1. Catalogs
  2. MANVIA
  3. DEGASSED CATION CONDUCTIVITY DG 105

DEGASSED CATION CONDUCTIVITY DG 105

DEGASSED CATION CONDUCTIVITY DG 105

DEGASSED CATION CONDUCTIVITY DG 105

Product catalog summary
Description: The DG-105 is a degassed cation conductivity device designed to measure the conductivity of a sample by eliminating dissolved gases. The process involves boiling the sample to release gases, cooling it to 25-30°C, and then measuring conductivity. This parameter is crucial for boilers, as manufacturers require steam cation conductivity to be below 0.5 microsecond/cm to prevent damage.
Applications: The device is used in steam-water analysis systems and manual sampling systems.
Features: The DG-105 is a complete unit with necessary components, ready for use with simple piping and plug-in. It includes shell protection to prevent hot injuries, a compact wall mount design, and a sample cooler to ensure consistent sample temperatures for probes, reducing errors.
Specifications: The device operates at a pressure rating of 4 bar(g) with a sample flow of 50 to 250 cc/min. It has specific sample and cooling water connections, consumes 2 l/min of cooling water, and requires an input power of 110-120/220-240 VAC, 50/60 Hz with a power consumption of 1.6 KVA. It is mounted on stainless steel.
Spare Parts: Available parts include heating cartridges for 120 VAC and 230 VAC, and a solid-state relay for 110-230 VAC, 50/60 Hz.
Main Elements: The device includes a mounting plate, temperature controller, isolating valve, rotameter with low flow alarm, SC, CC, and DCC sensors, transmitter, sample cooler, electrical box, resin column, heater, and thermocouple.
Dimensions: Detailed drawings are available for reference, indicating the layout and dimensions of the device components.
See more

Catalog excerpts

DEGASSED CATION CONDUCTIVITY DG 105-1

DEGASSED CATION CONDUCTIVITY DG 105 DESCRIPTION The cation conductivity degasser DG-105 is a device designed to obtain the value of the conductivity of a sample, eliminating the contribution to it of possible gases that may be dissolved. After subjecting the sample to boiling, releasing any gas can be dissolved, a cooling is performed at 25-30 ° C, and then the treated sample to a conductivity cell is introduced. Degassed cationic conductivity is an important parameter for the boiler, most boiler manufacturers have specified that the value of the cation conductivity in the steam must be below a certain limit of conductivity (less than 0.5 microsecond / cm), to avoid damage to it or the turbine. APPLICATIONS • Steam-water analysis systems. • Manual sampling systems. FEATURES • Including all necessary components as a complete unit, just piping, plug-in and ready to use. • Provide with necessary shell protection to prevent operators from hot injury. • Compact wall mount design. • Sample cooler ensures CC and DCC probes receive identical sample temperatures, eliminating errors. DEGASSED CATION CONDUCTIVITY

 Open the catalog to page 1
DEGASSED CATION CONDUCTIVITY DG 105-2

DEGASSED CATION CONDUCTIVITY DG 105 SPECIFICATIONS Pressure Rating: Sample Flow: Sample Connection: Sample Cooler Type: Cooler Cooling Water Connections: Cooling Water Consumption: Input Power: Power Consumption: Plate mounting : Protection No Flow 0.- None 1.- Switch protection No Flow Electrical supply F.- 120 VAC 60 Hz G.- 230 VAC 60 Hz SPARE PARTS Part # H1500-120 H-1500-230 RD-105 www.analyzersmanvia.com Poligono Riano III, Parcela 43 33920 Riano, Asturias, Spain Phone:+34 984204549. Email: [email protected]

 Open the catalog to page 2
DEGASSED CATION CONDUCTIVITY DG 105-3

DEGASSED CATION CONDUCTIVITY DG 105 DIMMENSIONS Drawings according to Ref. DG-105-D-1-1-F/G in mm MAIN ELEMENTS 1.- Mounting Plate 2.- Temperature controller 3.- Isolating valve 1/4” 4.- Rotameter with low flow alarm 5.- SC Sensor 6.- CC Sensor 7.- DCC Sensor 8.- Transmitter 9.- Sample Cooler 10.-Electrical box 11.-Resin column 12.- Heater & Thermocouple www.analyzersmanvia.com Poligono Riaño III, Parcela 43 33920 Riaño, Asturias, Spain Phone:+34 984204549. Email: [email protected] DEGASSED CATION CONDUCTIVITY DG 105. S

 Open the catalog to page 3
DEGASSED CATION CONDUCTIVITY DG 105-4

DEGASSED CATION CONDUCTIVITY DG 105 DIMMENSIONS Drawings according to Ref. DG-105-A-0-1-F/G in mm MAIN ELEMENTS 1.- Mounting Plate 2.- Temperature controller 3.- Isolating valve 1/4” 4.- Rotameter with low flow alarm 5.- Sample Cooler 6.-Electrical box 7.- Heater & Thermocouple www.analyzersmanvia.com Poligono Riaño III, Parcela 43 33920 Riaño, Asturias, Spain Phone:+34 984204549. Email: [email protected] DEGASSED CATION CONDUCTIVITY DG 105. S