1. Catalogs
  2. Meatest spol. s r.o.
  3. M550 Impedance Calibrator

M550 Impedance Calibrator

M550 Impedance Calibrator

M550 Impedance Calibrator

Product catalog summary
Overview
The M550 Impedance Calibrator is designed for the calibration of LCR meters, offering automated calibration capabilities with a frequency range from 20 Hz to 1 MHz. It supports four-terminal pair (4TP), four-wire (4W), and two-wire (2W) configurations, and includes open, short, and load compensation to mitigate test cable influences.
Specifications
  • Resistance: Ranges from 0.1 Ω to 100 MΩ in 4TP and 4W modes, and 1 Ω to 10 MΩ in 2W mode. Calibration uncertainty varies by resistance value and configuration.
  • Capacitance: Ranges from 10 pF to 100 µF in 4TP mode, with a tolerance of 5% at 1 kHz.
  • Inductance: Ranges from 10 µH to 10 H in 4TP mode, with a tolerance of 15% at 1 kHz.
Key Features
  • Automated LCR bridge calibration with remote control via RS-232 and GPIB interfaces.
  • Calibration data displayed on a large LCD screen, including voltage, current, and frequency readouts.
  • Supported by CALIBER/WinQbase automated calibration software.
General Data
  • Warm-up time: 15 minutes.
  • Operating temperature: +15 °C to +30 °C.
  • Power supply: 115/230V - 50/60 Hz, 35 VA max.
  • Dimensions: 450 x 150 x 430 mm; Weight: 12 kg.
See more

Catalog excerpts

M550 Impedance Calibrator-1

Impedance Calibrator Automated LCR bridge calibration Calibration data from 20 Hz to 1 MHz Four-terminal pair, 4W and 2W configurations Open, Short and Load compensation Model M550 Impedance calibrator is designed for calibration of LCR meters. The calibrator contains stable and temperature independent resistance standards, partial capacitance standards and partial inductance standards simulated by RC T-type passive network. Open and Short reference positions are available to eliminate influence of test cables. Calibration values of both complex parameters of partial standards are displayed on large LCD screen in preselected pairs of parameters as well as voltage, current and frequency readouts from built-in test signal meter. Philosophy of M550 is based on remote control and automated calibration. For this reason, the calibrator is equipped with RS-232 and GPIB interfaces and supported in CALIBER/WinQbase automated calibration software. Zelezna 3 | 619 00 Brno | Czech Republic | 420 543 250 886 | [email protected] | www.meatest.com

 Open the catalog to page 1
M550 Impedance Calibrator-2

SPECIFICATION Specifications below describe 1-year absolute uncertainty at a confidence interval of 95%, including long-term stability, linearity, load and line regulation and reference standard measurement uncertainty as well as ambient conditions within specified limits. Resistance range summary Frequency range Available parameter pairs Z/ɵ, Y/ɵ, Rs/Ls, Rs/Cs, Rp/Cp, Rp/Lp, R/X, G/B Series resistance (Rs) parameter overview Nominal value 1 year typical stability Temperature coefficient Capacitance range summary Frequency range Available parameter pairs Parallel capacitance (Cp) parameter overview...

 Open the catalog to page 2
M550 Impedance Calibrator-3

Inductance range summary Frequency range Temperature coefficient Available parameter pairs Series inductance (Ls) parameter overview Nominal value Typical series resistance Rs Inductance is simulated in 4TP mode using T-network RC circuit Test level meter Voltage measurement Current indication Warm-up time Reference temperature Operating temperature Storage temperature Output terminals 4TP mode: 4 BNC connectors 4W & 2W modes: 4 banana sockets Power supply GENERAL DATA Frequency measurement LCR bridge calibration (application) Zelezna 3 | 619 00 Brno | Czech Republic | 420 543 250 886 |

 Open the catalog to page 3
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.