1. Catalogs
  2. Linseis Thermal Analysis
  3. L78 / RITA (DIL Dilatometer)

L78 / RITA (DIL Dilatometer)

L78 / RITA (DIL Dilatometer)

L78 / RITA (DIL Dilatometer)

Product catalog summary
Overview: The document describes the RITA L78 Quenching Dilatometer, a high-speed instrument used to measure dimensional changes in steels during heating and cooling. These changes are due to thermal expansions and phase transformations, which are critical for understanding material properties.
Specifications: The RITA L78 features a high-speed quenching dilatometer with capabilities as per ASTM A 1033 - 04. It includes an induction furnace, LVDT for expansion measurement, and can achieve heating and cooling rates up to 2500 K/sec. The temperature range is from -150°C to 1600°C, with a maximum vacuum of 10E-5mbar. The data acquisition rate is 1000 samples per second, and it can operate in horizontal or vertical modes.
Operation Method: The dilatometer heats a sample to a temperature plateau and then cools it at various rates. The phase transformations are indicated by changes in length, which are used to create TTT (Time Temperature Transformation) and CCT (Continuous Cooling Transformation) diagrams. These diagrams help in selecting the optimal steel and processing route.
Advantages: The L78 RITA system can perform measurements under various atmospheres and temperatures in one run. It offers fast controlled heating and cooling, and can analyze non-metallic samples with an optional susceptor. It is designed for creating CHT, CCT, and TTT diagrams.
Software: The instrument is PC-controlled with software that runs on Microsoft Windows. It includes modules for temperature control, data acquisition, and evaluation. Features include text editing, data security, thermocouple break protection, and data export capabilities. The software supports multi-method analysis and statistical evaluation.
Technical Specifications: The sample holder is made of fused silica or Al2O3, with a diameter of 4-5mm and length of 10mm. The temperature range is from room temperature to 1600°C, with various atmosphere options. Helium gas is recommended for cooling, and the power requirement is 230VAC, 16A, 50-60 Hz.
Applications: The document includes diagrams illustrating austenitisation and phase transformations, highlighting the instrument's capability to measure and analyze these processes.
Contact Information: LINSEIS GmbH in Germany and LINSEIS Inc. in the USA are the contact points for further information and services.
See more

Catalog excerpts

L78 / RITA (DIL Dilatometer)-1

QUENCHING DILATOMETER RITA L78

 Open the catalog to page 1
L78 / RITA (DIL Dilatometer)-2

This practice is based upon the principle that, during heating and cooling of steels, dimensional changes occur as a result of both thermal expansions associated with temperature change and phase transformation. In this practice, sensitive high-speed quenching dilatometer equipment is used to detect and measure the changes in dimension that occur as functions of both time and temperature during dened thermal cycles. The resulting data are converted to discrete values of strain for specic values of time and temperature during the thermal cycle. Strain as a function of time or temperature, or both,...

 Open the catalog to page 2
L78 / RITA (DIL Dilatometer)-3

Operation Method Operation Method of Quenching Dilatometer A solid or hollow sample is continuously heated to a certain temperature plateau. At this plateau a continuous cooling is started. The cooling is repeated at different (linear or exponential) cooling rates. The phase transformations occurring in the different cooling cycles are indicated by the measured change of length. An array of cooling curves with different cool down cycles represents a continuous or isothermal TTT diagram (Time Temperature Transformation diagram). Start and stop of the transformation dene the elds of e.g. ferrite,...

 Open the catalog to page 3
L78 / RITA (DIL Dilatometer)-4

Technical Specifications Features -Software • Program capable of text editing All LINSEIS thermo analytical instruments are PC controlled. The indi- • Data security in case of power failure vidual software modules exclusively run under Microsoft® Windows® • Thermocouple break protection operating systems. The complete software consists of 3 modules: tem- • Repetition measurements with minimum parameter input perature control, data acquisition and data evaluation. The 32 bit soft- • Evaluation of current measurement ware incorporates all essential features for measurement preparation, • Curve...

 Open the catalog to page 4
L78 / RITA (DIL Dilatometer)-8

LINSEIS GmbH Vielitzerstr. 43 95100 Selb Germany Tel.: (+49) 9287–880 - 0 Fax: (+49) 9287–70488 E-mail: [email protected] LINSEIS Inc. 109 North Gold Drive Robbinsville, NJ 08691 USA Tel.: (+1) 609 223 2070 Fax: (+1) 609 223 2074 E-mail: [email protected] Products: DIL, TG, STA, DSC, HDSC, DTA, TMA, MS/FTIR, In-Situ EGA, Laser Flash, Seebeck Effect Services: Service Lab, Calibration Service