MPM P6
1 /34Pages

MPM P6

MPM P6
1 /34Pages

Catalog excerpts

MPM P6-3

INDEX INDEX ------------------------------------------------------------------------------------------- 3 1.0 GENERAL POINTS--------------------------------------------------------------- 4 1.1 TECHNICAL CHARACTERISTICS------------------------------------------------- 4 1.1.1 DISPLAY SIGNALS --------------------------------------------------------------- 4 1.2 CONNECTION DESCRIPTIONS ---------------------------------------------------- 5 1.2.1 WIRING DIAGRAM FOR CURRENT INPUT -------------------------------- 7 1.2.2 WIRING DIAGRAM FOR VOLTAGE INPUT -------------------------------- 8 1.3 PROGRAMMING...

 Open the catalog to page 3
MPM P6-5

1.0 GENERAL POINTS The MPM series instrument in the P6 housing (depth 75mm) has 2 analogue inputs and can show one input or the sum, differance, product or ratio of the two. The instrument can be ordered for 2 current (0÷20mA or 4÷20mA ) or voltage (0÷10V) inputs. The main characteristics are as follows: • two alarms with exchange relay output (3 if 220V power supply) that can be assigned to display, channel 1, channel 2, sum, difference, product or ratio. • 16V not regulated for 2 or 3 wire transducer power supply • alarm programming and calibration by 6 key on instrument front panel for easy...

 Open the catalog to page 5
MPM P6-6

KEYBOARD DESCRIPTION ' : alarm setup 2. Into the menu it fulfils the exit function. d p : increments blinking digit in menu or decimal point set-up 1: shifts blinking digit in menu. If "dISP = CAnl / CAn2" toggles between 1: setup of reading scale start point : set up of reading scale end point + : menu access point

 Open the catalog to page 6
MPM P6-7

TERMINAL BOARD DESCRIPTION serial outputs - transducer power supply (16V) - instrument power supply (Vcc, Vpt, 24Vac, 90-=-260Vac). DC power supply, terminal 13 is positive and terminal 15 is negative. - analogue output (if requested) - Hold: shortcircuiting terminals 16 and 18 will - Recovery tara: 1) Terminals 16-17 open: the reading is the same as the programmed values 2) Terminals 16-17 closed: at the moment the terminals become shortcircuited the display is zeroed (rec.tara) If requested serial output (the "hold" and "Rec. tara" functions is not available): Terminal 16: see paragraph "serial...

 Open the catalog to page 7
MPM P6-8

1.2.1 WIRING DIAGRAM FOR CURRENT INPUT 1) Connection for 2 wire transducer 2)Connection for 2 wire transducer with external power supply

 Open the catalog to page 8
MPM P6-9

1.2.2 WIRING DIAGRAM FOR VOLTAGE INPUT 1) Connection for 3 wire transducer 2) Connection for 3 wire transducer with external power supply 3) Connection for voltage input

 Open the catalog to page 9
MPM P6-10

I ^+2 = numer relevant to use table keys and reach the program If you want to reach an item, use the To skip into the item press If you want to change a number, use the I_3 key to increment the blinking digit, and the I ^ 1 key to shift the blinking digit. To exit from the menu, press modified parameters will be stored.

 Open the catalog to page 10
MPM P6-12

analog output serial output 1- Execute the connections as indicated at 6, 7, 8 pages 3- Get into the programming menu to set the type of visualization required (by the "SCAL" menu item) and the set up of the two inputs (ISI1, ISL1, FSI1, FSL1 e 4- The "zero" key is enabled just for the visualization of the single input (canl and can2). Using the "zero" key it is possible to set the reading value which must be the same of the minimum input (0 for the 0-10V and 0-20mA scales, 4 for the 4- 20mA scale to change this value see paragraph "Instrument configuration".

