uniTOC
8Pages

{{requestButtons}}

Catalog excerpts

uniTOC - 1

TOTAL ORGANIC CARBON

Open the catalog to page 1
uniTOC - 2

UNITOC SYSTEMS THE PRECISE HIGH DEFINITION INSTRUMENTS FOR THE ONLINE AND OFF-LINE DETERMINATION OF TOC FROM DRINKING WATER TO ULTRA PURE WATER. Two innovative instruments uniTOC online and uniTOC lab combine high end analytical components according to DIN EN 1484. The graphical control with the 10” touchscreen display and the data management of the control and analysis software fulfill all international standards for the determination of TOC. The instruments are designed for monitoring i.e. process water in the pharmaceutical industry, microlelectronics, chemistry as well as the production...

Open the catalog to page 2
uniTOC - 3

uniTOC system calibration Monitoring low or high TOC concentrations TOC determination requires calibration which is completely automised. uniTOC lab works with multipoint calibration and autosampler with auto dilution to measure also higher concentration of TOC. Also the software allows the automised System Suitability Test (SST). The uniTOC online is made for pure and ultra pure water monitoring in industrial and pharmaceutical applications. The display shows the results of online-measurements of ultra pure water even with low TOC concentration. With uniTOC lab you will be able to measure...

Open the catalog to page 3
uniTOC - 4

• ALL FLUIDIC LINES OF THE INSTRUMENTS ARE MANUFACTURED BY MATERIALS OF HIGHEST PURITY LIKE PEEK, PFA OR PTFE WITH ALLMOST NO DEAD VOLUMES. • DOSAGE OF SAMPLES IS ACCOMPLISHED BY A CORROSION FREE PERISTALTIC PUMP. UNITOC ADVANTAGES • TWO INLETS OFFER THE MONITORING OF TWO DIFFERENT MEASURING POINTS OR MEASURING OF ONE GRAB SAMPLE DURING THE ON-LINE ANALYSIS. • UV RADIATION IS GENERATED BY A HIGH ENERGY, LOW PRESSURE UV LAMP, WHICH GUARANTEES A LIFESPAN OF MINIMUM 12 MONTHS. • THE CO 2 CONCENTRATION IS ME ASURED WITH A NDIR DETECTOR SYSTEM WHICH IS CO2 SELECTIVE AND INDEPENDENT FROM SAMPLE...

Open the catalog to page 4
uniTOC - 5

TOTAL ORGANIC CARBON ANALYZER

Open the catalog to page 5
uniTOC - 6

uniTOC-online • analysis mode UV oxidation with NDIR-detection TOTAL ORGANIC CARBON ANALYZER • working range • detection limit • analysis time TECHNICAL SPECIFICATIONS • sample inlet 2 separated Ports (3 different sample modes) • sample volume • enhanced oxidation with reagent (configurable) • carrier gas compressed air, nitrogen, oxygen • user interface (or key board with mouse) • data acces • IQ/OQ available • norms • power supply • degree of protection IP 45 • dimensions

Open the catalog to page 6
uniTOC - 7

uniTOC-lab • analysis mode UV oxidation with NDIR-detection • working range • detection limit • analysis time • sample inlet 1 port or auto sampler • sample volume multi point • enhanced oxidation with reagent (auto range) compressed air, nitrogen, oxygen • user interface • data acces 36 Sample with 30ml vials • power supply 400x250x600 mm (without auto sampler)

Open the catalog to page 7
uniTOC - 8

membraPure GmbH Wolfgang-Kuentscher-Str. 14, 16761 Hennigsdorf/Berlin, Germany +49 3302 - 201 20 0 info@membraPure.de www.membraPure.de Managing director Emil Flindt / Registration o ce Neuruppin district court / Trade register: HRB 9604NP WEEE-Reg.-Nr. DE 14696925 / VAT-ID number according to § 27a German VAT law: DE 160260388

Open the catalog to page 8

All MembraPure catalogs and technical brochures

  1. miniTOC

    8 Pages

  2. aracus

    20 Pages

  3. Aquinity² P10

    12 Pages

  4. IONUS

    20 Pages

  5. Astacus²

    12 Pages