General Information
The Leyro Portable Process Calibrator LX529 is a compact, lightweight, and easy-to-use handheld instrument. It features a split-screen display and calibration functions with an accuracy of +0.01%. The device comes with a one-year warranty against factory defects.
Highlights
  • Compact and light design
  • Easy to use
  • Maintenance and service-friendly
Specifications
  • Dimensions: 223.5 x 111 x 45.5 mm
  • Weight: Approximately 670g
  • Power Requirement: 5V, 2A
  • Power Adapter Requirement: 100-240 VAC
  • Operating Temperature Range: -10°C to 55°C
  • Safety Standards: IP54
  • EMC Standards: EN55022, EN55024
  • Vibration: Random, 2g, (5...500)Hz
  • Shock: Half-sine wave, 30g, 11ms
Applications
The LX529 is suitable for on-site calibration across various industries, including machinery construction. It is designed to measure and source volts, current, thermocouples, resistance temperature detectors (RTD), resistance, and frequency.
Performance
  • DC Voltage Accuracy: +/- (0.01% reading + 0.01% range)
  • Frequency Measurement: 1-1100 Hz with 0.1 Hz resolution and +/- 0.05% accuracy
  • Resistance Measurement: 0-400Ω with +/- (0.00-0.01% reading + 0.01% range)
  • RTD Accuracy: PT100 (385) -200 to 800°C with +/- 0.1°C accuracy
  • Thermocouple Accuracy: Type J -200 to 1200°C with +/- 0.2°C accuracy
Standard Delivery
  • Process calibrator
  • MC leads, U-clamp, short jumpers
  • Power adapter and micro USB cable
  • Portable pack
  • User Manual and Conformity Certificate
See more

Catalog excerpts

LX529-1

LX529 The Leyro Portable Process Calibrator LX529 is a handheld, batteryoperated instrument that measures and sources a variety of electrical parameters. It can be widely applied in industrial fields and laboratories. The upper display allows you to measure volts, current and thermocouples. The lower display allows you to measure and source volts, resistance temperature detectors (RTD), thermocouples (TC), resistance and frequency. APPLICATIONS On site calibration Maintenance and service All kind of industry Machinery Constructions HIGHLIGHTS Compact and light Easy to use Accuracy +0.01% Zero Clearing Auto flash terminals General Information Compact, light and easy to use A split-screen display and calibration functions Accuracy +0.01% Warranty: 1 year agains factory defects

 Open the catalog to page 1
LX529-2

Environmental performance Operating Temperature Range(ºC) *All specifications apply from 18 to 28ºC and assume a 5 min warm-up period Safety standards and certifications Physical Dimensions Power requirement Power adapter requirement Sensitivity not less than 1V (peack to peack) Wave form : Square wave Wave form : (0...22)V +-0.5V peack to peack Load drive capability 3mA Excitation current: 0,5 mA Resolution: Excitation current Before sourcing or measuring, offset errors caused by i

 Open the catalog to page 2
LX529-3

3 wire, assumes matched leads with a total resistance not exceeding 100Ω Before sourcing or measuring, offset errors caused by internal elements or external factors must be cleared Thermocouple Type Resolution: 0,1ºC Cold junction error: +-0.5ºC Typical error of the temperature detector :+-0.2ºC Standard delivery: Process calibrator, MC leads, U-clamp, short jumpers Power adapter and micro usb cable Po