1. Catalogs
  2. LEWA GmbH
  3. LEWA microflow-metering pumps FC
video corpo

LEWA microflow-metering pumps FC

LEWA microflow-metering pumps FC

LEWA microflow-metering pumps FC

Product catalog summary
Overview: The LEWA microflow-metering pumps FC are designed for high-pressure applications up to 400 bar, suitable for laboratories and pilot plants. They offer process safety and versatility in applications.
Specifications:
  • Max rod thrust: 800 N
  • Stroke length: Infinitely variable from 0 to 10 mm, adjustable manually or via pneumatic/electric actuator
  • Construction: Solid materials, oil-bath lubrication, weather and splash protection
  • Pump head versions: Diaphragm and plunger
Features:
  • Solid bearings for power transmission
  • Positive spring return for eliminating play
  • High plunger acceleration for improved metering accuracy
  • Operation: Eccentric shaft driven by worm gear, with stroke adjustment via handwheel or actuator
Performance Overview:
  • Theoretical stroke volume and effective flow rates are specified per pump head and stroke frequency
  • Materials: Options include various steels, Hastelloy, PVC, PTFE, and others upon request
  • Standard connections: Internal thread to DIN or NPT, with optional flanges
  • Admissible suction pressure available upon request
Dimensions:
  • Type FC 1: L 375 mm, B 164 mm, ∆B 120 mm, H 400 mm
  • Type FC 3: L 375 mm, B 325 mm, ∆B 297 mm, H 403 mm
  • Dimensions vary based on pump head fitted
See more

Catalog excerpts

LEWA microflow-metering pumps FC-1

Creating Fluid Solutions LEWA MICROFLOW-METERING PUMPS FC The microflow-metering pump FC for pressures up to 400 bar has been developed especially for laboratories and pilot plants, where process safety and a broad range of applications are very important. ■ Stroke length infinitely and linearly variable from 0 to 10 mm while pump stopped or in operation; with manual adjustment (scale division 0.1 mm) or with pneumatic or electric actuator ■ Long life through solid construction, best materials, oil-bath lubrication of all moving parts, weather and splash protection ■ Pump head versions: diaphragm, plunger > D1-502en ■ 08.2011 ■ Subject to change without notice ■ Printed in Germany

 Open the catalog to page 1
LEWA microflow-metering pumps FC-2

LEWA microflow-metering pumps FC Simplex metering pump FC 1 with manual stroke adjustment; diaphragm pump head Simplex metering pump FC 1 with pneumatic stroke actuator PHV-C; diaphragm pump head Features: Solid, statically sound bearings for all power transmitting parts Positive spring return eliminates play High plunger acceleration at part-stroke E limination of play, constant front dead center and high acceleration of plunger improve metering accuracy of small metering rates at part stroke M ethod of operation: The drive rotates, via the worm gear, the eccentric shaft and thus initiates the...

 Open the catalog to page 2
LEWA microflow-metering pumps FC-3

LEWA microflow-metering pumps FC Performance overview Standard plunger diameter [mm] Qtheor [l/h]1) Displacement per pump head at maximum stroke length and stated speed n [min-1] 2) Permissible operating pressure of standard pump heads p [bar gauge] 5) Type Plunger pump heads Diaphragm pump heads 1) theor stroke volume x stroke per minute Q Q eff is quoted on specification sheet For multiplex pump units, determine total metered flow by multiplying by the number of pump heads 2) ptional stroke frequencies at 50 Hz: O n = 26, 43, 52, 80, 143, 160 min -1 3) = 13% Cr steel; 3 = stainless steel CrNiMo...

 Open the catalog to page 3
LEWA microflow-metering pumps FC-4

Creating Fluid Solutions LEWA MICROFLOW-METERING PUMPS FC The dimensions differ depending on the pump head fitted. The dimensions differ depending on the pump head fitted. > D1-502en ■ 08.2011 ■ Subject to change without notice ■ Printed in Germany

 Open the catalog to page 4
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.