1. Catalogs
  2. Larson Davis
  3. CAL150, Precision Acoustic Calibrator

CAL150, Precision Acoustic Calibrator

CAL150, Precision Acoustic Calibrator

CAL150, Precision Acoustic Calibrator

Product catalog summary
Overview: The Larson Davis CAL150 is a precision sound level calibrator designed for both field and laboratory use. It is battery-operated and provides a stable output level of either 94.0 or 114.0 dB at a frequency of 1 kHz, independent of battery condition. The device is compliant with IEC 60942-1:2003 standards and is traceable to the National Institute of Standards and Technology (NIST).
Specifications:
  • Output Levels: 94 dB or 114 dB
  • Output Frequency: 1 kHz
  • Microphone Opening: ½”
  • Compliance: IEC 60942 Class 2, ANSI S1.40-2006 Class 2
  • Battery: 9 V alkaline, with automatic shut-off feature
Applications: The CAL150 is suitable for calibrating sound level meters, noise dosimeters, and noise monitoring stations. It supports ½”, ¼”, and 3/8” microphones with appropriate adaptors.
Environmental Conditions:
  • Operating Temperature: -10 ºC to +50 ºC
  • Storage Temperature: -40 ºC to +60 ºC
  • Humidity Range: 10% to 90% RH (non-condensing)
  • Static Pressure Range: 65 kPa to 108 kPa
Physical Characteristics:
  • Dimensions: 106.1 mm x 63.4 mm x 25.9 mm
  • Weight: 156 g
  • Effective Volume: 4.15 cm³
Accessories: Standard accessories include adaptors for 3/8” and 1/4” microphones, and a 9 V battery. Optional accessories include CAL250 and CAL200 precision acoustic calibrators.
Traceability: The calibrator's accuracy is traceable to NIST, ensuring reliable and standardized measurements.
See more

Catalog excerpts

CAL150, Precision Acoustic Calibrator-1

CAL150 Precision Acoustic Calibrator IEC 60942 Class 2 Handheld Sound Level Calibrator for ½” microphones at 1000 Hz Used for Sound Level Meters and Noise Dosimeters Highlights I Internal battery Output level independent of battery condition Adaptors: 1/4”, 3/8” microphones Applications Field calibration or laboratory calibration of: I Sound level meters Noise dosimeters Noise monitoring stations CAL150 Portable Sound Calibrator The Larson Davis CAL150 Sound Level Calibrator is a battery operated precision microphone calibrator used for the calibration of sound level meters and other sound measurement equipment. It can provide an output level of either 94.0 or 114.0 dB (switch-selectable) at a frequency of 1 kHz. It has been designed for both field and laboratory use and the accuracy has been calibrated to a reference traceable to the National Institute of Standards and Technology (NIST). The Larson Davis CAL150 features a stable sound pressure independent of the battery condition. In addition, the Larson Davis CAL150 will turn off automatically to preserve battery and guarantee a stable output. Larson Davis

 Open the catalog to page 1
CAL150, Precision Acoustic Calibrator-2

Acoustic 114.0 dB and 94.0 dB ± 0.3 dB SPL re: 20µPa (114.0 dB is the principal sound pressure level) -0.12 dB for 1/2” microphones Calibration sound pressure level Equivalent free-field level Frequency Harmonic distortion Stability after pressing On Minimum stabilizing time 10 seconds after coupling microphone and calibrator 101.3 kPa, 23 ºC and 50 % RH. Temperature range SPL variation < ± 0.4 dB SPL variation < ± 0.6 dB Frequency variation < ± 7 Hz SPL variation < ± 0.5 dB Frequency variation < ± 7 Hz Storage temperature Storage humidity Humidity range For use with sound level meters and noise...

 Open the catalog to page 2

All Larson Davis catalogs and technical brochures

  1. AEC206