Catalog excerpts
USER'S MANUAL
Open the catalog to page 1INDEX INDEX ------------------------------------------------------------------------------------------- 3 1.0 GENERAL POINTS--------------------------------------------------------------- 4 1.1 TECHNICAL CHARACTERISTICS ------------------------------------------------- 4 1.2 DISPLAY SIGNALS-------------------------------------------------------------------- 4 1.3 CONNECTION DESCRIPTIONS ---------------------------------------------------- 5 WIRING DIAGRAM FOR POTENTIOMETER INPUT ---------------------------- 7 2.0 INSTALLATION NOTES-------------------------------------------------------- 7...
Open the catalog to page 31.0 GENERAL POINTS The instruments of the MPP series in the P6 type container (75mm depth) use the potentiometer tranducer. The main characteristics are as follows: • two alarms with exchange relay output • alarm programming and 6 key calibration on instrument front panel for ease of use • 9999 point display • the 'hold' function operates directly from the terminal board ( with memory of value displayed ) and ' tara recovery ' (automatic zeroing). • possibility of programming delay time and hysteresis alarms using hidden menu which is password protected. • possibility of fixed zero setup •...
Open the catalog to page 41.3 CONNECTION DESCRIPTIONS KEYBOARD DESCRIPTION : alarm setup 1 : alarm setup 2 : increments blinking digit in menu or decimal point set-up : shifts blinking digit in menu : setup of reading scale start point : set up of reading scale end point + : menu access point
Open the catalog to page 5Terminals 4,5,6 Terminals 7,8,9 Terminals 13 and 15 agnd out mA out V rec. tara hold dgnd 2.5Vref InIn+ TERMINAL BOARD DESCRIPTION - alarm output 2 - alarm output 1 - instrument power supply (25Vac, 90÷260Vac/Vdc) in dc power supply, terminals 13 is positive and 15 negative Terminals 29 and 30 - input measure Terminal 28 - potentiometer power supply (MPP) Terminals 22, 23 and 24: - analog outputs (if requested) Terminal 26 - Hold: shortcircuiting terminals 26 and 27 will memorize the reading Terminal 25 - Recovery tara: 1) Terminals 25-27 open: the reading is the same as the programmed...
Open the catalog to page 6rec. tara hold dgnd 2.5Vref InIn+ WIRING DIAGRAM FOR POTENTIOMETER INPUT 2.0 INSTALLATION NOTES 2.1 INSTALLATION PROCEDURE 1- Undertake the linkups on page 5, 6, 7 2- Switch on the instrument 3- Use the “zero” key to set up the value that must coincide with the minimum input value (minimum potentiometer) 4- Use the “FS↵“ key to set up the value that must coincide with the maximum input value (maximum potentiometer) 5- Use the “5d.p.” key to set up the decimal point as required 6- Use the SET1 key to set up alarm value 1 7- Use the SET2 key to set up alarm value 2. 8- The instrument is now...
Open the catalog to page 7See the paragraph "Instrument Setup". There now follow examples of zero calibration which are also true for end scale and alarms using the personalized keys F.S., SET1 and SET2. Touch the zero key. The display will visualize the value which it keeps in memory with the first left hand digit Touch the " ^ " key to move the flashing flashing value. Confirm the programmed number by To program the decimal point observe the following indications. Touch the ^ key with the instrument in measuring mode. A decimal point will light up. In order to move this point touch the indicated key until the...
Open the catalog to page 8MPP P6 mect s.r.l. 2.3 RECOVERY TARA FUNCTION For “Recovery Tara” we mean a function that when in use zeroes the instrument readings. The Recovery Tara function with terminals 25 and 27 open is not in use and in these circumstances the instrument reading is linked to the programming parameters ISL, FSL and OFFS. At the moment when the terminals 25 and 27 are short circuited zero is displayed and the reading zeroed. This operation is indicated when the led 'Reset' lights up on the instrument front. Let us assume that the instrument, linked to a transducer, indicates a 100mm number. Shutting...
