SP2500-H/C/I/BB - 4 Pages

Add to MyDirectIndustry favorites
SP2500-H/C/I/BB

Catalog excerpts

Embracing Challenge Gas Sample Probe Series SP® Version SP2500-H, SP2500-H/C/I/BB, SP2500-H/C/I/BB/F Special Features Extractable sample tube or prefilter, without dismounting the probe I ntegrated back purge possibility with closing off the sample gas outlet Optimum operational reliability Universal applicability Adaption to nearly all process condi- tions due to its compact and modular design Easy mounting Problem-free maintenance Application The gas sample probes M&C type SP2500..., based on version SP2000-H (see data sheet 2.2.2), are used for continuous ­ as sampling g in processes with gases of high dust content, high temperatures and/or high humidity. They make it possible to remove e.g. for cleaning purposes the preliminary filter or the sample tube out of the process without dismounting the complete gas sample probe. The special design of the heating element of the SP2500... (with protective hood) ­ ermits p controlled heating of the complete filter housing, including the mounting flange up tp 180 °C. This ensures reliable operation e ­ xternal to the process preventing the temperature ­ alling f The probes type SP2500-H/C/I/BB and SP2500-H/C/I/BB/F with special back purge possibility are used in case of very high dust content. For this purpose, they are equipped with an additional back purge valve and a pneumatic check valve in the sample gas outlet. The temperature of the standard probe is controlled by an integral compact design capillary sensor ­ hermostat with an excess t temperature limit switch and an alarm function for ­ emperature failure. The probe t SP2500-H provides the possibility to feed the test gas /C optionally via a check valve. Additional functions of the probe SP2500H/C/I/BB(/F): Small stagnant space Description The M&C gas sample probes are designed for easy installation, reliable operation, troublefree maintenance and universal applicability. Depending on the application, different sample tubes or prefilters (see data sheets 2-1.1.0.6 and 2-1.1.0.8), not included in the scope of delivery of the probe, are screwed into the thread (G3/4”i) of the filter housing. The large surface ceramic filter element (also glass fiber elements or spun glass fillings are available) is placed in a housing with low stagnant space outside the process. The M&C gas sample probes are designed so that changing a filter element does not involve the use of tools. In this operation the sample line does not need to be removed, thus avoiding contamination of the clean gas path and also maintaining the ­ntegrity of the i system. • Calibration gas will be injected into the probe through a check valve /C directly to the sample outlet. No calibration gas is lost into the stack. • An isolation valve with pneumatic control /I shuts off the sample outlet from the internal filter area. • Through a high flow rate check valve / BB, which is fixed to the internal probe area, blow back of the filter housing area and the insitu probe tube will be done. • Through a high flow rate check valve /BB/F, which is fixed to the filter housing wall, blow back of the incorporated ceramic filter will be done incl. the filter area and the insitu probe tube. The cleaning of the sample tube or the preliminary filter can be effected by extracting the filter from the probe. Technical specifications and illustrations are without obligation, subject to modifications. 03.03/06.06 M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

 Open the catalog to page 1

Technical Data Gas Sample Probe Version Part No. Integrated backflush Protective cover Terminal box Filter housing material Sealing materials Probe flange sealing material Insitu probe tube/prefilter Sample pressure max. Ambient temperature Filter chamber volume Filter element, porosity Thermostat, Temperature adjustment Ready for operation Low temperature alarm contact Sample gas outlet connection Backflush/Test gas connection Shut off valve connection /I Pressure range contol air /I Power supply Electrical connections Electrical equipment standard Mounting flange Stainless steeel 316 /...

 Open the catalog to page 2

345 optional: extension tube Insitu tube Alarm contact 1/4" NPTi Option: 2x Second Sample out optional: extension tube Insitu sample tube Power Alarm contact 230V,50Hz (115V,60Hz) Blowback Control Cal.gas 2-6bar 3-10bar >0,7bar Technical specifications and illustrations are without obligation, subject to modifications. 03.03/06.06 M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

 Open the catalog to page 3

optional: extension tube Insitu sample tube Alarm contact Power 230V,50Hz (115V,60Hz) M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111 Technical specifications and illustrations are without obligation, subject to modifications. 03.03/06.06

 Open the catalog to page 4

All M&C TechGroup Germany GmbH catalogs and technical brochures

  1. Pump series MP

    4 Pages

  2. Cooler EC30C

    3 Pages

  3. SP180-H

    3 Pages

  4. ECPX000C

    4 Pages

  5. FPF-0.1GF

    2 Pages

  6. FP-2GF

    4 Pages

  7. SP2000/V20

    3 Pages

  8. SP2000/PV

    2 Pages

  9. PSP4000-H

    4 Pages

  10. SP3000/Ex

    4 Pages

  11. PSP4000-H/C

    4 Pages

  12. SP3200/EX

    6 Pages

  13. SP3100/EX

    6 Pages

  14. SP2020-H

    2 Pages

  15. SP2200-H/C/I/BB

    2 Pages

  16. SP2300-H

    4 Pages

  17. SP10-H

    3 Pages

  18. SP10

    3 Pages

  19. SP2100-H

    2 Pages

Archived catalogs

  1. SP3000/EX

    4 Pages