DCC1000e System - 8 Pages

Add to MyDirectIndustry favorites
DCC1000e System

Catalog excerpts

Degassed Cation Conductivity DCC1000e System All conductivity measurements Calculated pH, ammonia and CO2 Trend graphs Resin health monitoring Degassed Cation Conductivity Measurement Reliable and Accurate Cycle Chemistry Monitoring

 Open the catalog to page 1

DCC1000e System for Precise Detection of Corrosive Contaminants The METTLER TOLEDO Thornton DCC1000e System offers an enhanced design for conductivity measurement for power cycle chemistry monitoring. By providing conductivity measurement in compliance with ASTM D4519, the DCC1000e is an easy to use system that ensures water purity to maximize power production and minimize corrosion. Features • Precise detection of corrosive contaminants • Multi-parameter M800 Transmitter with single-screen display of all measurements • Integrated flow sensor with automatic heater shut-off if flow stops •...

 Open the catalog to page 2

Applications Feed water and steam monitoring during plant startup enable the best decisions to be made on increasing load. By monitoring specific, cation and degassed conductivity, the technician can ensure the water quality is adequate to bring the plant online. Power steam quality monitoring ensures contaminant limits specified in turbine warranties are met. By eliminating interference from CO2 and ammonia, the degassed cation conductivity measures the level of contaminants that turbine manufacturers are concerned about. Power condensate monitoring can help distinguish air in-leakage from...

 Open the catalog to page 3

DCC1000e System for Conductivity and Contamination Detection With accurate and reliable conductivity measurements using UniCond Sensors with ISM, the DCC1000e System confirms water purity to maximize power production and minimize corrosion. Conductivity – a key measurement Conductivity measurement is the method for determining the concentration of dissolved ionic species present in sample water. Ultrapure water is very poorly ionized with conductivity as low as of 0.055 uS/cm. Even if a small amount of contaminant (for example, NaCl) is added, the conductivity increases substantially. Such...

 Open the catalog to page 4

Optimized design to meet the challenge The DCC1000e System uses the latest technology in conductivity measurement to accurately and reliably measure conductivity at each stage of sample conditioning to give insight into the cycle chemistry (Figure 2). UniCond® Sensors used in the DCC1000e are the best available for the measurement of conductivity in pure waters, giving the greatest accuracy at the lowest levels of detection. With ISM® features such as having the measurement circuit close and inseparable from the sensor elements, the UniCond Sensor can deliver enhanced accuracy and greater...

 Open the catalog to page 5

DCC1000e System DCC1000e System Product Specifications 11.80' [300 mm] NOTE: 1. All dimensions are in inches (mm) unless otherwise noted. 1. Sample tubing: 1/4' OD 2. Sample inlet connection: 1/4' OD SS-316 bulkhead union 3. Sample outlet connection 1/2' MNPT UniCond Sensor with ISM Specifications Accuracy    ± 1% for 0.02-5,000 pS/cm; Temperature sensor    Pt 1000 RTD, IEC 60751, Class A, with NIST-traceable calibration Temperature accuracy    ± 0.1 °C at 25 °C Wetted materials    Titanium, PEEK Response time    90% of value in    <5 s Signal to transmitter    Digital (modified    RS485)

 Open the catalog to page 6

Electrical Specifications Current (analog) outputs    8 X 0/4 to 20 mA, 22 mA alarm, galvanically isolated from input and from earth/ground Analog output accuracy    ± 0.05 mA over 1 to 20 mA range Analog output configuration    Linear, bi-linear, logarithmic, autoranging Analog output load    500 Q max. Digital communication    USB, Type B connector User interface    Color touchscreen 5.7" Resolution 320 X 240 px 256 colors Update time (meas. update rate) 1 per second Hold input    Selectable Alarm control delay    Selectable, 0 to 999 s Connection terminal    Spring cage terminals...

 Open the catalog to page 7

DCClQQQe System with separate cooling input ports, 200-240 VAC    30 421 481 www.mt.com/thornton_ Visit for more information METTLER TOLEDO Group Process Analytics Division Local contact: www.mt.com/pro-MOs Subject to technical changes ©11/2017 METTLER TOLEDO. All rights reserved UniCond and ISM are registered trademarks of the METTLER TOLEDO Group. 58 087 103 Rev A 11/17