Open the catalog to page 9Table 2 NOTE n seq. Touch Written key on display 1 PASS FS ↵ + Touch FS ↵ + SET1 keys for some seconds SET1 2 0 000 Digit the personal password ** (confirm with " FS FS ↵ ↵" ) 3 CPAS CHANGE PASSWORD ( if required see paragraph ) 5 4 AbSP SET POINT IN USE 5 5 on on = Set1 and Set2 keys in use FS ↵ oFF = Set1 and Set2 keys out of use To change use "5" key and confirm with " FS ↵" 6 AbSP 7 AbOF ZERO AND END SCALE IN USE 5 8 on on = zero and FS keys in use; OFF = zero and FS FS ↵ keys out of use. To change use "5" key and confirm with " FS ↵" 9 AbOF 10 AbPd DECIMAL POINT IN USE 5 11 on on =...
Open the catalog to page 10NOTE Written on display 0000 Use this item to effect a zeroing that will be retained in memory. The number written can vary between -1999 and 1999. ** confirm with " FS ↵" OFFS LSPL LOWER LIMIT SETPOINT 0000 Setup the lower limit for the alarm value **(confirm with " FS ↵") LSPL LSPH UPPER LIMIT SETPOINT 0000 Setup the uppers limit for the alarm value **(confirm with " FS ↵") LSPH rL1 EXCHANGE RELAY 1 SET UP nA nA = normal open relay; nC = normal closed relay To change this use the "5" key and confirm with " FS ↵" rL1 HY1 HYSTERESIS ALARM 1 200 Set up required hysteresis (see Fig.2) with a...
Open the catalog to page 11NOTE Written on display Sd2 DELAY SELECTION RELAY 2 no no = no delay time; EC = switch on delay; di = switch off delay; ECdi = delay switch on + off . To change use the "5" key and confirm with " FS ↵" Sd2 dL2 DELAY TIME ALARM 2 20.0 Delay set up for period between 0 and 20.0 secs. ** (confirm with " FS ↵") dL2 ZEFI FIXED ZERO SELECTION on OFF=standard display; ON=fixed zero display . To change use the "5" key and confirm with " FS ↵" ZEFI SEL.A ANALOGUE OUTPUT SCALE 4 20 4 20= analogue ouput 4÷20mA; 0 20= analogue output 0÷20mA; 0 10= analogue output 0÷10V. To change use the "5" key and...
Open the catalog to page 12All MECT SRL catalogs and technical brochures
-
TPAC1008
22 Pages
-
TPAC1007
14 Pages
-
TPAC1006
22 Pages
-
MPA-MPV M6
42 Pages
-
MPA386 M6
39 Pages
-
MPA-MPV P6
18 Pages
-
MP40 P6 Simple imput
16 Pages
-
MP40 P6 Multi input
17 Pages
-
MPX28
8 Pages
-
MPSN20
2 Pages
-
MPCV015
20 Pages
-
MPT390 M6
61 Pages
-
MPT91 M1
66 Pages
-
MPT90 M1
46 Pages
-
MPT60 M1
45 Pages
-
MPT100 M1
35 Pages
-
MPCIB396 P6
41 Pages
-
MPCIB20 M1
25 Pages
-
MPCT301 P6
53 Pages
-
MPCT300 P6
37 Pages
-
MPCT30 P6
36 Pages
-
MPCT20 M1
35 Pages
-
MP2200 M6
44 Pages
-
MP1200 P6
36 Pages
-
MPPV340 M6
40 Pages
-
MPO347 M6
40 Pages
-
MPP323 M6
42 Pages
-
MPPV010 P6
40 Pages
-
MP45 P6
18 Pages
-
MPM P6
34 Pages
-
MP30 P2
15 Pages
-
MP20 M1
35 Pages
-
MPNC030
12 Pages
-
TPLC005
28 Pages
-
MPNC020
13 Pages
-
MPNC010
25 Pages
-
TP1070
15 Pages
-
TP1057
15 Pages