AquaCal 2000-1

AquaCal 2000 Pure Water Conductivity and Resistivity Portable Kit Features • Conductivity measurement from 0.055 to 99.99 µS/cm • Scaleable voltage outputs for conductivity, resistivity or temperature • Resistivity measurement to 18.20 MΩ/cm • Alkaline or NiCad rechargeable batteries or mains adaptor operation • Temperature measurement from -10 to 110°C • User selectable security code to protect settings and configuration • Calibrated to traceable standards – conforms to ASTM D 5391-99 and current USP Test Method • Diagnostics to indicate instrument and sensor faults and incorrect configuration • Ideal for validation of pure water treatment plants

 Open the catalog to page 1
AquaCal 2000-2

The AquaCal 2000 hand held system has been developed to provide a method of measuring the conductivity or resistivity of pure water to traceable standards making it suitable to validate pure water treatment plants. Currently it is only possible to calibrate instrumentation on site by using precision resistors, however it is impossible to check the complete measurement system including the cell. The AquaCal 2000 overcomes this problem by providing a portable calibrated system that can be used to check on-line instrumentation by comparison. The AquaCal 2000 measures both nontemperature compensated...

 Open the catalog to page 2
AquaCal 2000-3

Specification Range of measurement Conductivity Resistivity 0-0.9999 µS/cm 0-999.9 KΩ/cm 0-9.999 µS/cm 0-9.999 MΩ/cm 0-99.99 µS/cm 0-99.99 MΩ/cm User selectable or full auto ranging Temperature compensated system accuracy including linearity and repeatability: Conductivity 0.3% of range Ambient temperature: 0-50°C Ambient temperature variation: 0.01% of range Temperature compensation: Selectable in or out fixed curve compensation for ultra pure water. Selectable slope of 0.0 to 3.9%/°C for the impurity contribution Temperature compensation base: Selectable 20 or 25°C Display: 16 x 2 character...

 Open the catalog to page 3
AquaCal 2000-4

LTH Electronics Ltd, Chaul End Lane, Luton, Bedfordshire. LU4 8EZ England Telephone: +44 (0)1582 593693 Fax: +44 (0)1582 598036 email: [email protected] web: www.lth.co.uk www.lth.co.uk Order Codes AquaCal 2000 Pure water measuring kit Type No Portable conductivity, resistivity and temperature measuring kit supplied with traceable certification. Kit comprises of: AquaCal 2000 measurement instrument, CMC26/001/PT43 0.5" BSP Male threaded insertion conductivity cell, 2 metre cell connection cable, stainless steel flow cell carrier with 0.25" BSP inlet & outlets, c/w hose connectors & washers. Carrying...

 Open the catalog to page 4
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.