Water Vapor Permeation and Barrier Test System
6Pages

{{requestButtons}}

Catalog excerpts

Water Vapor Permeation and Barrier Test System - 1

PERME®W3/230 Water Vapor Transmission Rate Test System Professional, High-efficiency and Intelligent WVTR System W3/230 Water Vapor Transmission Rate Test System is designed for the determination of water vapor transmission rate test of film and package specimens. It has three test modes and is applicable to various materials with high, medium and low water vapor permeability. It is also embedded with high precision infrared sensor, which provides a wide test range of 0.0011000g/m2·24h and durable service life. Wide range and high-precision of automatic temperature and humidity control to support various combinations of non-standard test conditions Proportional, cycling and manual testing modes facilitate to test different materials with distinct water vapor permeability Convenient fast-access calibration port for temperature and humidity Reference Standard film or standard weight for fast and accurate calibration

Open the catalog to page 1
Water Vapor Permeation and Barrier Test System - 2

High Efficiency W3/230 system is based on the infrared sensor method and utilizes Labthink’s latest exclusive patent design of three diffusion cells integrated in one instrument for individual or multiple tests, which improves test efficiency by 3 times. It can be easily connected with 9 satellite bases to accomplish up to 30 tests at the same time. Three diffusion cells integrated in one instrument with lower space occupancy rate and higher test efficiency Three distinct or equivalent specimens can be tested individually with independent test results at one operation The system can be...

Open the catalog to page 2
Water Vapor Permeation and Barrier Test System - 3

Test Principle Infrared method is used by W3/230 Water Vapor Transmission Rate Test System: The test specimen is mounted in the diffusion cell, which is subsequently divided into a dry chamber and a controlled-humidity chamber. The dry side of the specimen is swept by a flow of dry nitrogen, and the water vapor permeating through the specimen from the controlled-humidity chamber is carried by dry nitrogen to the infrared sensor where proportion electrical signals will be generated. The water vapor transmission rate is obtained by analyzing and calculating the electrical signals. For package...

Open the catalog to page 3
Water Vapor Permeation and Barrier Test System - 4

Package Caps LCD Monitor Film Solar Back-sheets Test seal performance of different package caps WVTR test of LCD monitor films WVTR test of solar back sheets Test water vapor transmission rate of plastic bottles for drug and health care products , e.g. eye drop bottles, infusion Health Care bags and health care product packages Products Plastic Pipes Pharmaceutical Blister Package Extended Protecting Film Blister package for pharmaceutical products Aseptic Wound Including various sorts of pipes, e.g. PPR Aseptic Wound Protecting Film and Medical Plaster Plaster Plastic fuel tanks are widely...

Open the catalog to page 4
Water Vapor Permeation and Barrier Test System - 5

Temperature control for one Package Ф180 mm, Height380 mm Specimen Size Temperature control for 3 Packages Ф100 mm, Height380 mm No temperature control Number of Specimen 3 (Independent results for each specimen) Extension Specimen Thickness Test Area Temperature Range Temperature Accuracy Humidity Range Humidity Accuracy Carrier Gas Flow of Carrier Gas Pressure of Carrie Gas Port Size Instrument Size Power Supply Weight 0%RH, 35%RH90%RH, 100%RH ±1%RH 99.999% high-purified nitrogenNot in supply scope 0200 ml/min 0.28 MPa/40.6 psi 1/8 inch copper tubing 690 mm (L) × 350 mm (W) × 360 mm (H)...

Open the catalog to page 5
Water Vapor Permeation and Barrier Test System - 6

Standard Configuration Optional Parts Mainframe, Professional Software, Precision Pressure Regulator, Gas Inlet Pipe and Connector, Communication Cable, Round Sample Cutter, Vacuum Grease, Sponge Cushion Satellite Base, Accessories for Packaging Testing, Temperature Control System, Reference Film, Package Mouth Sealing Accessories, Sample Cutter, Vacuum Grease, Printer 1. The gas supply port of the instrument is 1/8 inch copper tubing 2. Customers will need to prepare for gas supply and distilled water.

Open the catalog to page 6