video corpo

OX2/230 Oxygen Transmission Rate Test System
4Pages

{{requestButtons}}

Catalog excerpts

OX2/230 Oxygen Transmission Rate Test System - 1

PERME® OX2/230 Oxygen Transmission Rate Test System Professional, High Efficiency and Intelligent OTR Test System Professional This instrument can be used to measure oxygen transmission rate of film and package specimens. It has three test modes and is applicable to various materials with high, medium and low oxygen permeability. It is also embedded with high precision oxygen sensor, which provides a wide test range of 0.01 ~ 65,000 cm3/m2^d and durable service life. • Wide range and high precision of temperature and humidity control, specimen, to support various combinations of non-standard test conditions • World leading pressure compensation technology ensures accurate and repeatable test results • Standard, proportional and continuous testing modes facilitate to minimize errors caused by human factors during test process • Equipped with catalyst deaerator to purify carrier gas and reduce system errors • 2 types of carrier gases are available: High purity nitrogen gas or hydrogen-mixed nitrogen gas • Convenient fast-access calibration port for temperature and humidity • Reference film for accurate calibration High Efficiency OX2/230 system is based on the equal pressure method and adopts Labthink’s latest exclusive patent design of three diffusion cells integrated in one instrument for individual or multiple tests, which improves test efficiency by 3 times. It can be easily connected with 9 satellite bases together to accomplish up to 30 tests at the same time. • 3 diffusion cells are integrated in one instrument with lower space occupancy rate and higher test efficiency • 3 distinct or equivalent specimens can be tested individually with independent test results at one operation • The system can be easily connected with a maximum of 10 instruments to accomplish up to 30 tests at the same time • Oxygen transmission rate and water vapor transmission rate can be obtained at one operation by working with Labthink’s water vapor transmission rate test system together under the control of one computer Intelligent The instrument utilizes Labthink’s latest operating software with user-friendly operating interface and intelligent data management. It also supports LystemTM Lab Data Sharing System, which ensures uniform management of test results and test reports. • Based on the user-friendly Windows operating interface for easy operation • Saves test data in different formats for convenient data transfer • Intelligent historical data searching, comparison, analysis and printing functions • Supports Lystem™ Lab Data Sharing System for uniform management of test results and test reports Test Principle Labthink Instruments Co., Ltd. 144 Wuyingshan Road, Jinan, P.R.China (250031) Phone: +86-531-85068566 FAX: +86-531-85062108 Labthink International, Inc. 200 River's Edge Drive, Medford, MA, 02155, U.S.A. Phone: +1-617-830-2190 FAX: +1-781-219-3638 www.labt

Open the catalog to page 1
OX2/230 Oxygen Transmission Rate Test System - 2

The pre-conditioned specimen is mounted between the upper and lower chambers at ambient atmospheric pressure. One chamber contains oxygen or air and the other chamber is slowly purged by a stream of nitrogen. Due to the concentration difference between the two chambers, oxygen molecules permeate through the specimen into the nitrogen side and are taken to the coulometric sensor where proportional electrical signals are generated. The oxygen transmission rate is then obtained by analyzing and calculating the signals. For package samples, high purity nitrogen flows inside the package, and...

Open the catalog to page 2
OX2/230 Oxygen Transmission Rate Test System - 3

Number of Specimens Resolution Temperature Range Temperature Accuracy Humidity Range Humidity Accuracy Test Gas Specimen Thickness Test Area 1~3 with independent test results 0.001cm3/(m2-d) 0.00001 cm3/(pkg-d) 15°C - 55°C (standard) _±0.1°C (standard)_ 0%RH, 35%RH - 90%RH ±1%RH Oxygen and air (outside of supply scope) < 3 mm (customization is available for other thickness ) 50 cm2 - Specimen Size Port Size Instrument Dimension Power Supply Net Weight One package with temperature control device: specimen should be smaller than ®150 mm and lower than 380 mm Three packages with...

Open the catalog to page 3
OX2/230 Oxygen Transmission Rate Test System - 4

Labthink Instruments Co., Ltd. 144 Wuyingshan Road, Jinan, P.R.China (250031) Phone: +86-531-85068566 FAX: +86-531-85062108 Labthink International, Inc. 200 River's Edge Drive, Medford, MA, 02155, U.S.A. Phone: +1-617-830-2190 FAX: +1-781-219-3638