  3. Pyroelectric Detectors for Oriel Merlin™ Radiometry System
video corpo

Pyroelectric Detectors for Oriel Merlin™ Radiometry System

Pyroelectric Detectors for Oriel Merlin™ Radiometry System

Product catalog summary
Overview: This document provides technical specifications and details for Pyroelectric Detectors used in the Oriel Merlin™ Radiometry System. These detectors are designed for broad spectral range sensitivity and stable response across ultraviolet to infrared wavelengths.
Specifications: The detectors are available in two sizes: 2 mm and 5 mm diameter. The 2 mm detectors are calibrated to 2000 V/W, while the 5 mm detectors are calibrated to 1000 V/W. They use lithium tantalate crystals with a proprietary black coating to enhance responsivity and stability.
Performance: The noise equivalent power (NEP) for the 2 mm detector is 5 x 10-10 for a 1 Hz bandwidth, allowing for measurement of power levels down to a nanowatt with appropriate shielding. The upper power limit is 2 mW for the 2 mm detector, with higher powers measurable using an attenuator.
Calibration: Each detector is calibrated at 633 nm, and the flat response eliminates the need for calibration at additional wavelengths.
Components: Each detector package includes a pyroelectric detector with a window, detector housing with a built-in preamplifier, and necessary cables. Additional components required for a complete system include the Merlin Control Unit and a chopper.
Ordering Information: The document lists models 70362 and 70364 with CaF2 windows, and models 70363 and 70365 with KRS 5 windows, available in 5 mm and 2 mm diameters respectively.
See more

Catalog excerpts

Pyroelectric Detectors for Oriel Merlin™ Radiometry System  -1

1230 Phone: 1-800-222-6440 • Fax: 1-949-253-1680 S p e c t r o s c o p y I n s t r u m e n t s MONOCHROMATORS AND SPECTROGRAPHS BENCHTOP SPECTROMETERS MINISPECTROMETERS SINGLE CHANNEL DETECTION SYSTEMS AND CCD CAMERAS ACCESSORIES FOR FT-IR SPECTROMETERS SPECTROSCOPY INSTRUMENTS Pyroelectric Detectors for Oriel Merlin™ Radiometry System 1.0 0.8 0.6 0.4 0.2 0.0 0.1 1 10 WAVELENGTH (ìm) RELATIVE RESPONSIVITY 70362, 70364 CaF2 70363, 70365 KRS 5 Relative spectral response of Pyroelectric detectors. The window material sets the useful limits. Pyroelectric Detectors are sensitive over a broad spectral range. The responsivity varies very little from the ultraviolet to the infrared. We calibrate these detectors for response in V/W. The Merlin™ Control Unit can then display the reading in watts. Stability Because the responsivity of pyroelectric detectors is fairly temperature insensitive, you get good stability of response. The high absorptance of our durable coating has excellent long term stability. Measurable Power The noise equivalent power (NEP) of the 2 mm detector is 5 x 10-10 for a 1 Hz bandwidth. Since the NEP falls as bandwidth is reduced, you can, in principle, measure even lower powers if you are prepared to wait, and if all the noise is "well behaved." With reasonable electrical and acoustic shielding, we measure power levels down to a nW with a 1 s time constant at a 2:1 signal to noise ratio, in our laboratory. We regard this as a practical lower limit. The upper power limit for Merlin, with a 2 mm pyroelectric detector, is 2 mW, based on the calibration factor of 2000 V/W. With an attenuator you can measure higher powers. The lower practical limit for the 5 mm unit is approximately 3 nW due to the noise associated with the large area of the detector element. Calibration Each unit is calibrated at 633 nm. We do not offer detectors calibrated at more wavelengths, as the flatness of these detectors' response is quite unique and eliminates the need for more wavelength specific data. All the 2 mm units have their responsivity adjusted to 2000 V/W. The 5 mm units are shipped adjusted to 1000 V/W. Lithium Tantalate Elements These detectors use lithium tantalate crystals for high responsivity, to minimize sensitivity to microphonics and for long term stability. The crystals have a proprietary black coating to enhance the responsivity and remove the sharp spectral variations beyond 10 ìm due to crystal absorption bands. We offer these detectors with either CaF2 or KRS 5 windows. This is done for detector protection and noise suppression, since these thermal detectors are very sensitive to air drafts. Detector Systems We offer 2 mm and 5 mm diameter detectors. The smaller detectors are useful for many laser or focused beam applications; the 5 mm models are useful for monitoring the output of a monochromator or for applications where it is not possible to tightly focus the radiation. Each pyroelectric detector package includes: • Pyroelectric detector with window • Detector housing with built-in preamplifier • All necessary interconnecting cables To complete a Merlin System, you will also need to order: • Merlin Control Unit • Chopper • Sensitive over broad spectral range • Good, stable response Ordering Information Model Detector Active area Window Material 70362 5 mm diameter CaF2 70363 5 mm diameter KRS 5 70364 2 mm diameter CaF2 70365 2 mm diameter KRS 5 See our website WEBfor more info

 Open the catalog to page 1

All MICRO-CONTROLE / Spectra-Physics catalogs and technical brochures

  1. 10BPF10-310

    1  Page

  2. M-401

    1  Page

  3. newport ressource

    1640  Pages

  4. Motion PL30

    24  Pages

  5. Nanopositioners

    40  Pages

  6. Beamsplitters

    34  Pages

  7. Optrical mirror

    33  Pages

  8. Motion System

    2  Pages

  9. Base Clamps

    1  Page

  10. Rod Clamps

    1  Page

  11. V-Blocks

    1  Page

  12. Blackbodies

    2  Pages

  13. Beam Dump

    1  Page

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.