Sample Carryover NanoDrop 8000
1Pages

{{requestButtons}}

Catalog excerpts

Sample Carryover NanoDrop 8000 - 1

Sample Carryover Introduction A study was conducted to demonstrate the lack of sample carryover on the Thermo Scientific NanoDrop™ 8000 spectrophotometer. The protocol included measuring 4 individual aliquots of water, followed by 4 aliquots of a high concentration DNA sample. Carryover was then assessed by the measuring of another 4 aliquots of water. The test results demonstrate that there is no significant sample carryover on the instrument measurement pedestals. The tables below display the results of the study using 3 separate measurement pedestal positions. Position/Replicate Sample ID Date Time ng/ul Table 1. Pedestal Position A Table 2. Pedestal Position D Table 2. Pedestal Position H water water water water 3600 ng/ul 3600 ng/ul 3600 ng/ul 3600 ng/ul water water water water Sample ID water water water water 3600 ng/ul 3600 ng/ul 3600 ng/ul 3600 ng/ul water water water water water water water water 3600 ng/ul 3600 ng/ul 3600 ng/ul 3600 ng/ul water water water water Thermo Fisher Scientific - NanoDrop products Technical support: info@nanodrop.com