 Open the catalog to page 12
MPM P6-13

MPM P6 mect s.r.l. 5- The “F.S.” key is enabled just for the visualization of the single inputs (cAn1 and cAn2). Using this key it is possible to set the reading value which must be the same of the input “fondo scala” (10V for the 0-10V scale, 20mA for the 0-20mA and 4-20mA scales to change this value see paragraph “Instrument configuration”). 6- Use the key to set up the decimal point as required. key to set up alarm value 2 8- Use the 9- The instrument is now ready for use. 10- For automatic zeroing check the recovery Tara function. 11- To install other input parameters or alarms see the instrument...

 Open the catalog to page 13
MPM P6-14

mect s.r.l. 2nd OPERATION Touch the “4” key to move the flashing digit to the right. 3rd OPERATION Touch the “5” key to increase the flashing number. 4th OPERATION Confirm the programmed number by touching the 'zero' key. To program the decimal point observe the following indications. Al1 Al2 In1 In2 Touch the “5” key with the instrument in measuring mode. A decimal point will light up. In order to move this point touch the indicated key until the desired point is reached. 2.3 RECOVERY TARA FUNCTION Function enabled only for 0-10V and 4-20mA inputs. For “Recovery Tara” we mean a function that...

 Open the catalog to page 14
MPM P6-15

MPM P6 mect s.r.l. Let us assume that the instrument, linked to a transducer, indicates a 100kg number. Shutting down the connection between terminals 16 and 17 will zero the instrument reading , thus rendering the indications negative for weight values that are less than 100 Kg and positive for values that are greater than 100 Kg. N.B. The recovery tara function is used when it is necessary to zero the instrument reading continuously as the data contained in the instrument memory will not be retained after instrument switch off. To retain in the permanent memory the zeroed data you should use...

 Open the catalog to page 15
MPM P6-16

MPM P6 mect s.r.l. Table 2 n seq. Touch Written NOTE key on display PASS 1 FS ↵ + Touch FS ↵ + SET1 keys for some seconds. SET1 0 000 Digit the personal password ** (confirm with " FS 2 FS ↵ ↵" ) CPAS CHANGE PASSWORD ( if required see paragraph ) 3 5 AbSP SET POINT IN USE 4 5 on on = Set1 and Set2 keys in use 5 FS ↵ off = Set1 and Set2 keys out of use. To change use "5" key and confirm with " FS ↵" AbSP 6 SCAL SELECTION INPUT CURRENT (only for MPM 157 22 instrument) 4 20 4 20 = 4-20mA input, 8 FS ↵ 0 20 = 0-20mA input. To change use "5" key and confirm with " FS ↵" SCAL 9 dISP SELECTION READ...

 Open the catalog to page 16

All MECT SRL catalogs and technical brochures

  1. TPAC1008

    22  Pages

  2. TPAC1007

    14  Pages

  3. TPAC1006

    22  Pages

  4. MPA-MPV M6

    42  Pages

  5. MPA386 M6

    39  Pages

  6. MPP P6

    19  Pages

  7. MPA-MPV P6

    18  Pages

  8. MPX28

    8  Pages

  9. MPSN20

    2  Pages

  10. MPCV015

    20  Pages

  11. MPT390 M6

    61  Pages

  12. MPT91 M1

    66  Pages

  13. MPT90 M1

    46  Pages

  14. MPT60 M1

    45  Pages

  15. MPT100 M1

    35  Pages

  16. MPCIB396 P6

    41  Pages

  17. MPCIB20 M1

    25  Pages

  18. MPCT301 P6

    53  Pages

  19. MPCT300 P6

    37  Pages

  20. MPCT30 P6

    36  Pages

  21. MPCT20 M1

    35  Pages

  22. MP2200 M6

    44  Pages

  23. MP1200 P6

    36  Pages

  24. MPPV340 M6

    40  Pages

  25. MPO347 M6

    40  Pages

  26. MPP323 M6

    42  Pages

  27. MPPV010 P6

    40  Pages

  28. MP45 P6

    18  Pages

  29. MP30 P2

    15  Pages

  30. MP20 M1

    35  Pages

  31. MPNC030

    12  Pages

  32. TPLC005

    28  Pages

  33. MPNC020

    13  Pages

  34. MPNC010

    25  Pages

  35. TP1070

    15  Pages

  36. TP1057

    15  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.