KA-R series

KA-R series

KA-R series

Product catalog summary
Overview: The KA-R series is an ultra-low ripple, highly stable high voltage (HV) power supply designed for Scanning Electron Microscopy (SEM) and Mass Spectrometry (MASS) applications. It features reversible polarity controlled by an external signal, making it versatile for various scientific applications.
Specifications: The power supply offers output voltages ranging from ±10kV to ±30kV with a high stability of 0.001% over 5 minutes. It has an ultra-low ripple noise and provides output voltage and current monitoring. The input voltage required is +24Vdc ±1V, and it supports external control voltage from 0 to 10V.
Performance: The device ensures line regulation and load regulation at ±0.001% of full scale for input changes and load changes, respectively. It has a temperature coefficient of 25ppm/°C and stability of 10ppm/5min after a 3-hour warm-up.
Protection and Environment: The KA-R series includes overvoltage and short circuit protection. It operates within a temperature range of 0 to 50°C and can be stored between -20 to 60°C. The humidity range for operation is 20 to 80% RH without condensation.
Options and Accessories: The series offers options like the D-Sub connector assignment and additional control connectors. Accessories include a high voltage shielded cable and control connectors like -LTc10 and -LX.
Warranty: The products come with a one-year warranty against defects in material and workmanship under normal use. However, this warranty excludes applications requiring extremely high reliability and safety, such as nuclear power or medical equipment.
Contact Information: The document provides contact details for various offices, including those in San Jose, New York, Dallas, Boston, and the international office in Osaka, Japan.
See more

Catalog excerpts

KA-R series-1

y? Matsusada Prcision Rversible Polarity & High Stability NEW KA-R ■ -±10kV to ±30kV R驩versible Polarity- > Designed for SEM.MASS applications > High Stability 0.001%/5min > Ultra Low ripple noise > Output polarity change by external signal > Output voltage, current monitor APPLICATIONS * SEM * MASS Spectrometry SUMMARY KA-R series is an ultra low ripple, highly stable HV power supply. It is the perfect choice for SEM and MASS applications. The output polarity can be controlled with an external signal. LINEUP Output Voltage (kV) Output Current Output Power (W) Model Ripple (mVp-p) Input Current (A typ.) 0 to ±10 1mA 10 KA-10R1 30mV 1 10mA 100 KA-10R10 500mV 6 0 to ±20 400|jA 8 KA-20R0.4 70mV 1 1.5mA 30 KA-20R1.5 150mV 2 5mA 100 KA-20R5 1V 6 0 to ±30 250|jA 7.5 KA-30R0.25 100mV 1 3.3mA 100 KA-30R3.3 2V 6 SPECIFICATION Input Voltage Output Voltage Control Reverse Polarity Line regulation Load regulation Stability Temperature coef Output monitor HV Output cable Protection Temperature Range Humidity Accessory +24Vdc ±1V External control voltage 0 to 10V(Vcon-in) TTL(Reversing time 90% to 90% 1.2sec typ.) ±0.001% of F.S. for ±1V input change ±0.001% of F.S. for 100|jA to 100% load change 10ppm/5min, 50ppm/Hr(After 3Hr warm-up) 100ppm/Hr(After 1/2Hr warm-up) 25ppm/°C Voltage: 0 to 10Vdc Current: 0 to 10Vdc HV shielded cable 1m Overvoltage and short circuit protection Operating: 0 to 50°C Storage: -20 to 60°C 20 to 80%RH(no condensation) Control connector ID-Sub CONNECTOR ASSIGNMENT - 1 2 3 4 5 6 7 8 - O DDDDDDDD v D D □ □ □ □ □ I O 9 10 11 121314 15 Ref. +10V Output Voltage Control 0 to 10V Internal Potentiometer* Internal Potentiometer* Output Voltage Monitor 0 to 10V Reverse Polarity TTL"Hi" is Negative TTL"Lo" is Positive 7 COM 8 + 24Vdc input return 9 COM 10 Polarity status (+) 11 Polarity status (-) 12 HV ON/OFF(GND=ON, OPEN=OFF) 13 Output Current Monitor 0 to 10V 14 COM 15 + 24Vdc input *Switch to internal potentiometer by shorting Pin 3 and Pin4 OPTION -LTc10 Temperature coef 10ppm/°C -LX Control connector Molex 12-pin Base plate 200W x 240L(mm) (20kV, 30kV) 200W x 155L(mm) (10kV) *Add 1-LTc101 "-LX" to the Model number at time of order.(i.e. KA-10R1-LTc10X)

 Open the catalog to page 1
KA-R series-2

DIMENSIONS inch(mm) 10kV 100W(standard) 20kV,30kV (standard) 10kV 100W(-LX) 20kV.30kV (-LX) Side Output cable © ® © ®,- -/© © © © Input side Top 9.06(230) Output side GND(M4) 0.20(5) Input side Bottom 15PIN D-sub(male) 5) 5) CD O) .4 .8 8. 6. 5) (21 6( .4 8. co O) O CO 4-04.5 8.07(205) 0.79(20) 9.45(240) 0.59(15) 12.01(51). Side 3.76(95.5) 1 3.76(95.5) |j.93(49)j] 0.59(15) © © 1 © ©/- -/© © © © 8.07(205) /GND(M4)/ M3 X 8mm max 10kV 10W (standard) 10kV 10W (-LX) Side © © _©__©/- -'® © © © Output cable Top Output side GND(M4) Input side 0) 6) 09 oj t- .8 .7 . 5.12 130) ; nput side Bottom Output...

 Open the catalog to page 2
KA-R series-3

USA/canada : 1-888-652-8651 other countries : 1-81-77-561-2112 (HV modules) Please copy this page and above fax number after filling out form below. I would like □ A quotation Q An explanation of product □ A demonststration □ To purchase □ Other ( ) Give us your requirement / comment Please fill in below. Address: Company: Dept.:_ Title: Name: Tel: Fax: E-mail: Warranty We warrant that products contained in this catalog (hereinafter, the "Products") are free from defects in material and workmanship under normal use for a period of one (1) year from the date of shipment thereof. However, the warranty...

 Open the catalog to page 3

All Matsusada catalogs and technical brochures

  1. W series

    3  Pages

  2. RC series

    2  Pages

  3. TP series

    1  Page

  4. TZ series

    2  Pages

  5. TG series

    2  Pages

  6. TF series

    1  Page

  7. PZJ series

    2  Pages

  8. XWR series

    2  Pages

  9. X series

    1  Page

  10. XR series

    1  Page

  11. Xg Series

    4  Pages

  12. AMP series

    4  Pages

  13. DRA series

    5  Pages

  14. DOS series

    8  Pages

  15. VOL Series

    12  Pages

  16. ECA series

    2  Pages

  17. EJC series

    2  Pages

  18. TA/TC Series

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.