1. Catalogs
  2. Meatest spol. s r.o.
  3. MTE Laboratory Impedance Standards

MTE Laboratory Impedance Standards

MTE Laboratory Impedance Standards

MTE Laboratory Impedance Standards

Product catalog summary
Highlights
- Resistance and Capacitance standards are available in four package types for various applications.
- Custom nominal values and terminal configurations can be requested.
- Designed for calibration of ohmmeters, LCR meters, and insulation testers.
- Can be used as transfer standards for maintaining traceability between laboratories.
Description
The MTE series offers resistance and capacitance standards suitable for calibration purposes. They are available in four terminal configurations to accommodate different applications, including OPEN and SHORT standards for zero calibration of LCR meters. Custom configurations are available upon request.
Package Types
  • HV: For megohmmeters and insulation testers, featuring 3x banana sockets with extra insulation.
  • A: For ohmmeters and multimeters, featuring 4x banana sockets.
  • B: For LCR bridges with wired connections, featuring 4x female BNC connectors.
  • C: For LCR bridges with direct connections, featuring 4x male BNC connectors.
Specifications
General Data
- Reference temperature: +21 °C to +25 °C
- Operating temperature: 0 °C to +50 °C
- Dimensions: 125 x 60 x 105 mm
Resistance Standards
- Construction: Foil resistor up to 1 MΩ, Ceramic resistor above 1 MΩ
- Frequency range: DC to 20 kHz
- Reference voltage: 1 kV for 10 MΩ to 10 GΩ range
- Voltage coefficient: 0.5 ppm/V (typical) for 10 MΩ to 10 GΩ range
Nominal Values and Uncertainty
  • 100 mΩ: Max deviation 0.1%, Uncertainty 0.05%, Temp. coefficient 10 ppm/K, Rating 3 W
  • 1 Ω to 1 MΩ: Max deviation 0.01%, Uncertainty 0.005%, Temp. coefficient 1 ppm/K, Rating varies
  • 10 MΩ to 10 GΩ: Max deviation varies, Uncertainty varies, Temp. coefficient 100 ppm/K, Rating varies
Capacitance Standards
- Construction: Multi-layer mica capacitor up to 100 nF, PP film capacitor above 100 nF
- Maximum voltage: 30 Vrms
Nominal Values and Uncertainty
  • 10 pF: Max deviation 2%, Uncertainty 0.1%, Temp. coefficient 30 ppm/K, Dissipation factor < 0.005
  • 100 pF to 100 µF: Max deviation varies, Uncertainty varies, Temp. coefficient 30 ppm/K, Dissipation factor varies
See more

Catalog excerpts

MTE Laboratory Impedance Standards-1

Impedance standards Resistance and Capacitance Four package types for various applications Custom nominal values and terminals upon request Low time constant MTE series resistance and capacitance standards are designed for calibration of ohmmeters, LCR meters and insulation testers. Alternatively, MTE standards can be used as transfer standards, keeping traceability between primary and secondary laboratories. MTE standards come in four terminal configurations to fit every application. OPEN and SHORT standards for zero calibration of LCR meters as well as custom nominals and terminal configurations are available upon request. Package types Package type megohmmeters, insulation testers 3x banana socket with extra insulation (2W measurement + case ground) ohmmeters, multimeters LCR bridges – wired connection LCR bridges – direct connection Zelezna 3 | 619 00 Brno | Czech Republic | 420 543 250 886 | [email protected] | www.meatest.com

 Open the catalog to page 1
MTE Laboratory Impedance Standards-2

SPECIFICATION Specifications below describe 1-year absolute accuracy of this product including long-term stability, linearity and reference standard measurement uncertainty as well as ambient conditions within specified limits. Reference temperature up to 1 MΩ: Foil resistor above 1 MΩ: Ceramic resistor Frequency range Available packages Operating temperature GENERAL DATA up to 1 MΩ: A, B and C above 1 MΩ: HV; DC applications only Reference voltage Voltage coefficient MTE RP Resistance Standards Max. deviation Temperature coefficient Nominal value Maximum dissipation power at 23 °C or maximum...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.