9010

9010

9010

Product catalog summary
Highlights
  • AC/DC voltage/current capabilities up to 1050 V/20.5 A with a basic uncertainty of 35 ppm.
  • Supports AC/DC power, energy, resistance, capacitance, frequency, thermocouples (TC), and resistance temperature detectors (RTD).
  • Scope option available up to 400 MHz and high voltage resistance option for 1.5 kV insulation testers.
  • Built-in process multimeter and multiple interface options including RS232, LAN, USB, and GPIB.
Description

The 9010 Multifunction Calibrator is a versatile tool designed for electrical calibration laboratories. It is capable of calibrating a wide range of devices such as multimeters, clamp meters, ohm meters, power meters, energy meters, transducers, insulation testers, and scopes. It features high load capacity for both voltage and current outputs, allowing calibration of high-consumption analogue meters. The calibrator supports harmonic and non-harmonic signal testing and is compatible with Meatest's CALIBER/WinQbase software for automated calibrations.

Specifications
  • General Data: Warm-up time of 30 minutes, operating temperature range of +13 °C to +33 °C, and storage temperature range of -10 °C to +55 °C. Power supply is 115/230V - 50/60 Hz, 450 VA max.
  • DC/AC Voltage: Voltage range from 0 mV to 1050 V with various internal ranges and frequency accuracy of 25 ppm.
  • DC/AC Current: Current range from 0.0000 µA to 20.50000 A with frequency accuracy of 25 ppm.
  • Resistance: Resistance range from 0.0000 Ω to 1.100000 GΩ with 2W and 4W modes.
  • Capacitance: Capacitance range from 0.800000 nF to 120.0000 mF in 2W mode.
  • Temperature (RTD, TC): Supports various RTD and thermocouple types with manual or automatic cold junction compensation.
  • AC/DC Power & Energy: Power range from 40 µW to 5.74 kW with total uncertainty based on voltage, current, phase shift, and energy period specifications.
Options
  • Frequency/Scope Option: HF mode with amplitude range from 1.400 mVpk to 1.5000 Vpk and frequency range up to 400 MHz.
  • High Voltage Resistance Option: Maximum test voltage up to 1500 Vdc with resistance uncertainty varying by range.
  • Multimeter Option: Includes measurement functions for DC voltage, DC current, frequency, resistance, and temperature with specified uncertainties.
See more

Catalog excerpts

9010-1

Multifunction Calibrator AC/DC voltage/current up to 1050 V/20.5 A Basic uncertainty 35 ppm AC/DC power, energy, resistance, capacitance, frequency, TC, RTD Scope option up to 400 MHz High voltage resistance option for 1.5 kV insulation testers Built-in process multimeter Interface RS232, LAN, USB, GPIB Multifunction calibrator 9010 is designed as universal calibration tool for electrical calibration laboratories, covering most of their workload like multimeters clamp meters, ohm meters, power meters and power analyzers, energy meters, transducers, insulation testers, process meters, scopes and many others. High load capacity of both voltage (up to 50 mA) and current output allows for calibration of high-consumption analogue meters. Installed harmonic and non -harmonic shape signals allow for testing meter sensitivity to distorted signals by a signal with various crest factor. Advancing from previous M14x calibrator series, 9010 can now calibrate even 400 MHz scopes, 1.5 kV insulation testers and 287 kW power meters. On the other hand we kept all the popular functions including complete transducer and external sensor calibration (strain gauge, pressure, torsion, strength, etc.) using built-in multimeter, automatic uncertainty calculation, remote control and easy recalibration. 9010 calibrator is fully compatible with Meatest calibration SW package CALIBER/WinQbase which allows for time saving automated calibrations using any of the four installed remote control interfaces. Zelezna 3 | 619 00 Brno | Czech Republic | 420 543 250 886 | [email protected] | www.meatest.com

 Open the catalog to page 1
9010-2

SPECIFICATION Specifications below describe 1-year absolute uncertainty at a confidence interval of 95%, including long-term stability, linearity, load and line regulation and reference standard measurement uncertainty as well as ambient conditions within specified limits. Warm-up time Reference temperature Operating temperature GENERAL DATA Storage temperature 10 % of accuracy / °C outside Tref Power supply Voltage range summary DC: 0 mV – 1050 V AC sine: 1 mVrms – 1050 Vrms Non-sine: 1 mVrms – 200 Vrms Internal ranges DC/AC Voltage Temperature coefficient Frequency accuracy and resolution Non-sine...

 Open the catalog to page 2
9010-3

DC/AC Current Current range summary DC: 0.0000 µA – 20.50000 A AC Sine: 10.0000 µArms – 20.50000 Arms Non-sine: 100.0000 µArms – 2.000000 Arms Internal ranges Frequency accuracy and resolution Non-sine waveform types saw, triangle, square, truncated sin; 1kHz max. Non-sine amplitude uncertainty Ranges, resolution, 1 year uncertainty [% of value] Range Distortion and Load Characteristics Parameter THD in bandwidth up to 100 kHz Additional uncertainty for compliance voltage above 0.5 Vrms Voltage from current Voltage range Amplitude uncertainty Current coil (option 140-50) Source impedance Applicable...

 Open the catalog to page 3
9010-4

Resistance range summary 2W and 4W continuous range 2W and 4W fixed decadic standards 100 GΩ High Voltage Resistance (optional) Basic resistance modes and 1 year uncertainty [ppm of value + absolute] Continuous range mode Nominal standard value Capacitance range summary 2W continuous range 2W fixed decadic standards Capacitance modes, 1 year uncertainty and frequency limits Continuous range mode Nominal standard value Thermocouple types TC cold junction compensation Manual or automatic with adapter 91

 Open the catalog to page 4
9010-5

AC/DC Power & Energy Range summary power: voltage: current: frequency: time period: Total uncertainty based on voltage, current, phase shift and energy period specifications. Phase shift uncertainty Energy period uncertainty Additional features Harmonic distortion, voltage from current, current coil multiplication Total 1 year uncertainty overview [% of value] Current range Harmonic distortion (all AC functions) Fundamental harmonic uncertainty amplitude: frequency: phase shift: Frequency range Harmonic product amplitude range Harmonic product phase shift unc. Frequency / Scope option HF mode...

 Open the catalog to page 5
9010-6

High Voltage Resistance option Maximum test voltage Resistance uncertainty Test voltage uncertainty Multimeter option Measurement function Using 9000-60 4W measurement adapter (comes as standard with MER option) Zelezna 3 | 619 00 Brno | Czech Republic | 420 543 250

 Open the catalog to page 6
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.