Overview: This document provides detailed information on capacitance sensors designed for ungrounded or poorly grounded targets. It highlights the unique push/pull technology developed by MTII, which integrates two high-precision sensors into one body, suitable for various applications such as semiconductor thickness measurement, sheet metal thickness, and automotive brake rotor run-out.
Capacitance Technology: Traditional capacitive sensors require the target to be electrically grounded, which can lead to inaccuracies if the target's electrical conductivity or ground path changes. The push-pull design by MTII addresses this by using two capacitance sensors with a 180-degree phase shift, allowing current to travel across the target surface, thus eliminating inaccuracies from poorly grounded targets. This technology also enables measurement of highly resistive targets.
Push/Pull Technology: The push/pull amplifier design cancels common mode electrical noise, which is beneficial in environments with magnetic bearing surfaces, semiconductor wafers, and air bearing floating surfaces. The document details the types of push/pull probes and their specifications, emphasizing the importance of grounding the probe's shell to the amplifier for optimal performance.
Specifications: The document includes detailed specifications for various probe models, including dimensions, range, sensitivity, and resolution. It also provides information on compatible MTI capacitance amplifiers and optional accessories like low noise extension cables and calibration fixtures.
Calibration and Fixturing: The KD-CH-IIID calibrator is highlighted as a precision fixture for securing non-contact displacement sensors and varying target positions for calibration. The FS-2 fixture stand is described as an adjustable support system for positioning probes over targets.
Conclusion: MTII's push/pull capacitance sensors offer a robust solution for measuring ungrounded or poorly grounded targets, with applications across various industries. The document provides comprehensive technical specifications and accessory options to support these advanced measurement solutions.
See more

Catalog excerpts

CAPACITANCE PROBES - Special Push-Pull-1

A worldwide leader in precision measurement solutions Capacitance Sensors for Ungrounded Targets or Poorly Grounded Targets PUSH/PULL PROBES TWO HIGH PRECISION SENSORS BUILT INTO ONE BODY

 Open the catalog to page 1
CAPACITANCE PROBES - Special Push-Pull-2

Capacitance Technology: Standard capacitive sensors require the target to be electrically grounded. Current flows from the probe face to the target and back to the amplifier to complete the circuit. The capacitance between probe and target is proportional to the distance and converted to a 0 - 10V output from the amplifier. The measurement of electrically grounded targets can be, however, affected by changes in the electrical conductivity or ground path of the target. Standard Capacitance Sensor Ground Shell Guard Sensing Area TARGET GROUND Ground Return Unique Push/Pull Technology: Two Probes...

 Open the catalog to page 2
CAPACITANCE PROBES - Special Push-Pull-3

11.58 mm (0.456 inch) 20 mm (0.787 inch) 100 mm (3.937 inch) Spot Size 11.58 mm (0.456 inch) 16 mm (0.63 inch) 100 mm (3.937 inch) Spot Size Range Extension Range Extension 9.45 mm (0.372 inch) 14.8 mm (0.583 inch) 34.04 mm (1.34 inch) Spot Size 7.11 mm (0.28 inch) 16 mm (0.63 inch) 50 mm (1.969 inch) Spot Size Range Extension Probe Connector Types: (Ground shell must be grounded to amplifier for proper performance) ILA Integrated Lead Axial ILR Integrated Lead Radial

 Open the catalog to page 3
CAPACITANCE PROBES - Special Push-Pull-4

E 7.42 mm (0.292 inch) D 10.26 mm (0.404 inch) L 31.75 mm (1.25 inch) Spot Size Range Extension E 10.26 mm (0.404 inch) D 13.97 mm (0.55 inch) L 31.75 mm (1.25 inch) Spot Size Range Extension E 12.83 mm (0.505 inch) D 19.05 mm (0.75 inch) L 31.75 mm (1.25 inch) Spot Size Range Extension E 16.15 mm (0.636 inch) D 22.96 mm (0.904 inch) L 31.75 mm (1.25 inch) Spot Size Probe Connector: (Ground shell must be grounded to amplifier for proper performance) ILA Integrated Lead Axial

 Open the catalog to page 4
CAPACITANCE PROBES - Special Push-Pull-5

Push/Pull Probe Specifications Reference Range Spot Size Standard Push/Pull probes are operated at 5X range extension with AccumeasureTM 500 capacitance modular rack systems and AS-562 OEM Boards. Compatible with the following MTI Capacitance Amplifiers Analog Output Benchtop Capacitance Modular System Analog Output Capacitance OEM Board The FS-2 is an adjustable fixture to support and position probes over the target being measured. The system consists of a probe clamp, 2 positioning arms, magnetic base, c-clamp base and a micrometer with a 7mm (0.28") travel. The probe clamp holds probes up...

 Open the catalog to page 5

All MTI Instruments catalogs and technical brochures

  1. 1510A

    2  Pages

  2. Proforma 300i

    2  Pages

  3. PBS-4100+

    6  Pages

  4. MICROTRAK 3

    4  Pages

  5. PBS-4100R+

    8  Pages

  6. PV Series

    6  Pages

  7. 1510A

    6  Pages

  8. 42CA

    1  Page

  9. 41CA

    1  Page

  10. 45CA

    1  Page

  11. TSC-4800A

    6  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.