HET63

HET63

HET63

Product catalog summary
Overview: The HET63 Electrodeless Conductivity Transmitters are designed for measuring conductivity, solution concentration, salinity, and temperature. They feature an IP66 enclosure and are available in wall or head mounting versions with detachable sensors for easy maintenance.
Key Features:
  • Measurement capabilities include conductivity, % solution concentration, salinity, and temperature.
  • IP66 rated enclosure suitable for harsh environments.
  • Detachable sensors for easy maintenance.
  • Two isolated 4-20mA current outputs and two relay outputs.
  • Relays configurable for on/off, time, and pulse proportional control.
  • Digital inputs for external range changing.
  • Low voltage operation (15-30v DC).
  • Preset curves for common chemicals and salinity, with custom curve entry available.
Models:
  • Base Model: Single 4-20mA output.
  • Advanced Model: Two 4-20mA outputs for measuring and transmitting both process temperature and conductivity or solution concentration.
Specifications:
  • Measurement range: 0 to 999.9 uS/cm, 9.999, 99.99, and 999.9 mS/cm.
  • Accuracy: ±1% of range; Linearity: ±0.1% of range; Repeatability: ±0.1% of range.
  • Temperature range: -50°C to +300°C with ±0.2°C accuracy.
  • Temperature compensation: Automatic or manual, 0 to 100°C.
  • Display: Custom backlit LCD with multilingual options.
  • Power supply: 15–30vDC at 200 mA.
Applications: Ideal for cooling tower bleed, rinse water, and solution concentration applications.
Additional Features:
  • Multi-level security for operator access.
  • Configuration data backup for fast reconfiguration and recovery.
  • Simulate mode for testing set point and current outputs.
  • Automatic temperature compensation standard on all models.
Compliance: Complies with European Directives, EMC 2004/108/EC, and Low Voltage Directive 2006/95/EC.
Order Codes:
  • HET63 1142: Wall mounting transmitter with single 4-20mA output.
  • HET63 1143: Advanced wall mounting transmitter with dual 4-20mA outputs and relay outputs.
  • HET63 1138: Head mounting transmitter with ECS64 sensor and single 4-20mA output.
  • HET63 1139: Advanced head mounting transmitter with ECSC64 sensor, dual 4-20mA outputs, and relay outputs.
See more

Catalog excerpts

HET63-1

HET63Electrodeless Conductivity Transmitters Features • Measurement of conductivity, % solution concentration, salinity and temperature • IP66 enclosure. • Wall mounting or head mounting versions • Detachable sensors on all variants for easy maintenance. • Up to two isolated 4-20mA current and two relay outputs. • Relays can be configured for on/off, time and pulse proportional control • Digital inputs for external range changing • Cable termination via detachable connectors, no need to open the enclosure • Low voltage operation 15-30v DC • Ideal for cooling tower bleed, rinse water and solution concentration applications • Preset curves stored for common chemicals and salinity • Custom curve from 2 to 9 points can be entered.

 Open the catalog to page 1
HET63-2

Two different options of the HET63 transmitter are available with the base model offering a single industry standard 4-20mA output while the Advanced HET63 offers two 4-20mA outputs enabling the process temperature to be measured and transmitted as well as the conductivity or solution concentration value. The Advanced HET63 model also has three digital inputs for remote range changing via a plc. Both of the HET63 versions are available as head mounted transmitters mounted on an ECS64 Electrodeless conductivity sensor or as a wall-mounting version with a detachable cable and separate LTH ECS20...

 Open the catalog to page 2
HET63-3

Specification Measurement input ECS20, ECS40 or ECS60 series Electrodeless conductivity sensor. Cell constant adjustment 0-10.0 Range of measurement 0 to 999.9 uS/cm, 9.999, 99.99 and 999.9 mS/cm. 0 to 16.00% NaOH 0 to 30.00% NaCl 0 to 15.00% HCl 0 to 25.00% H2SO4 0 to 25.00% H3PO4 0 to 41.00 ppt Salinity 0 to 99.9% custom from 2 to 9 points can be entered. Range selection Internal single or auto range, or external range selection via digital inputs. Accuracy ±1% of range Linearity ±0.1% of range Repeatability ±0.1% of range Operator adjustment Solution ±20% offset, Conductivity ±10% slope Connection...

 Open the catalog to page 3
HET63-4

LTH Electronics Ltd, Chaul End Lane, Luton, Bedfordshire. LU4 8EZ England Telephone: +44 (0)1582 593693 Fax: +44 (0)1582 598036 email: [email protected] web: www.lth.co.uk Electronics www.lth.co.uk Type No Part No Description HET63 1142 HET63 IP66 wall mounting transmitter complete with mounting kit. Single 4-20mA output. HET63 1143 HET63 Advanced IP66 wall mounting transmitter complete with mounting kit. 2 x 4-20mA outputs Head Mounted Transmitters Type No Part No Description HET63 1138 HET63 IP66 head mounting transmitter complete with ECS64 Electrodeless conductivity...

 Open the catalog to page 4
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.