9000 Series pH and Redox Insertion Sensors Combination glass pH electrode and Redox (ORP) electrode Sealed Ag/AgCl reference with coaxial ceramic junction Single junction reference design 6 Metres of cable with Tag ends The 9000 series pH & Redox sensors are designed for direct installation into a pipe or in the wall of a tank for continuous pH or Redox monitoring and control. These sensors have a large non-clogging peripheral reference junction, which is constructed in annular form around the sensing electrode. The 9000 series sensors have a Kynar body with a 0.5” NPT male threaded process connection and must always be installed at least 15 ° above horizontal for correct operation.
Open the catalog to page 1Temp Range Reference System Reference Electrolyte 3.8M KCl, Saturated with AgCl Glass Membrane Type Membrane Resistance Zero Potential Wetted Materials Kynar, Glass and Ceramic Standard cable length 6 Metres low noise coaxial cable with Tag ends Order Codes: Type No 9027 Description 0.5” NPT Male threaded pH electrode with 6 metres of cable and tag ends 0.5” NPT Male threaded Redox electrode with 6 metres of cable and tag ends These products comply with current European Directives Issue No. 0510