CMC15,CMC24 & CMC25    Conductivity & Resistivity Pure water cells-1

CMC15, CMC24 and CMC25 Conductivity Cells for Pure Water Applications Features • High quality robust industrial sensors • IP66 quick release connector simplifies installation and maintenance • 0.5" BSP or triclamp process connection • FDA approved wetted materials Conductivity is an extremely sensitive measurement of the levels of dissolved materials in water. The CMC15, CMC24 & CMC25 cells in conjunction with an analyser are able to detect levels as low as 1ppb concentration. The construction of the cells ensures that they are ideally suited for industrial measurement of pure and ultra pure water while particular attention in the design has been made to eliminate traps where contaminated water may collect. The use of specially treated high-grade Stainless Steel electrodes ensures that the cells do not require initial periodic calibration. Cable connection is made via an IP66 plug that simplifies installation and maintenance. Temperature effects on the measurement of pure water are large and non-linear. The LTH range of cells incorporate a precision PT1000 temperature element whose position in the cells is critical to the overall measurement accuracy. A 0.5°C temperature difference between the element and the solution could introduce an error in excess of 5%.

 Open the catalog to page 1
CMC15,CMC24 & CMC25    Conductivity & Resistivity Pure water cells-2

www.lth.co.uk The CMC15/001/PT43 conductivity cell has a 0.5" BSP process connection and a cell constant of K=0.01. The CMC24 conductivity cell is available with a triclamp process connection and a cell constant of K=0.01 or 0.1. The CMC25/001/PT43 conductivity cell has a triclamp process connection and a cell constant of K=0.01. This cell has a high temperature and pressure specification to allow for steam sterilisation at up to 135°C. The cells can be supplied with traceable certification and are manufactured from FDA approved materials making them ideal for pharmaceutical applications.

 Open the catalog to page 2
CMC15,CMC24 & CMC25    Conductivity & Resistivity Pure water cells-3

HETTEC MATERIALS SILICONE SEALS NETTED MATERIALS EPOXr RESIN SILICONE SEAL

 Open the catalog to page 3
CMC15,CMC24 & CMC25    Conductivity & Resistivity Pure water cells-4

WETTED PARTS. SILICONE SEALS Cable Assemblies COAX SCREEN MIGHT ANGLE

 Open the catalog to page 4
CMC15,CMC24 & CMC25    Conductivity & Resistivity Pure water cells-5

Cable Assemblies for use with HCT63 Series Transmitters REAR VIEW OVERALL OUTER SCREEN REAR VIEW X XXXX X XXXX X X XXXXXX XX COAX SCREEN OVERALL OUTER SCREEN AND COAX SCREENS MUST NOT CONTACT OTHER LEADS OR EACH OTHER conductivity cells Cell constant: Cell constant accuracy: Process connection: Wetted materials: 316L stainless steel, PEEK, silicone seals Maximum operating temperature: Maximum operating pressure: Temperature compensation: conductivity cells Cell constant: Cell constant accuracy: Process connection: Wetted materials: 316L stainless steel, PEEK, silicone seals Maximum operating...

 Open the catalog to page 5
CMC15,CMC24 & CMC25    Conductivity & Resistivity Pure water cells-6

LTH Electronics Ltd, Chaul End Lane, Luton, Bedfordshire. LU4 8EZ England Telephone: +44 (0)1582 593693 Fax: +44 (0)1582 598036 email: [email protected] web: www.lth.co.uk www.lth.co.uk Order Codes Type No K=0.01, 0.5" BSP process connection. Terminated with a C16P connector. K=0.01, 2" triclamp process connection, 70mm insertion. Terminated with a C16P connector. K=0.01, 1.5" triclamp process connection, 70mm insertion. Terminated with a C16P connector. K=0.1, 2" triclamp process connection, 70mm insertion. Terminated with a C16P connector. K=0.1, 1.5" triclamp process connection, 70mm insertion....

 Open the catalog to page 6