Overview: The EL-USB-ULT-LCD+ is a high accuracy, ultra-low temperature cryogenic vaccine data logger designed for monitoring vaccines in cryogenic dry ice storage. It features a high contrast LCD screen and can store over 32,000 temperature readings.
Specifications: The data logger operates within a temperature range of -100 to +100°C (-148 to +212°F) and has an accuracy of ±1.8°C (±3.2°F). It supports logging rates from 1 second to 12 hours and has a battery life of approximately 2 years at 25°C with a 1-minute logging rate.
Display and Status Indication: The LCD displays current, maximum, and minimum temperature readings. Status indications are provided via red and green LEDs, which signal alarm conditions and battery status.
Software: The EasyLog software, available for free download, allows users to set up the logger, download data, and export it in various formats. It supports configuration of parameters such as logger name, measurement units, logging rate, and alarm thresholds.
Battery Information: The logger uses a 3.6V ½AA lithium metal battery, recommended for annual replacement. The logger retains stored readings even when the battery is discharged or replaced, but logging will stop until the battery is replaced and the logger is restarted.
Calibration Certificates: Lascar offers a Traceable Calibration Certificate Service for temperature data loggers, ensuring accuracy through equipment calibrated by accredited laboratories.
Accessories: The package includes a battery, thermocouple probe, and mounting bracket. Additional accessories such as replacement batteries and probes are available.
See more

Catalog excerpts

EL-USB-ULT-LCD+-1

High Accuracy, Ultra Low Temperature Cryogenic Vaccine Data Logger with LCD screen ▪ ▪ ▪ ▪ ▪ ▪ ▪ ▪ ▪ Your data, anytime, anywhere Measurement range -100 to +100°C (-148 to +212°F) High accuracy T-Type probe on 2m cable Stores over 32,000 readings High contrast LCD, with four digit temperature display function Logging rates between 1 second and 12 hours Immediate, delayed and push-to-start logging User-programmable alarm thresholds Status indication via red and green LEDs EasyLog software available as a free download The EL-USB-ULT-LCD+ is designed to monitor vaccines in cryogenic dry ice storage. This standalone data logger measures and stores more than 32,000 temperature readings from its high accuracy thermocouple temperature probe. The user can easily set up the logger and view downloaded data by plugging the data logger into a PC’s USB port and using the free EasyLog software. Data can then be graphed, printed and exported to other applications for detailed analysis. The high contrast LCD can show a variety of temperature information. At the touch of a button, the user can cycle between the current, maximum and minimum stored values for temperature. The data logger is supplied with a lithium metal battery which typically gives 2 years’ logging life. Temperature measurement range Internal resolution Accuracy (overall error) Logging rate User selectable between 1 second & 12 hours Operating temperature range -10 to +40°C (+14 to +104°F) (data logger only) Battery Life 2 years (at 25°C and 1 minute logging rate) Probe dimensions Replacement battery Replacement T-type thermocouple probe Handheld datalogger programmer & collector Thermocouple probe EL-LCD WALL BRACKET Mounting Bracket CALIBRATION CERTIFICATES NOW AVAILABLE Lascar now offers a Traceable Calibration Certificate Service on Temperature Data Loggers. Using reference equipment which has been calibrated by a UKAS/NIST accredited laboratory and using apparatus traceable to national or international standards. For more information, please see www.lascarelectronics.com.

 Open the catalog to page 1
EL-USB-ULT-LCD+-2

High Accuracy, Ultra Low Temperature Cryogenic Vaccine Data Logger with LCD screen Your data, anytime, anywhere EasyLog Software Lascar’s EasyLog control software is available as a free download from www.easylogusb.com. Easy to install and use, the control software is compatible with 32-bit and 64-bit versions of Windows 7, 8 & 10. The software is used to set up the logger, download, graph and annotate data or export in Excel, PDF and jpeg formats. The software allows the following parameters to be configured: ▪ ▪ ▪ ▪ ▪ Logger name Measurement parameter (°C or °F) Logging rate (user selectable...

 Open the catalog to page 2
EL-USB-ULT-LCD+-3

High Accuracy, Ultra Low Temperature Cryogenic Vaccine Data Logger with LCD screen Your data, anytime, anywhere LED STATUS INDICATION EL-USB-ULT-LCD+ features two LEDs: ▪ The first LED flashes red to indicate that the EL-USB-ULT-LCD+ is in an alarm condition. It will flash when the logged temperature has exceeded a Low or High alarm level. ▪ The second LED flashes green to indicate that the EL-USB-ULT-LCD+ is not in an alarm condition. Using the EasyLog software it is possible to set the alarm to remain active even if the reading has returned to normal, in which case the alarm LED will continue...

 Open the catalog to page 3
EL-USB-ULT-LCD+-4

High Accuracy, Ultra Low Temperature Cryogenic Vaccine Data Logger with LCD screen Your data, anytime, anywhere BATTERY INFORMATION Replacement We recommend that you replace the battery annually, or prior to logging critical data. Only use 3.6V ½AA lithium metal batteries. The data logger does not lose its stored readings when the battery is discharged or replaced; however, the data logging process will stop and will not resume until the battery is replaced and the logger restarted by the EasyLog software or an EL-DataPad. Before replacing the battery, remove the data logger from the PC. Please...