 - 10

Analytical and Technical Features Gas carrier Leak test Furnace temperature Oven temperature Oxygen volume need Flow rate Gas separation Detector Software data analysis Calibration Active calibration Helium Automatic Left Furnace : max 1100°C RightFurnace : max 1100°C Max 110°C Automatically calculated by the oxygen doser Electronic Flow Rate 3 m GC Column High Sensitivity TCD EAS Clarity Linear, Quadratic, Cubic As needed

Open the catalog to page 10
ECS8022 - 11

Innovative Elemental p-Analysis Q Via Milano,15/A - 20060 Bussero (MI), Italy V* Phone: +39 02 950 34 69 (|||) www.nctechnologies.it

Open the catalog to page 11

All NC Technologies catalogs and technical brochures

  1. ECS8024

    11 Pages

  2. EA ID MICRO

    16 Pages